Abstract
Background: In China, the traceability of clinical chemistry methods is still immature. Therefore, it is necessary to establish a reference measurement procedure and evaluate commercial reagent kits using such established procedures.
Methods: We reproduced the reference measurement procedure for serum total protein, as recommended by the American Association for Clinical Chemistry (AACC). We evaluated the performance by Clinical Laboratory Standard Institute (CLSI) guidelines EP15-A and EP6-A. Subsequently, four commercial reagent kits were evaluated by the reproduced reference procedure following CLSI guideline EP9-A2.
Results: The performance of the reproduced reference procedure was as follows: CVs ranged from 0.47% to 0.85% at medical decision levels (Xc) of 45 g/L, 60 g/L and 80 g/L. Linearity was Y=1.0022X–0.2121 (r=0.9999), and recovery ranged from 100.2% to 102.4%. The External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine (RELA) was applied, and the result was within the limit of equivalence. The linear relationships of four commercial reagent kits, Merit Choice, KHB, Leadman, and Olympus, were, respectively: Y=0.9922X+0.5776 (r=0.9961); Y=0.9936X+0.4316 (r=0.9992); Y=0.9949X+0.9129 (r=0.9987) and Y=0.9923X+0.8876 (r=0.9989). KHB showed slight negative bias, and the mean±SD was –0.03±0.60 g/L. Merit Choice, Leadman, and Olympus all showed positive bias, and the mean±SDs were 0.02±0.63 g/L, 0.55±0.77 g/L and 0.34±0.71 g/L, respectively.
Conclusions: The correlation and bias of four commercial reagent kits for serum total protein were found to be acceptable. Thus, these reagent kits can be used reliably in China.
©2011 by Walter de Gruyter Berlin Boston
Articles in the same Issue
- Editors’ and Publisher’s Note
- CCLM Award for The Most Cited Paper Recently Published
- Reviews
- Some views on proteomics in diabetes
- Cancer biomarkers for hepatocellular carcinomas: from traditional markers to recent topics
- Minireviews
- Low-density lipoprotein receptor-related protein 1: new functions for an old molecule
- Could serum measurements of S100 proteins be reliable markers to predict recurrence in meningiomas?
- Opinion Paper
- Focusing on the clinical impact of standardization of creatinine measurements: a report by the EFCC Working Group on Creatinine Standardization
- General Clinical Chemistry and Laboratory Medicine
- Use of a fully automated immunoassay for celiac disease screening in a pediatric population
- Evaluation of four commercial biuret reagent kits of serum total protein by the American Association for Clinical Chemistry reference measurement procedure
- Renal function parameters during early and late recovery periods following an all-out 21-km run in trained adolescent runners
- Increased urinary neutrophil gelatinase associated lipocalin in urinary tract infections and leukocyturia
- The severity of pseudohyperkalaemia is not dependent upon the stage of chronic kidney disease: a prospective study
- Modified phosphatidylserine-dependent antithrombin ELISA enables identification of patients negative for other antiphospholipid antibodies and also detects low avidity antibodies
- Perchlorate (Irenat®) may falsely lower measured ionised calcium
- Comparison of thyroglobulin antibody interference in first- and second-generation thyroglobulin immunoassays
- A case of consistent discrepancies between urine and blood human chorionic gonadotropin measurements
- Reference Values and Biological Variations
- Reference change values for monitoring dehydration
- Cancer Diagnostics
- Diagnostic value of pro-gastrin-releasing peptide for small cell lung cancer: a meta-analysis
- Plasma concentrations of angiogenetic factors and angiogenetic inhibitors in patients with ductal pancreatic neoplasms. A pilot study
- Cardiovascular Diseases
- High-sensitivity troponin T improves the prognostic value of N-terminal pro-B-type natriuretic peptide in patients with stable coronary artery disease: results from the LURIC Study
- Letters to the Editor
- Results of a pilot external quality assessment study on free protoporphyrin in erythrocytes
- Trueness assessment of Korean nation-wide glucose proficiency testing
- Genetic polymorphism of cytochrome P450 1A1 in a northern Italian population: high frequency of the CYP1A1*2C allele
- Prestaining of PCR products with SYBR Green for agarose gel electrophoresis: advantages and limitations
- A bias in quantitative RT-PCR limit of detection is induced by the use of cancer cell lines in the molecular detection of circulating tumor cells
- Treatment of hepatitis B virus-associated membranous nephritis patients in Chinese: an open parallel controlled trial
- Evidence on the cause of false positive troponin I results with the Beckman AccuTnI method
- Cross-reactivity of insulin analogs with the Diasorin Liaison Insulin assay
- Improved turnaround time for neonatal hematology profile tests (complete blood count) using a new microcollection tube
- Erratum
- Pharmacogenetics of tacrolimus after renal transplantation: analysis of polymorphisms in genes encoding 16 drug metabolizing enzymes
Articles in the same Issue
- Editors’ and Publisher’s Note
- CCLM Award for The Most Cited Paper Recently Published
- Reviews
- Some views on proteomics in diabetes
- Cancer biomarkers for hepatocellular carcinomas: from traditional markers to recent topics
- Minireviews
- Low-density lipoprotein receptor-related protein 1: new functions for an old molecule
- Could serum measurements of S100 proteins be reliable markers to predict recurrence in meningiomas?
- Opinion Paper
- Focusing on the clinical impact of standardization of creatinine measurements: a report by the EFCC Working Group on Creatinine Standardization
- General Clinical Chemistry and Laboratory Medicine
- Use of a fully automated immunoassay for celiac disease screening in a pediatric population
- Evaluation of four commercial biuret reagent kits of serum total protein by the American Association for Clinical Chemistry reference measurement procedure
- Renal function parameters during early and late recovery periods following an all-out 21-km run in trained adolescent runners
- Increased urinary neutrophil gelatinase associated lipocalin in urinary tract infections and leukocyturia
- The severity of pseudohyperkalaemia is not dependent upon the stage of chronic kidney disease: a prospective study
- Modified phosphatidylserine-dependent antithrombin ELISA enables identification of patients negative for other antiphospholipid antibodies and also detects low avidity antibodies
- Perchlorate (Irenat®) may falsely lower measured ionised calcium
- Comparison of thyroglobulin antibody interference in first- and second-generation thyroglobulin immunoassays
- A case of consistent discrepancies between urine and blood human chorionic gonadotropin measurements
- Reference Values and Biological Variations
- Reference change values for monitoring dehydration
- Cancer Diagnostics
- Diagnostic value of pro-gastrin-releasing peptide for small cell lung cancer: a meta-analysis
- Plasma concentrations of angiogenetic factors and angiogenetic inhibitors in patients with ductal pancreatic neoplasms. A pilot study
- Cardiovascular Diseases
- High-sensitivity troponin T improves the prognostic value of N-terminal pro-B-type natriuretic peptide in patients with stable coronary artery disease: results from the LURIC Study
- Letters to the Editor
- Results of a pilot external quality assessment study on free protoporphyrin in erythrocytes
- Trueness assessment of Korean nation-wide glucose proficiency testing
- Genetic polymorphism of cytochrome P450 1A1 in a northern Italian population: high frequency of the CYP1A1*2C allele
- Prestaining of PCR products with SYBR Green for agarose gel electrophoresis: advantages and limitations
- A bias in quantitative RT-PCR limit of detection is induced by the use of cancer cell lines in the molecular detection of circulating tumor cells
- Treatment of hepatitis B virus-associated membranous nephritis patients in Chinese: an open parallel controlled trial
- Evidence on the cause of false positive troponin I results with the Beckman AccuTnI method
- Cross-reactivity of insulin analogs with the Diasorin Liaison Insulin assay
- Improved turnaround time for neonatal hematology profile tests (complete blood count) using a new microcollection tube
- Erratum
- Pharmacogenetics of tacrolimus after renal transplantation: analysis of polymorphisms in genes encoding 16 drug metabolizing enzymes