Evaluation of an alternative centrifugation protocol for reducing total turnaround time
-
Reyes Nicolás de Blas
, Miriam Menacho Román
Abstract
Objectives
Centrifugation is a key process that should be controlled to ensure an adequate sample quality. To achieve rapid, standardized, and consistent sample centrifugation, we aimed to evaluate an alternative centrifugation protocol and its impact on the results of 20 biochemical determinations in serum.
Methods
The study included 45 ambulatory patients. Two serum-separating tubes were collected from each patient (Becton Dickinson (BD) Vacutainer® SST™ II Advance, 8.5 mL Ref. 366468). One of the tubes was centrifuged at 2,530 × g for 10 min (control method), while the other tube was centrifuged under alternative centrifugation conditions, namely 2,530 × g for 7 min.
Results
The analysis of results revealed that calcium, total bilirubin, and magnesium exhibited a constant and proportional systematic bias. However, considering the proportional mean difference, all analytes met the desirable bias performance established by our laboratory, except for magnesium, which met the minimum bias criteria.
Conclusions
Our study demonstrates that both centrifugation protocols are interchangeable for measuring the studied analytes, thereby ensuring adequate sample quality.
Introduction
Centrifugation is a pre-analytical procedure repeatedly performed in routine clinical laboratory practice. However, there is paucity of data documenting the influence of centrifugation on laboratory results [1], [2], [3], [4], [5], [6]. Centrifugation is a key process that should be controlled to ensure adequate sample quality.
Laboratories are required to establish a standardized centrifugation protocol. This protocol should define the optimal relative centrifugal force (RCF), temperature, and time required to obtain high-quality samples for each type of tube, based on the manufacturer’s recommendations and turnaround time (TAT) requirements. Any deviation from these recommendations requires prior validation [7].
An adequate TAT is crucial for clinical laboratory quality. The laboratory is expected to provide rapid, reliable, and accurate reports enabling appropriate and timely medical decision-making. TAT is considered as a key performance indicator in quality assurance programs. In fact, routine TAT measurement and evaluation are essential to ensure extra-analytical quality. TATs are recorded and reviewed to adopt corrective measures, should it be necessary [8].
General centrifugation recommendations for the serum biochemistry parameters analyzed are 1000–3000 × g for 10–15 min at a temperature of 15 to 24 °C [7]. There are validated alternative centrifugation procedures involving a higher RCF and a shorter centrifugation time, resulting in shorter TATs [7].
The alternative centrifugation conditions (3,000 × g-5 min) validated by Becton Dickinson (BD) in the BD White Paper VS7228 study [9], were proven to be interchangeable with control conditions (1,300 × g-10 min). In our case, since 3000 × g centrifugation was not recommended by the manufacturer (Kubota S700TR), centrifugation was performed at 2,530 × g (maximum speed available) for 7 min [9].
The purpose of this study was to establish a harmonized, standardized, and faster centrifugation protocol for all sample types without significantly affecting plasma/serum quality. As a first approach, the study was conducted only with serum samples for biochemical parameter determination.
In this study, we aimed to assess the clinical performance of BD Vacutainer® SST™ II Ad Advance tubes under centrifugation conditionsof 2,530 × g for 7 min. Control centrifugation conditions were set at 2,530 × g for 10 min.
Materials and methods
The study was conducted by the Department of Clinical Biochemistry of Ramón y Cajal University Hospital (HURYC) between May and June 2024.
The project was approved by the HURYC Ethics Committee under code 046/24.
The sample consisted of 45 ambulatory male and female patients with ages ranging from 18 to 70 years. Informed consent was obtained from all participants. Cancer patients were excluded.
For each patient, two serum-separating tubes were collected (Becton Dickinson (BD) Vacutainer® SST™ II Advance, 8.5 mL Ref. 366468). Blood collection was performed using a Vacuette safety-lok 21G set (0.80 ×19 mm) with a pre-attached holder (Ref. 450085V1 Greiner Bio-One) following the Clinical and Laboratory Standards Institute (CLSI) GP41 guidelines [10] and the joint guidelines of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) and the Latin America Confederation of Clinical Biochemistry [11].
All sample pairs were handled under the same pre-analytical conditions (collection, shipping, centrifugation time and temperature) to limit interference from other variables.
Samples were stored at room temperature in vertical position for a minimum of 30 min to ensure complete clot retraction.
All samples were centrifuged in a Kubota S700TR swinging-bucket rotor centrifuge (Rotor RS-1440 M) (preventive maintenance was performed prior to initiation of the study). As a control method, one of the two samples from each patient was centrifuged at 2,530 × g for 10 min. The other tube was centrifuged at alternative centrifugation conditions, namely 2,530 × g for 7 min. All samples were centrifuged at room temperature (18–22 °C). The order of tube collection and centrifugation protocol for each group was randomized.
A total of 20 biochemical parameters (Table 1) were analyzed on an Alinity c analyzer (Abbott Diagnostics, IL, USA). Samples were processed in duplicate immediately after centrifugation.
Comparison of the results obtained with the two centrifugation conditions. Descriptive study.
Analyte | Units | n | Study range | Mean ± SD/median [p25-p75] | |
---|---|---|---|---|---|
2530 × g 10 min | 2530 × g 7 min | ||||
Glucose | mg/dL | 44 | 72–142 | 95.66 ± 14.27 | 95.90 ± 14.47 |
Sodium | mmol/L | 45 | 136–145 | 140.59 ± 2.02 | 140.62 ± 2.15 |
Potassium | mmol/L | 45 | 3.71–6.57 | 4.54 ± 0.47 | 4.51 ± 0.45 |
Chlorine | mmol/L | 45 | 100–114 | 106.44 ± 2.89 | 106.31 ± 2.97 |
Creatinine | mg/dL | 45 | 0.53–2.82 | 1.02 ± 0.45 | 1.02 ± 0.45 |
Phosphate | mg/dL | 44 | 2.29–4.63 | 3.37 ± 0.59 | 3.36 ± 0.59 |
Calcium | mg/dL | 45 | 8.5–10.9 | 9.60 ± 0.48 | 9.64 ± 0.50 |
Cholesterol | mg/dL | 45 | 107–256 | 178.74 ± 37.74 | 178.49 ± 37.41 |
Total protein | g/dL | 45 | 6.25–8.05 | 7.08 ± 0.37 | 7.08 ± 0.36 |
Total bilirubin | mg/dL | 45 | 0.24–2.69 | 0.66 ± 0.40 | 0.66 ± 0.40 |
Alkaline phosphatase | U/L | 45 | 36–173 | 81.59 ± 72.78 | 81.28 ± 27.74 |
Alanine aminotransferase | U/L | 44 | 7–50 | 20.56 ± 9.67 | 20.55 ± 9.65 |
Lactate dehydrogenase | U/L | 45 | 92–317 | 187.94 ± 37.88 | 188.21 ± 39.29 |
Iron | µg/dL | 45 | 28–159 | 88.30 ± 31.79 | 88.31 ± 31.86 |
Magnesium | mg/dL | 45 | 1.53–2.34 | 1.94 ± 0.16 | 1.97 ± 0.15 |
Urea | mg/dL | 44 | 18–126 | 33.25 [28.50 to 45.75] | 33.00 [28.50 to 46.25] |
Gamma glutamyltransferase | U/L | 45 | 11–247 | 25.00 [20.37 to 36.62] | 25.5 [20.00 to 36.00] |
Aspartate aminotransferase | U/L | 45 | 11–51 | 20.00 [17.87 to 23.12] | 21.00 [17.87 to 23.50] |
-
SD, standard deviation; p25, 25th percentile; p75, 75th percentile.
Statistical analysis
Outliers were excluded according to the recommendations in the CLSI EP09-A2 guidelines [12].
Descriptive analysis of quantitative variables was performed using mean and standard deviation for parametric variables and median with 25th and 75th percentiles for non-parametric variables.
The Kolmogorov-Smirnov test was used to assess the normal distribution of data.
Once the linear relation between the series was confirmed, the level of agreement between the two protocols was evaluated using the Bland-Altman [13] plot and Lin’s concordance correlation coefficient [14].
In accordance with our analytical performance specifications, the maximum allowable systematic error (SE) was based on desirable biological variation (BV) (between- and within-subject variation, CVI and CVG), obtained from the EFLM database, using the following formula [15]:
Estimations were performed using MedCalc® 12.3 (MedCalc Software, Ostende, Belgium). A p-value<0.05 was considered statistically significant.
Results
Uric acid and triglyceride determinations were excluded, as they exceeded the allowable number of outliers for comparison. For all other parameters, adequate values were obtained in a minimum of 44 subjects (Table 1).
Bland-Altman plots revealed that calcium, total bilirubin, and magnesium showed a constant and proportional systematic bias, as the 0 value was not included in any of the confidence intervals generated for absolute and relative differences. However, considering the proportional mean difference, all analytes met the desirable bias requirement accepted by our laboratory, except for magnesium.
As shown in Figure 1, the alternative centrifugation method consistently produced slightly higher magnesium values compared to the control method. This difference kept constant in all the concentrations studied. However, considering biological criteria and despite not meeting the desirable bias criterion, the relative differences obtained for magnesium were lower than the minimum bias recommended by the EFLM [15].
![Figure 1:
Results of Bland-Altman test for magnesium, provided as an example. Yi: mean value for magnesium obtained with the evaluated method (7 min). Xi: mean value for magnesium obtained with the control method (10 min). D’i: D’i percent relative difference: Xi-Yi/[(Xi+Yi)/2]*100. Data are commented in the manuscript.](/document/doi/10.1515/almed-2024-0170/asset/graphic/j_almed-2024-0170_fig_001.jpg)
Results of Bland-Altman test for magnesium, provided as an example. Yi: mean value for magnesium obtained with the evaluated method (7 min). Xi: mean value for magnesium obtained with the control method (10 min). D’i: D’i percent relative difference: Xi-Yi/[(Xi+Yi)/2]*100. Data are commented in the manuscript.
The results obtained for all the analytes included in the study are displayed in Tables 1 and 2.
Comparison of the results obtained with the two centrifugation conditions. Agreement analysis.
Analyte | Bland-Altman |
Limit of maximum acceptable difference, % | Lin’s concordance correlation coefficient |
---|---|---|---|
Relative mean difference (95 %CI) | |||
Glucose | 0.26 (−0.18 to 0.70) | 1.2 | 0.9985 |
Sodium | −0.03 (−0.17 to 0.13) | 0.1 | 0.9421 |
Potassium | 0.43 (−0.64 to 1.51) | 0.8 | 0.9358 |
Chlorine | −0.13 (−0.26 to 0.01) | 0.2 | 0.9855 |
Creatinine: | 0.45 (−0.12 to 1.01) | 2.1 | 0.9991 |
Phosphorus | 0.22 (−0.23 to 0.66) | 1.6 | 0.9970 |
Calcium | −0.42 (−0.65 to −0.19) | 0.8 | 0.9847 |
Cholesterol | 0.11 (−0.20 to 0.43) | 2 | 0.9987 |
Total protein | −0.02 (−0.35 to 0.30) | 0.5 | 0.9774 |
Total bilirubin | −0.71 (−1.36 to −0.06) | 4 | 0.9995 |
Alkaline phosphatase | 0.40 (−0.01 to 0.81) | 2.7 | 0.9991 |
Alanine aminotransferase | 0.23 (−0.74 to 1.20) | 5 | 0.9984 |
Lactate dehydrogenase | −0.07 (−1.55 to 1.41) | 1.6 | 0.9626 |
Iron | −0.02 (−0.59 to 0.56) | 4.8 | 0.9985 |
Magnesium | −1.87 (−2.43 to −1.30) | 2.4 | 0.9448 |
Urea | 0.14 (−0.31 to 0.59) | 3.1 | 0.9997 |
Gamma glutamyltransferase | 0.71 (−0.09 to 1.51) | 5.7 | 0.9998 |
Aspartate aminotransferase | −0.98 (−2.18 to 0.22) | 2.7 | 0.9941 |
-
The desirable bias (%) was established as the allowable limit, except for magnesium (minimum bias)- EFLM database. Lin’s concordance correlation coefficient>0.99: almost perfect agreement; 0.95–0.99: very strong agreement; 0.90–0.95. Moderate agreement; <0.90: poor agreement. 95 %CI, 95 % confidence interval. Results for calcium, total bilirubin and magnesium, as discussed in the text, are indicated in bold.
Discussion
To ensure an adequate TAT, centrifugation conditions can be adjusted as long as sample quality is not compromised.
Following BD-validated recommendations [9], we proposed and evaluated an alternative centrifugation protocol that was feasible in our laboratory (2,530 × g for 7 min).
Multiple studies have demonstrated that alternative centrifugation times and RCFs using different collection tubes and instruments can provide equivalent analytical precision [1], [2], [3], [4], [5]. There is cumulative evidence that centrifugation time can be reduced below 10 min without negatively impacting sample quality for most parameters studied. However, differences in the recommended centrifugation conditions limits the generalization of results from studies using a specific collection tube to tubes from other manufacturers. Other studies demonstrate sample quality variations with RCF exceeding 4000 × g [6].
Further studies are needed to assess the role of other factors related to sample quality, such as gel barrier formation and serum separation as examined by BD in the study BD White Paper VS7228 [9]. Other factors to consider include the presence of residual fibrin or cells in serum/plasma.
Laboratories using centrifugation conditions different from those recommended by the tube and/or reagent manufacturer should first perform a validation study.
In our laboratory, near 3000 samples are centrifuged daily in approximately 40 centrifugation runs, of which 28 % are preferential. Theoretically, a 3-min reduction of centrifugation time would save 120 min per day. A follow-up study is needed to confirm this TAT reduction in real practice.
Given the high sample volume requiring centrifugation and the need for rapid results, the 3-min reduction in the centrifugation protocol is justified.
The results of this study confirm that the two centrifugation protocols are interchangeable for measuring the analytes studied, thereby ensuring adequate sample quality.
A larger-scale study including other parameters and specimens (plasma) would be needed to obtain more robust conclusions.
-
Research ethics: The study was conducted in accordance with the World Medical Association’s Helsinki Declaration on the Ethical Conduct of Research Involving Human Subjects (revised 2013).
-
Informed consent: Informed consent was obtained from all participants, according to a protocol approved by the Ethics Committee of the Ramón y Cajal University Hospital, with code 046/24, as of April 10, 2024.
-
Author contributions: All authors have accepted responsibility for all content of this manuscript and have approved its submission.
-
Use of Large Language Models, AI and Machine Learning Tools: None declared.
-
Conflict of interest: The authors state no conflict of interest.
-
Research funding: None declared.
-
Data availability: Raw data can be obtained on request from the corresponding author.
-
Article Note: The original article can be found here: https://doi.org/10.1515/almed-2023-0119.
References
1. Minder, EI, Schibli, A, Mahrer, D, Nesic, P, Plüer, K. Effects of different centrifugation conditions on clinical chemistry and Immunology test results. BMC Clin Pathol 2011;11:6. https://doi.org/10.1186/1472-6890-11-6.Search in Google Scholar PubMed PubMed Central
2. Cadamuro, J, Mrazek, C, Leichtle, AB, Kipman, U, Felder, TK, Wiedemann, H, et al.. Influence of centrifugation conditions on the results of 77 routine clinical chemistry analytes using standard vacuum blood collection tubes and the new BD-Barricor tubes. Biochem Med (Zagreb). 2018;28. https://doi.org/10.11613/BM.2018.010704.Search in Google Scholar PubMed PubMed Central
3. Møller, MF, Søndergaard, TR, Kristensen, HT, Münster, A-MB. Evaluation of a reduced centrifugation time and higher centrifugal force on various general chemistry and immunochemistry analytes in plasma and serum. Ann Clin Biochem 2017;54:593–600. https://doi.org/10.1177/0004563216674030.Search in Google Scholar PubMed
4. Mensel, B, Wenzel, U, Roser, M, Lüdemann, J, Nauck, M. Considerably reduced centrifugation time without increased hemolysis: evaluation of the new BD Vacutainer SSTTMII Advance. Clin Chem 2007;53:794–5. https://doi.org/10.1373/clinchem.2006.079582.Search in Google Scholar PubMed
5. Holland, L, DomBourian, M. Evaluation of an abbreviated centrifugation protocol for chemistry testing. Lab Med 2012;43:78–81. https://doi.org/10.1309/LM920FFGFXLRAORQ.Search in Google Scholar
6. Miler, M, Nikolac Gabaj, N, Culej, J, Unic, A, Vrtaric, A, Milevoj Kopcinovic, L. Integrity of serum samples is changed by modified centrifugation conditions. Clin Chem Lab Med 2019;57:1882–7. https://doi.org/10.1515/cclm-2019-0244.Search in Google Scholar PubMed
7. Transport, CLSI. Handling. processing, and storage of blood specimens for routine laboratory examinations. In: CLSI guideline PRE-04, 1st ed. USA: Clin Lab Stand Inst; 2023.Search in Google Scholar
8. Asociación Española de Normalización y Certificación. Requisitos particulares para la calidad y la competencia. In: Norma española. UNE 15189:2023. Spain: Asociación Española de Normalización y Certificación; 2023.Search in Google Scholar
9. Becton, Dickinson, Company. Performance of BD Vacutainer® SST™ II advance tubes at four and five minute centrifugation times. Switzerland: Becton, Dickinson and Company. White Paper VS7228; 2004.Search in Google Scholar
10. CLSI. Procedures for collection of diagnostic blood specimens by venipuncture, 7th ed. USA: Clinical and Laboratory Standards Institute; 2017. CLSI guideline GP41.Search in Google Scholar
11. Simundic, AM, Bolenius, K, Cadamuro, J, Church, S, Cornes, MP, van Dongen-Lases, EC, et al.. Joint EFLM-COLABIOCLI Recommendation for venous blood sampling. Clin Chem Lab Med 2018;56:2015–38. https://doi.org/10.1515/cclm-2018-0602.Search in Google Scholar PubMed
12. CLSI. Measurement procedure comparison and bias estimation using patient samples, 3th ed. USA: Clin Lab Stand Inst; 2013. CLSI guideline EP09-A3.Search in Google Scholar
13. Bland, JM, Altman, DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–10.10.1016/S0140-6736(86)90837-8Search in Google Scholar
14. Lin, LI-K. A concordance correlation coefficient to evaluate reproducibility. Biometrics 1989;45:255–68. https://doi.org/10.2307/2532051.Search in Google Scholar
15. Aarsand, AK, Fernandez-Calle, P, Webster, C, Coskun, A, Gonzales-Lao, E, Diaz-Garzon, J, et al.. The EFLM biological variation database. https://biologicalvariation.eu/ [Accessed 05 Jun 2024].Search in Google Scholar
© 2024 the author(s), published by De Gruyter, Berlin/Boston
This work is licensed under the Creative Commons Attribution 4.0 International License.
Articles in the same Issue
- Frontmatter
- Editorial
- Lights and shadows of artificial intelligence in laboratory medicine
- Luces y sombras de la inteligencia artificial en la medicina de laboratorio
- Review / Artículo de Revisión
- Fundamentals of lipoprotein(a) request and quantification in the clinical laboratory
- Aspectos fundamentales en la solicitud y determinación de la lipoproteína(a) en el laboratorio clínico
- Original Article / Artículo Original
- Evaluating seven bioinformatics platforms for tertiary analysis of genomic data from whole exome sequencing in a pilot group of patients
- Evaluación de siete programas bioinformáticos para el análisis terciario de datos genómicos generados a partir de la secuenciación del exoma completo en un grupo piloto de pacientes
- Parameters of glycemic variability in continuous glucose monitoring as predictors of diabetes: a prospective evaluation in a non-diabetic general population
- Parámetros de variabilidad glucémica de la monitorización continua de glucosa como predictores de diabetes: evaluación prospectiva en una población general sin diabetes
- Reference intervals of hematological parameters in the Chilean adult population and the Mapuche ethnic group
- Intervalos de referencia de parámetros hematológicos en población chilena adulta y en la etnia mapuche
- Lipid metabolism in overweight/obese children vs. normal weight children in a north-eastern region of Spain
- Estudio del metabolismo lipídico en niños aragoneses con sobrepeso/obesidad vs. niños normopeso
- Short Communication / Comunicación Breve
- Evaluating the research parameters available on the Sysmex® XN-series hematology analyzers as markers of dysplasia in peripheral blood
- Valoración de los parámetros de investigación de los analizadores hematológicos de la serie XN (Sysmex®) como marcadores de displasia en sangre perifèrica
- Evaluation of an alternative centrifugation protocol for reducing total turnaround time
- Evaluación de un protocolo de centrifugación alternativo que permita reducir el tiempo de respuesta total
- Case Report / Caso Clínico
- Jordans’ anomaly in Chanarin-Dorfman syndrome
- Anomalía de Jordans en síndrome de Chanarin-Dorfman
- Letter to the Editor / Carta al Editor
- Considerations about the use of glucometers in glucose tolerance tests
- Consideraciones acerca del uso de glucómetros durante la prueba de tolerancia oral a la glucosa
- Reply to: “Considerations about the use of glucometers for testing glucose tolerance”
- Respuesta a la carta al editor: “Consideraciones acerca del uso de glucómetros durante la prueba de tolerancia oral a la glucosa” (https://doi.org/10.1515/almed-2024-0108)
Articles in the same Issue
- Frontmatter
- Editorial
- Lights and shadows of artificial intelligence in laboratory medicine
- Luces y sombras de la inteligencia artificial en la medicina de laboratorio
- Review / Artículo de Revisión
- Fundamentals of lipoprotein(a) request and quantification in the clinical laboratory
- Aspectos fundamentales en la solicitud y determinación de la lipoproteína(a) en el laboratorio clínico
- Original Article / Artículo Original
- Evaluating seven bioinformatics platforms for tertiary analysis of genomic data from whole exome sequencing in a pilot group of patients
- Evaluación de siete programas bioinformáticos para el análisis terciario de datos genómicos generados a partir de la secuenciación del exoma completo en un grupo piloto de pacientes
- Parameters of glycemic variability in continuous glucose monitoring as predictors of diabetes: a prospective evaluation in a non-diabetic general population
- Parámetros de variabilidad glucémica de la monitorización continua de glucosa como predictores de diabetes: evaluación prospectiva en una población general sin diabetes
- Reference intervals of hematological parameters in the Chilean adult population and the Mapuche ethnic group
- Intervalos de referencia de parámetros hematológicos en población chilena adulta y en la etnia mapuche
- Lipid metabolism in overweight/obese children vs. normal weight children in a north-eastern region of Spain
- Estudio del metabolismo lipídico en niños aragoneses con sobrepeso/obesidad vs. niños normopeso
- Short Communication / Comunicación Breve
- Evaluating the research parameters available on the Sysmex® XN-series hematology analyzers as markers of dysplasia in peripheral blood
- Valoración de los parámetros de investigación de los analizadores hematológicos de la serie XN (Sysmex®) como marcadores de displasia en sangre perifèrica
- Evaluation of an alternative centrifugation protocol for reducing total turnaround time
- Evaluación de un protocolo de centrifugación alternativo que permita reducir el tiempo de respuesta total
- Case Report / Caso Clínico
- Jordans’ anomaly in Chanarin-Dorfman syndrome
- Anomalía de Jordans en síndrome de Chanarin-Dorfman
- Letter to the Editor / Carta al Editor
- Considerations about the use of glucometers in glucose tolerance tests
- Consideraciones acerca del uso de glucómetros durante la prueba de tolerancia oral a la glucosa
- Reply to: “Considerations about the use of glucometers for testing glucose tolerance”
- Respuesta a la carta al editor: “Consideraciones acerca del uso de glucómetros durante la prueba de tolerancia oral a la glucosa” (https://doi.org/10.1515/almed-2024-0108)