Analytical validation of the new plasma calibrated Accu-Chek® Test Strips (Roche Diagnostics)
-
Cécile Meex
Abstract
Background: The Accu-Chek Inform glucose monitor is a point-of-care system for testing blood glucose. New test strips, calibrated to deliver glucose plasma-like values, were launched on the market in May 2005. The aim of our study was to perform analytical validation of these new strips.
Methods: We compared the new plasma strips with whole blood strips; results for the plasma strips with plasma values obtained using a clinical analyzer and with whole blood values given by the glucose electrode of a blood gas analyzer; and the influence of the type of blood (capillary or venous) on the results obtained by the glucose monitor with the plasma calibrated strips.
Results: Plasma strips give on average 7% higher results than the previous whole blood strips. However, the results given by the plasma strips on capillary whole blood, even if well correlated, are not completely comparable with those given by an analyzer for venous plasma. Nevertheless, these plasma strips and the glucose electrode of a blood gas analyzer give comparable results.
Conclusions: Accu-Chek Inform plasma strips are a good method for monitoring of blood glucose values in patients with diabetes.
Clin Chem Lab Med 2006;44:1376–8.
References
1. IFCC recommendation on reporting results for blood glucose. Clin Chim Acta 2001;307:205–9.10.1016/S0009-8981(01)00431-4Suche in Google Scholar
2. Kilpatrick ES, Rumley AG, Myint H, Dominiczak MH, Small M. The effect of variations in haematocrit, mean cell volume and red blood cell count on reagent strip tests for glucose. Ann Clin Biochem 1993; 30:485–7.10.1177/000456329303000513Suche in Google Scholar PubMed
3. Rao LV, Jakubiak F, Sidwell JS, Winkelman JW, Snyder ML. Accuracy evaluation of a new glucometer with automated hematocrit measurement and correction. Clin Chim Acta 2005; 356:178–83.10.1016/j.cccn.2005.01.027Suche in Google Scholar PubMed
4. Stahl M, Brandslund I. Measurement of glucose content in plasma from capillary blood in diagnosis of diabetes mellitus. Scand J Clin Lab Invest 2003; 63:431–40.10.1080/00365510310002590Suche in Google Scholar PubMed
5. Chan AY, Swaminathan R, Cockram CS. Effectiveness of sodium fluoride as a preservative of glucose in blood. Clin Chem 1989; 35:315–7.10.1093/clinchem/35.2.315Suche in Google Scholar
©2006 by Walter de Gruyter Berlin New York
Artikel in diesem Heft
- Renal function – estimation of glomerular filtration rate
- Research translation: a new frontier for clinical laboratories
- Association of APOA5 c.553G>T polymorphism with type 2 diabetes mellitus in a Chinese population
- MTRR 66A>G polymorphism in relation to congenital heart defects
- Increased homocysteine in heart failure: a result of renal impairment?
- Urine flow cytometry and detection of glomerular hematuria
- Chymotrypsin effects on the determination of sperm parameters and seminal biochemistry markers
- Evaluation of cardiac involvement following major orthopedic surgery
- Increased sensitivity in detecting renal impairments by quantitative measurement of marker protein excretion compared to detection of pathological particles in urine sediment analysis
- Clinical chemistry reference values for 75-year-old apparently healthy persons
- Serum pro-hepcidin concentrations and their responses to oral iron supplementation in healthy subjects manifest considerable inter-individual variation
- Comparability of five analytical systems for the determination of triiodothyronine, thyroxine and thyroid-stimulating hormone
- Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100
- Automation and validation of a fast method for the assessment of in vivo oxidative stress levels
- Analytical validation of the new plasma calibrated Accu-Chek® Test Strips (Roche Diagnostics)
- Use of insulin immunoassays in clinical studies involving rapid-acting insulin analogues: Bi-insulin IRMA preliminary assessment
- Analytical and clinical evaluation of a new heart-type fatty acid-binding protein automated assay
- A caveat for OV-Monitor (CA 125 antigen) measurement: something is improving, something is not
- Reply to the letter written by Dorizzi et al.
Artikel in diesem Heft
- Renal function – estimation of glomerular filtration rate
- Research translation: a new frontier for clinical laboratories
- Association of APOA5 c.553G>T polymorphism with type 2 diabetes mellitus in a Chinese population
- MTRR 66A>G polymorphism in relation to congenital heart defects
- Increased homocysteine in heart failure: a result of renal impairment?
- Urine flow cytometry and detection of glomerular hematuria
- Chymotrypsin effects on the determination of sperm parameters and seminal biochemistry markers
- Evaluation of cardiac involvement following major orthopedic surgery
- Increased sensitivity in detecting renal impairments by quantitative measurement of marker protein excretion compared to detection of pathological particles in urine sediment analysis
- Clinical chemistry reference values for 75-year-old apparently healthy persons
- Serum pro-hepcidin concentrations and their responses to oral iron supplementation in healthy subjects manifest considerable inter-individual variation
- Comparability of five analytical systems for the determination of triiodothyronine, thyroxine and thyroid-stimulating hormone
- Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100
- Automation and validation of a fast method for the assessment of in vivo oxidative stress levels
- Analytical validation of the new plasma calibrated Accu-Chek® Test Strips (Roche Diagnostics)
- Use of insulin immunoassays in clinical studies involving rapid-acting insulin analogues: Bi-insulin IRMA preliminary assessment
- Analytical and clinical evaluation of a new heart-type fatty acid-binding protein automated assay
- A caveat for OV-Monitor (CA 125 antigen) measurement: something is improving, something is not
- Reply to the letter written by Dorizzi et al.