Abstract
Background: Discrepancies in ionised calcium concentrations between results from several point-of-care devices derived from intensive care unit (ICU) patients were postulated to be caused by perchlorate, a thyroid blocking agent. The deviations were serious concerning the diagnosis of hyper- or hypocalcaemia and administration of calcium infusions.
Methods: The problem was studied from three perspectives. First: quantification of ionised calcium in heparinised blood samples spiked with sodium perchlorate from healthy volunteers measured using five blood gas analysers (BGAs from IL, Radiometer, Roche and Siemens). Second: verification of clinical concentrations of perchlorate in blood samples after routine use for blood gas analysis from ICU patients. Third: retrospective analysis of data stored from patients during their stay in general ICU of the Departments of Anesthesio-logy and General Surgery by a patient data management system.
Results: Category 1: three of the point-of-care testing devices measure clinically relevant falsely low ionised calcium concentrations when exposed to concentrations perchlorate >0.1 mmol/L. Two were not able to identify hypercalcaemia of up to 4 mmol/L ionised calcium when specific perchlorate concentrations are exceeded. Category 2: measured clinical concentrations of perchlorate ranged between the lower limit of quantification [LOQ=0.03 mmol/L and 2.75 mmol/L (median=0.29 mmol/L; mean=0.585 mmol/L)]. A concentration above 0.1 mmol/L perchlorate was found in 74.3% of all samples that tested positive. Category 3: between February 2006 and July 2009, 42 patients per year (2.2%) received sodium perchlorate with a median length of treatment of 25 days.
Conclusions: Perchlorate causes clinically relevant lowering of measured ionised calcium in BGAs from two providers at concentrations obtained in samples from ICU patients (affecting about 1.1% of all ICU patients).
©2011 by Walter de Gruyter Berlin Boston
Articles in the same Issue
- Editors’ and Publisher’s Note
- CCLM Award for The Most Cited Paper Recently Published
- Reviews
- Some views on proteomics in diabetes
- Cancer biomarkers for hepatocellular carcinomas: from traditional markers to recent topics
- Minireviews
- Low-density lipoprotein receptor-related protein 1: new functions for an old molecule
- Could serum measurements of S100 proteins be reliable markers to predict recurrence in meningiomas?
- Opinion Paper
- Focusing on the clinical impact of standardization of creatinine measurements: a report by the EFCC Working Group on Creatinine Standardization
- General Clinical Chemistry and Laboratory Medicine
- Use of a fully automated immunoassay for celiac disease screening in a pediatric population
- Evaluation of four commercial biuret reagent kits of serum total protein by the American Association for Clinical Chemistry reference measurement procedure
- Renal function parameters during early and late recovery periods following an all-out 21-km run in trained adolescent runners
- Increased urinary neutrophil gelatinase associated lipocalin in urinary tract infections and leukocyturia
- The severity of pseudohyperkalaemia is not dependent upon the stage of chronic kidney disease: a prospective study
- Modified phosphatidylserine-dependent antithrombin ELISA enables identification of patients negative for other antiphospholipid antibodies and also detects low avidity antibodies
- Perchlorate (Irenat®) may falsely lower measured ionised calcium
- Comparison of thyroglobulin antibody interference in first- and second-generation thyroglobulin immunoassays
- A case of consistent discrepancies between urine and blood human chorionic gonadotropin measurements
- Reference Values and Biological Variations
- Reference change values for monitoring dehydration
- Cancer Diagnostics
- Diagnostic value of pro-gastrin-releasing peptide for small cell lung cancer: a meta-analysis
- Plasma concentrations of angiogenetic factors and angiogenetic inhibitors in patients with ductal pancreatic neoplasms. A pilot study
- Cardiovascular Diseases
- High-sensitivity troponin T improves the prognostic value of N-terminal pro-B-type natriuretic peptide in patients with stable coronary artery disease: results from the LURIC Study
- Letters to the Editor
- Results of a pilot external quality assessment study on free protoporphyrin in erythrocytes
- Trueness assessment of Korean nation-wide glucose proficiency testing
- Genetic polymorphism of cytochrome P450 1A1 in a northern Italian population: high frequency of the CYP1A1*2C allele
- Prestaining of PCR products with SYBR Green for agarose gel electrophoresis: advantages and limitations
- A bias in quantitative RT-PCR limit of detection is induced by the use of cancer cell lines in the molecular detection of circulating tumor cells
- Treatment of hepatitis B virus-associated membranous nephritis patients in Chinese: an open parallel controlled trial
- Evidence on the cause of false positive troponin I results with the Beckman AccuTnI method
- Cross-reactivity of insulin analogs with the Diasorin Liaison Insulin assay
- Improved turnaround time for neonatal hematology profile tests (complete blood count) using a new microcollection tube
- Erratum
- Pharmacogenetics of tacrolimus after renal transplantation: analysis of polymorphisms in genes encoding 16 drug metabolizing enzymes
Articles in the same Issue
- Editors’ and Publisher’s Note
- CCLM Award for The Most Cited Paper Recently Published
- Reviews
- Some views on proteomics in diabetes
- Cancer biomarkers for hepatocellular carcinomas: from traditional markers to recent topics
- Minireviews
- Low-density lipoprotein receptor-related protein 1: new functions for an old molecule
- Could serum measurements of S100 proteins be reliable markers to predict recurrence in meningiomas?
- Opinion Paper
- Focusing on the clinical impact of standardization of creatinine measurements: a report by the EFCC Working Group on Creatinine Standardization
- General Clinical Chemistry and Laboratory Medicine
- Use of a fully automated immunoassay for celiac disease screening in a pediatric population
- Evaluation of four commercial biuret reagent kits of serum total protein by the American Association for Clinical Chemistry reference measurement procedure
- Renal function parameters during early and late recovery periods following an all-out 21-km run in trained adolescent runners
- Increased urinary neutrophil gelatinase associated lipocalin in urinary tract infections and leukocyturia
- The severity of pseudohyperkalaemia is not dependent upon the stage of chronic kidney disease: a prospective study
- Modified phosphatidylserine-dependent antithrombin ELISA enables identification of patients negative for other antiphospholipid antibodies and also detects low avidity antibodies
- Perchlorate (Irenat®) may falsely lower measured ionised calcium
- Comparison of thyroglobulin antibody interference in first- and second-generation thyroglobulin immunoassays
- A case of consistent discrepancies between urine and blood human chorionic gonadotropin measurements
- Reference Values and Biological Variations
- Reference change values for monitoring dehydration
- Cancer Diagnostics
- Diagnostic value of pro-gastrin-releasing peptide for small cell lung cancer: a meta-analysis
- Plasma concentrations of angiogenetic factors and angiogenetic inhibitors in patients with ductal pancreatic neoplasms. A pilot study
- Cardiovascular Diseases
- High-sensitivity troponin T improves the prognostic value of N-terminal pro-B-type natriuretic peptide in patients with stable coronary artery disease: results from the LURIC Study
- Letters to the Editor
- Results of a pilot external quality assessment study on free protoporphyrin in erythrocytes
- Trueness assessment of Korean nation-wide glucose proficiency testing
- Genetic polymorphism of cytochrome P450 1A1 in a northern Italian population: high frequency of the CYP1A1*2C allele
- Prestaining of PCR products with SYBR Green for agarose gel electrophoresis: advantages and limitations
- A bias in quantitative RT-PCR limit of detection is induced by the use of cancer cell lines in the molecular detection of circulating tumor cells
- Treatment of hepatitis B virus-associated membranous nephritis patients in Chinese: an open parallel controlled trial
- Evidence on the cause of false positive troponin I results with the Beckman AccuTnI method
- Cross-reactivity of insulin analogs with the Diasorin Liaison Insulin assay
- Improved turnaround time for neonatal hematology profile tests (complete blood count) using a new microcollection tube
- Erratum
- Pharmacogenetics of tacrolimus after renal transplantation: analysis of polymorphisms in genes encoding 16 drug metabolizing enzymes