A Two-Center Evaluation of the Blood Gas Immediate Response Mobile Analyzer (IRMA)

  • Ferruccio Ceriotti , Laura Del Buono , Pierangelo Bonini , Luca Germagnoli , Carlo A. Ferrero , Mario Pezzo and Alessandro Marocchi
Published/Copyright: June 1, 2005
Clinical Chemistry and Laboratory Medicine (CCLM)
From the journal Volume 40 Issue 2

Abstract

The Immediate Response Mobile Analyzer (IRMA) is a selective and portable point-of-care testing (POCT) blood gas, electrolyte and hematocrit (Hct) analyzer. The overall analytical performance was evaluated in a two-center study involving two Italian hospital laboratories, following the guidelines suggested by the manufacturer (based on the NCCLS protocol), after a preliminary evaluation of their formal validity. The IRMA was compared to the analyzers used in the routine laboratory as reference. The considered parameters were pH, pO2, pCO2, Na+, K+, ionized calcium and Hct. When using the aqueous quality control material provided by the manufacturer most of the parameters showed good precision, with the exception of pCO2 and pO2 that showed high CVs on two of the three levels of the aqueous control. We could demonstrate that this imprecision was material-related and was reduced when using a different material (blood equilibrated by tonometry). With tonometred blood for pO2 and pCO2 and the aqueous material for the remaining parameters the CVs were all below 5%, ranging from 0.08% to 2.8%. The IRMA results correlated adequately with the comparison instruments, with the exception of sodium and ionized calcium where contradictory results were obtained in the two centers.

:
Published Online: 2005-06-01
Published in Print: 2002-03-01

Copyright © 2002 by Walter de Gruyter GmbH & Co. KG

Articles in the same Issue

  1. Recommendations for the Routine Use of Pancreatic Amylase Measurement instead of Total Amylase for the Diagnosis and Monitoring of Pancreatic Pathology
  2. Bovine Spongiform Encephalopathy: Are the Cows Mad or Full of Carbohydrates?
  3. Plasma Total Antioxidant Capacity in Hemodialyzed Patients and Its Relationships to other Biomarkers of Oxidative Stress and Lipid Peroxidation
  4. Effects of Propofol and Taurine on Intracellular Free Amino Acid Profiles and Immune Function Markers in Neutrophils In Vitro
  5. Genotyping of Hemochromatosis-Associated Mutations in the HFE Gene by PCR-RFLP and a Novel Revers Hybridization Method
  6. Comparison of Different CD71 Monoclonal Antibodies for Enrichment of Fetal Cells from Maternal Blood
  7. The Effects of Marathon Swimming on Serum Leptin and Plasma Neuropeptide Y Levels
  8. Renal Function Exerts only a Minor Influence on High Plasma Homocysteine Concentrations in Patients with Acute Coronary Syndromes
  9. Changes of Biochemical Bone Markers during the Menopausal Transition
  10. Blood Sampling for Biochemical Analysis from Central Venous Catheters: Minimizing the Volume of Discarded Blood
  11. Experience with External Quality Assessment of Drugs of Abuse Testing in the Lombardy Region in Italy
  12. Reference Intervals for Thyroid Hormones on the Architect Analyser
  13. Intermethod Variation in Serum Carcinoembryonic Antigen (CEA) Measurement. Fresh Serum Pools and Control Materials Compared
  14. Turn-Around Time for Chemical and Endocrinology Analyzers Studied Using Simulation
  15. A Two-Center Evaluation of the Blood Gas Immediate Response Mobile Analyzer (IRMA)
  16. Painting, Poetry and Optics: Johannes Vermeer
  17. The Practice of Clinical Chemistry in the European Union
  18. Meetings and Awards
Downloaded on 30.12.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2002.031/html?lang=en
Scroll to top button