A case of consistent discrepancies between urine and blood human chorionic gonadotropin measurements

  • Arjan Albersen EMAIL logo , Evelien Kemper-Proper , Marc H.M. Thelen , Noush A. Kianmanesh Rad , Robert F. Hoedemaeker and Lianne S.M. Boesten
Published/Copyright: March 17, 2011

Abstract

Background: Our laboratory was confronted with two successive urine samples from a single patient which tested positive for human chorionic gonadotropin (hCG) when tested with both qualitative and quantitative assays, combined with no detectable hCG in corresponding plasma samples.

Methods: Serial dilution and recovery experiments were performed in order to investigate the presence of interfering substances or a high-dose hook effect. The ovarian cysts that were removed from this patient were immunohistochemically stained using polyclonal anti-human hCG antibodies. Furthermore, a urine sample was sent to the USA hCG Reference Service for hCG variant analysis.

Results: Dilution and recovery experiments in urine and plasma samples were unremarkable. The biopsy stained negative for human hCG and free β-subunit. hCG isoform analysis in the urine sample revealed that approximately 87.5% of the immunoreactive hCG lacked the β-subunit C-terminal peptide (CTP).

Conclusions: We report a rare case in which two successive urine samples test positive for hCG whereas in corresponding plasma samples hCG is undetectable. The majority of the total hCG contained a degraded form of β-subunit that lacks the CTP. This hCG variant, possibly of pituitary origin, is thought to have an extreme fast clearance rate possibly explaining the discordance between the hCG results in urine and plasma samples.


Corresponding author: Arjan Albersen, Department of Clinical Chemistry (AKL), IJsselland Ziekenhuis, Post box 690, 2900 AR Capelle a/d IJssel, The Netherlands Phone: +3110 2583139, Fax: +3110 2585393

Received: 2010-9-15
Accepted: 2011-1-9
Published Online: 2011-03-17
Published in Print: 2011-06-01

©2011 by Walter de Gruyter Berlin Boston

Articles in the same Issue

  1. Editors’ and Publisher’s Note
  2. CCLM Award for The Most Cited Paper Recently Published
  3. Reviews
  4. Some views on proteomics in diabetes
  5. Cancer biomarkers for hepatocellular carcinomas: from traditional markers to recent topics
  6. Minireviews
  7. Low-density lipoprotein receptor-related protein 1: new functions for an old molecule
  8. Could serum measurements of S100 proteins be reliable markers to predict recurrence in meningiomas?
  9. Opinion Paper
  10. Focusing on the clinical impact of standardization of creatinine measurements: a report by the EFCC Working Group on Creatinine Standardization
  11. General Clinical Chemistry and Laboratory Medicine
  12. Use of a fully automated immunoassay for celiac disease screening in a pediatric population
  13. Evaluation of four commercial biuret reagent kits of serum total protein by the American Association for Clinical Chemistry reference measurement procedure
  14. Renal function parameters during early and late recovery periods following an all-out 21-km run in trained adolescent runners
  15. Increased urinary neutrophil gelatinase associated lipocalin in urinary tract infections and leukocyturia
  16. The severity of pseudohyperkalaemia is not dependent upon the stage of chronic kidney disease: a prospective study
  17. Modified phosphatidylserine-dependent antithrombin ELISA enables identification of patients negative for other antiphospholipid antibodies and also detects low avidity antibodies
  18. Perchlorate (Irenat®) may falsely lower measured ionised calcium
  19. Comparison of thyroglobulin antibody interference in first- and second-generation thyroglobulin immunoassays
  20. A case of consistent discrepancies between urine and blood human chorionic gonadotropin measurements
  21. Reference Values and Biological Variations
  22. Reference change values for monitoring dehydration
  23. Cancer Diagnostics
  24. Diagnostic value of pro-gastrin-releasing peptide for small cell lung cancer: a meta-analysis
  25. Plasma concentrations of angiogenetic factors and angiogenetic inhibitors in patients with ductal pancreatic neoplasms. A pilot study
  26. Cardiovascular Diseases
  27. High-sensitivity troponin T improves the prognostic value of N-terminal pro-B-type natriuretic peptide in patients with stable coronary artery disease: results from the LURIC Study
  28. Letters to the Editor
  29. Results of a pilot external quality assessment study on free protoporphyrin in erythrocytes
  30. Trueness assessment of Korean nation-wide glucose proficiency testing
  31. Genetic polymorphism of cytochrome P450 1A1 in a northern Italian population: high frequency of the CYP1A1*2C allele
  32. Prestaining of PCR products with SYBR Green for agarose gel electrophoresis: advantages and limitations
  33. A bias in quantitative RT-PCR limit of detection is induced by the use of cancer cell lines in the molecular detection of circulating tumor cells
  34. Treatment of hepatitis B virus-associated membranous nephritis patients in Chinese: an open parallel controlled trial
  35. Evidence on the cause of false positive troponin I results with the Beckman AccuTnI method
  36. Cross-reactivity of insulin analogs with the Diasorin Liaison Insulin assay
  37. Improved turnaround time for neonatal hematology profile tests (complete blood count) using a new microcollection tube
  38. Erratum
  39. Pharmacogenetics of tacrolimus after renal transplantation: analysis of polymorphisms in genes encoding 16 drug metabolizing enzymes
Downloaded on 6.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2011.158/pdf
Scroll to top button