Results of the recalibration of creatinine measurement with the modular Beckman Coulter® Jaffe creatinine method
-
Catherine Fillée
, Myriam Othmane
Abstract
Background: Serum creatinine is important for detecting the beginning of a decline in kidney function. The Beckman Coulter Jaffe reagents for measuring creatinine have been standardized to the internationally accepted reference method: isotope dilution mass spectrometry (IDMS). The impact of this recalibration on the Beckman Coulter® modular or cup (stat) Jaffe method is studied.
Methods: Recalibrated Jaffe (calibrator set points IDMS traceable) and classic National Institute of Standards and Technology (NIST) creatinine methods (traceable to NIST 914a) were compared with an enzymatic method (Sentinel, traceable to NIST SRM 967). All measurements were performed on Synchron DxC 800 systems. Imprecision of the routine methods was studied using the Clinical and Laboratory Standards Institute (CLSI) protocols and laboratory quality survey. Thirteen plasma pools, with concentrations <354 μmol/L, were analyzed with a gas chromatography isotope dilution mass spectrometry (GC-IDMS) method and routine methods. Total error of 8.2% based on biological variability, set on the GC-IDMS values and acceptance criteria (bias <5%, imprecision <8% at concentrations ≥88.4 μmol/L and a maximum 10% increase in the relative error of estimated glomerular filtration rate (eGFR) of the National Kidney Disease Educational Program (NKDEP) were used for evaluating analytical performance of the routine methods studied.
Results: After recalibration of the Jaffe method, median bias (μmol/L) decreased from 12.4 (95% CI: 9.1–21.2) to 7.3 (95% CI: 1.5–10.5). Imprecision of the Jaffe method is in agreement with the claim of the manufacturer, namely <9 μmol/L or <3% below or above 292 μmol/L. Below creatinine of 88.4 μmol/L, imprecision of the recalibrated Jaffe and enzymatic methods is between 4.1% and 6.9%, and 5.0% and 7.1%, respectively, and significantly different (p=0.02 for both the Jaffe and enzymatic methods) from the goal, based on biological variability, of 2.7%. For the adult pools, all recalibrated Jaffe and enzymatic results fall within the total error of 8.2%. In the pediatric samples, one-third of the recalibrated Jaffe and three of the six enzymatic results fall within this total error goal.
Conclusions: Recalibration significantly reduced bias of the Jaffe method. For pediatric samples, recalibrated Jaffe results do not comply with either the imprecision goal or the total error based on biological variability. Adult recalibrated Jaffe results are in compliance with the goals based on biological variability and with the acceptance criteria from the NKDEP.
©2011 by Walter de Gruyter Berlin Boston
Artikel in diesem Heft
- Editorial
- Appropriate labelling of blood collection tubes: a step ahead towards patient’s safety
- Reviews
- New insights into adipose tissue dysfunction in insulin resistance
- Role of biomarkers in cardioncology
- Minireviews
- Measurement of the pro-hormone of brain type natriuretic peptide (proBNP): methodological considerations and pathophysiological relevance
- High-sensitivity troponin assays in the evaluation of patients with acute chest pain in the emergency department
- Opinion Paper
- Regulation of in vitro diagnostics (IVDs) for use in clinical diagnostic laboratories: towards the light or dark in clinical laboratory testing?
- Guidelines and Recommendations
- The Greek way to the Register: the establishment and operation of the Register for Clinical Chemists-Clinical Biochemists in Greece
- Genetics and Molecular Diagnostics
- Rapid HLA-B27 screening with real-time TaqMan PCR: a clinical validation in the Dutch population
- General Clinical Chemistry and Laboratory Medicine
- Results of the recalibration of creatinine measurement with the modular Beckman Coulter® Jaffe creatinine method
- Heterophilic antibody interference in commercial immunoassays; a screening study using paired native and pre-blocked sera
- Oxidative status in the lungs associated with tobacco smoke exposure
- Correlation of cortisol in 1-cm hair segment with salivary cortisol in human: hair cortisol as an endogenous biomarker
- Evaluation of the MassTrak Immunosuppressant XE Kit for the determination of everolimus and cyclosporin A in human whole blood employing isotopically labeled internal standards
- Simultaneous determination of cyclophosphamide and 4-hydroxycyclophosphamide in human plasma by high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry – application to Chinese systemic lupus erythematosus patients
- Sterol profiles in plasma and erythrocyte membranes in patients with Smith-Lemli-Opitz syndrome: a six-year experience
- Chemistry and haematology sample rejection and clinical impact in a tertiary laboratory in Cape Town
- Trefoil factor family peptides are increased in the saliva of children with mucositis
- Validation of commercial assays for measurements of trefoil factor family peptides in serum
- To mix or not to mix venous blood samples collected in vacuum tubes?
- Rapid and simple IgG specific test for the exclusion of heparin induced thrombocytopenia (HIT)
- β-Thalassemia mutations in subjects with borderline HbA2 values: a pilot study in North India
- Cancer Diagnostics
- Genetic variants in miR-146a, miR-149, miR-196a2, miR-499 and their influence on relative expression in lung cancers
- The diagnostic accuracy of two human epididymis protein 4 (HE4) testing systems in combination with CA125 in the differential diagnosis of ovarian masses
- Infectious Diseases
- Measurement of granulocyte maturation may improve the early diagnosis of the septic state
- Cardiovascular Diseases
- Influence of hemolysis on troponin testing: studies on Beckman Coulter UniCel Dxl 800 Accu-TnI and overview of the literature
- Methylenetetrahydrofolate reductase C677T polymorphism and congenital heart disease: a meta-analysis
- Determination of asymmetric dimethyl arginine in human serum by liquid chromatography-tandem mass spectrometry: clinical application in hypertensive subjects
- Diabetes
- The optimal cut-points of HbA1c for detecting newly diagnosed diabetes and pre-diabetes in the Chinese population living in Sichuan
- Letters to the Editor
- Specimen labelling: before or after collection?
- Effect of storage conditions and freeze/thaw cycles on aquaporin-4 antibody (NMO-IgG) serum levels
- Comparison of vancomycin concentrations in blood and interstitial fluid: a possible model for less invasive therapeutic drug monitoring
- Acknowledgement
- Acknowledgement
Artikel in diesem Heft
- Editorial
- Appropriate labelling of blood collection tubes: a step ahead towards patient’s safety
- Reviews
- New insights into adipose tissue dysfunction in insulin resistance
- Role of biomarkers in cardioncology
- Minireviews
- Measurement of the pro-hormone of brain type natriuretic peptide (proBNP): methodological considerations and pathophysiological relevance
- High-sensitivity troponin assays in the evaluation of patients with acute chest pain in the emergency department
- Opinion Paper
- Regulation of in vitro diagnostics (IVDs) for use in clinical diagnostic laboratories: towards the light or dark in clinical laboratory testing?
- Guidelines and Recommendations
- The Greek way to the Register: the establishment and operation of the Register for Clinical Chemists-Clinical Biochemists in Greece
- Genetics and Molecular Diagnostics
- Rapid HLA-B27 screening with real-time TaqMan PCR: a clinical validation in the Dutch population
- General Clinical Chemistry and Laboratory Medicine
- Results of the recalibration of creatinine measurement with the modular Beckman Coulter® Jaffe creatinine method
- Heterophilic antibody interference in commercial immunoassays; a screening study using paired native and pre-blocked sera
- Oxidative status in the lungs associated with tobacco smoke exposure
- Correlation of cortisol in 1-cm hair segment with salivary cortisol in human: hair cortisol as an endogenous biomarker
- Evaluation of the MassTrak Immunosuppressant XE Kit for the determination of everolimus and cyclosporin A in human whole blood employing isotopically labeled internal standards
- Simultaneous determination of cyclophosphamide and 4-hydroxycyclophosphamide in human plasma by high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry – application to Chinese systemic lupus erythematosus patients
- Sterol profiles in plasma and erythrocyte membranes in patients with Smith-Lemli-Opitz syndrome: a six-year experience
- Chemistry and haematology sample rejection and clinical impact in a tertiary laboratory in Cape Town
- Trefoil factor family peptides are increased in the saliva of children with mucositis
- Validation of commercial assays for measurements of trefoil factor family peptides in serum
- To mix or not to mix venous blood samples collected in vacuum tubes?
- Rapid and simple IgG specific test for the exclusion of heparin induced thrombocytopenia (HIT)
- β-Thalassemia mutations in subjects with borderline HbA2 values: a pilot study in North India
- Cancer Diagnostics
- Genetic variants in miR-146a, miR-149, miR-196a2, miR-499 and their influence on relative expression in lung cancers
- The diagnostic accuracy of two human epididymis protein 4 (HE4) testing systems in combination with CA125 in the differential diagnosis of ovarian masses
- Infectious Diseases
- Measurement of granulocyte maturation may improve the early diagnosis of the septic state
- Cardiovascular Diseases
- Influence of hemolysis on troponin testing: studies on Beckman Coulter UniCel Dxl 800 Accu-TnI and overview of the literature
- Methylenetetrahydrofolate reductase C677T polymorphism and congenital heart disease: a meta-analysis
- Determination of asymmetric dimethyl arginine in human serum by liquid chromatography-tandem mass spectrometry: clinical application in hypertensive subjects
- Diabetes
- The optimal cut-points of HbA1c for detecting newly diagnosed diabetes and pre-diabetes in the Chinese population living in Sichuan
- Letters to the Editor
- Specimen labelling: before or after collection?
- Effect of storage conditions and freeze/thaw cycles on aquaporin-4 antibody (NMO-IgG) serum levels
- Comparison of vancomycin concentrations in blood and interstitial fluid: a possible model for less invasive therapeutic drug monitoring
- Acknowledgement
- Acknowledgement