Multicenter evaluation of a new immunoassay for procalcitonin measurement on the Kryptor® System
-
Gerald Steinbach
Abstract
We compared the manually performed LUMItest® procalcitonin (PCT) assay with the newly developed fully mechanized Kryptor® PCT assay and determined the essential assay characteristics of this assay.
The new Kryptor® PCT assay was evaluated according to modified NCCLS EP-10/EP-6 protocols in five different laboratories. Samples from 696 patients were assayed using the original LUMItest® PCT assay and the new Kryptor® PCT assay. Possible interference by hemoglobin, triglycerides and bilirubin was evaluated by spiking patient plasma with the appropriate substances. The functional assay sensitivity (FAS) was determined by analyzing samples with low PCT concentrations.
The FAS of the new Kryptor® PCT assay was 0.04 ng/ml and the imprecision within- and between-series below 5% and below 10%, respectively. Within the smallest range of determination, from 0.3 ng/ml to 50 ng/ml, common to the LUMItest® PCT assay (x) and the Kryptor® PCT assay (y) the values correlated well: y=0.64+0.94x, s.xy=2.78 ng/ml.
The performance characteristics of the Kryptor® PCT assay are fully compatible with the intended clinical use. The assay allows determination of PCT in a turnaround time (TAT) of about 20 minutes and thus is adequate for STAT analyses.
Copyright © 2004 by Walter de Gruyter GmbH & Co. KG
Articles in the same Issue
- How useful are haematology analyser flags?
- Intravenous immunoglobulin (IVIG): modes of action in the clinical management of recurrent pregnancy loss (RPL) and selected autoimmune disorders
- Automated flagging influences the inconsistency and bias of band cell and atypical lymphocyte morphological differentials
- Extracellular concentration of homocysteine in human cell lines is influenced by specific inhibitors of cyst(e)ine transport
- Generation of eight adjacent mutations in a single step using a site-directed mutagenesis kit
- Prohepcidin accumulates in renal insufficiency
- Impurities from polypropylene microcentrifuge tubes as a potential source of interference in simultaneous analysis of multiple lipid-soluble antioxidants by HPLC with electrochemical detection
- Cerebrospinal fluid tau and Aβ42 concentrations in healthy subjects: delineation of reference intervals and their limitations
- The influence of smoking on plasma homocysteine and cysteine levels in passive and active smokers
- Reference change values and power functions
- A different approach to analyzing age-related HbA1c values in non-diabetic subjects
- Ratio of urinary free immunoglobulin light chain κ to λ in the diagnosis of Bence Jones proteinuria
- Influence of blood collection in plastic vs. glass evacuated serum-separator tubes on hormone and tumour marker levels
- Multicenter evaluation of a new immunoassay for procalcitonin measurement on the Kryptor® System
- Comparison of three commercial assays for the measurement of 17α-hydroxyprogesterone (17α-OHPR): limitations of the quality control system
- Analytical performance of the Albumin Cobalt Binding (ACB®) test on the Cobas MIRA® Plus analyzer
- Letter to the Editor
- Meetings and Awards
Articles in the same Issue
- How useful are haematology analyser flags?
- Intravenous immunoglobulin (IVIG): modes of action in the clinical management of recurrent pregnancy loss (RPL) and selected autoimmune disorders
- Automated flagging influences the inconsistency and bias of band cell and atypical lymphocyte morphological differentials
- Extracellular concentration of homocysteine in human cell lines is influenced by specific inhibitors of cyst(e)ine transport
- Generation of eight adjacent mutations in a single step using a site-directed mutagenesis kit
- Prohepcidin accumulates in renal insufficiency
- Impurities from polypropylene microcentrifuge tubes as a potential source of interference in simultaneous analysis of multiple lipid-soluble antioxidants by HPLC with electrochemical detection
- Cerebrospinal fluid tau and Aβ42 concentrations in healthy subjects: delineation of reference intervals and their limitations
- The influence of smoking on plasma homocysteine and cysteine levels in passive and active smokers
- Reference change values and power functions
- A different approach to analyzing age-related HbA1c values in non-diabetic subjects
- Ratio of urinary free immunoglobulin light chain κ to λ in the diagnosis of Bence Jones proteinuria
- Influence of blood collection in plastic vs. glass evacuated serum-separator tubes on hormone and tumour marker levels
- Multicenter evaluation of a new immunoassay for procalcitonin measurement on the Kryptor® System
- Comparison of three commercial assays for the measurement of 17α-hydroxyprogesterone (17α-OHPR): limitations of the quality control system
- Analytical performance of the Albumin Cobalt Binding (ACB®) test on the Cobas MIRA® Plus analyzer
- Letter to the Editor
- Meetings and Awards