Effects of common hemoglobin variants on HbA1c measurements in China: results for α- and β-globin variants measured by six methods

  • Anping Xu , Weidong Chen , Yong Xia , Yu Zhou and Ling Ji EMAIL logo
Published/Copyright: April 7, 2018

Abstract

Background:

HbA1c is a widely used biomarker for diabetes mellitus management. Here, we evaluated the accuracy of six methods for determining HbA1c values in Chinese patients with common α- and β-globin chains variants in China.

Methods:

Blood samples from normal subjects and individuals exhibiting hemoglobin variants were analyzed for HbA1c, using Sebia Capillarys 2 Flex Piercing (C2FP), Bio-Rad Variant II Turbo 2.0, Tosoh HLC-723 G8 (ver. 5.24), Arkray ADAMS A1c HA-8180V fast mode, Cobas c501 and Trinity Ultra2 systems.

Results:

DNA sequencing revealed five common β-globin chain variants and three common α-globin chain variants. The most common variant was Hb E, followed by Hb New York, Hb J-Bangkok, Hb G-Coushatta, Hb Q-Thailand, Hb G-Honolulu, Hb Ube-2 and Hb G-Taipei. Variant II Turbo 2.0, Ultra2 and Cobas c501 showed good agreement with C2FP for most samples with variants. HLC-723 G8 yielded no HbA1c values for Hb J-Bangkok, Hb Q-Thailand and Hb G-Honolulu. Samples with Hb E, Hb G-Coushatta, Hb G-Taipei and Hb Ube-2 produced significant negative biases for HLC-723 G8. HA-8180V showed statistically significant differences for Hb E, Hb G-Coushatta, Hb G-Taipei, Hb Q-Thailand and Hb G-Honolulu. HA-8180V yielded no HbA1c values for Hb J-Bangkok. All methods showed good agreement for samples with Hb New York.

Conclusions:

Some common hemoglobin variants can interfere with HbA1c determination by the most popular methods in China.

Acknowledgments

The authors thank the members of the Clinical Medical Laboratory of GuangDong Dongguan Houjie town Hospital for the HbA1c assay with the HLC-723 G8 analyzer, the members of the Clinical Medical Laboratory of Yantian District People’s Hospital for the HbA1c assay with the Cobas c501 and HA-8180V analyzers, and the members of the Clinical Medical Laboratory of Shenzhen Chinese Medicine Hospital for the HbA1c assay with the Trinity Ultra2 analyzer.

  1. Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

  2. Research funding: None declared.

  3. Employment or leadership: None declared.

  4. Honorarium: None declared.

  5. Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

References

1. Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977–86.10.1056/NEJM199309303291401Search in Google Scholar

2. UK Prospective Diabetes Study (UKPDS) Group. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS33). Lancet 1998;352:837–53.10.1016/S0140-6736(98)07019-6Search in Google Scholar

3. Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parrott M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2002;48:436–72.10.1093/clinchem/48.3.436Search in Google Scholar

4. Nathan DM, Balkau B, Bonora E, Borch-Johnsen K, Buse JB, Colagiuri S, et al. International Expert Committee report on the role of the A1c assay in the diagnosis of diabetes. Diabetes Care 2009;32:1327–34.10.2337/dc09-1777Search in Google Scholar

5. Sacks DB, Arnold M, Bakris GL, Bruns DE, Horvath AR, Kirkman MS, et al. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Diabetes Care 2011;34:e61–99.10.2337/dc11-9998Search in Google Scholar PubMed PubMed Central

6. Weykamp CW. HbA1c: a review of analytical and clinical aspects. Ann Lab Med 2013;33:393–400.10.3343/alm.2013.33.6.393Search in Google Scholar PubMed PubMed Central

7. Little RR, Rohlfing CL, Sacks DB. National Glycohemoglobin Standardization Program (NGSP) Steering Committee. Status of hemoglobin A1c measurement and goals for improvement: from chaos to order for improving diabetes care. Clin Chem 2011;57:205–14.10.1373/clinchem.2010.148841Search in Google Scholar PubMed

8. International Federation of Clinical Chemistry and Laboratory Medicine, IFCC Scientific Division, Mosca A, Goodall I, Hoshino T, Jeppsson JO, et al. Global standardization of glycated hemoglobin measurement: the position of the IFCC Working Group. Clin Chem Lab Med 2007;45:1077–80.10.1515/CCLM.2007.246Search in Google Scholar PubMed

9. Lin CN, Emery TJ, Little RR, Hanson SE, Rohlfing CL, Jaisson S, et al. Effects of hemoglobin C, D, E, and S traits on measurements of HbA1c by six methods. Clin Chim Acta 2012;413:819–21.10.1016/j.cca.2011.12.019Search in Google Scholar PubMed PubMed Central

10. Rohlfing C, Hanson S, Weykamp C, Siebelder C, Higgins T, Molinaro R, et al. Effects of hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by twelve methods. Clin Chim Acta 2016;455:80–3.10.1016/j.cca.2016.01.031Search in Google Scholar PubMed PubMed Central

11. Chen W, Zhang X, Shang X, Cai R, Li L, Zhou T, et al. The molecular basis of beta-thalassemia intermedia in southern China: genotypic heterogeneity and phenotypic diversity. BMC Med Genet 2010;11:31.10.1186/1471-2350-11-31Search in Google Scholar PubMed PubMed Central

12. Jaisson S, Leroy N, Desroches C, Tonye-Libyh M, Guillard E, Gillery P. Interference of the most frequent haemoglobin variants on quantification of HbA1c: comparison between the LC–MS (IFCC reference method) and three routinely used methods. Diabetes Metab 2013;39:363–9.10.1016/j.diabet.2013.01.004Search in Google Scholar PubMed

13. Lee ST, Weykamp CW, Lee YW, Kim JW, Ki CS. Effects of 7 hemoglobin variants on the measurement of glycohemoglobin by 14 analytical methods. Clin Chem 2007;53:2202–5.10.1373/clinchem.2007.093963Search in Google Scholar PubMed

14. Little RR, Vesper H, Rohlfing CL, Ospina M, Safar-Pour S, Roberts WL. Validation by a mass spectrometric reference method of use of boronate affinity chromatography to measure glycohemoglobin in the presence of hemoglobin S and C traits. Clin Chem 2005;51:264–5.10.1373/clinchem.2004.043141Search in Google Scholar PubMed

15. Kabytaev K, Connolly S, Rohlfing CL, Sacks DB, Stoyanov AV, Little RR. Higher degree of glycation of hemoglobin S compared to hemoglobin A measured by mass spectrometry: potential impact on HbA1c testing. Clin Chim Acta 2016;458:40–3.10.1016/j.cca.2016.04.027Search in Google Scholar PubMed PubMed Central

16. Lin M, Han ZJ, Wang Q, Zheng L, Wang Y, Yang H, et al. Molecular epidemiological survey of hemoglobinopathies in the Wuxi region of Jiangsu Province, eastern China. Hemoglobin 2013;37:454–66.10.3109/03630269.2013.807285Search in Google Scholar PubMed

17. Ji L, Yu J, Zhou Y, Xia Y, Xu A, Li W, et al. Erroneous HbA1c measurements in the presence of β-thalassemia and common Chinese hemoglobin variants. Clin Chem Lab Med 2015;53: 1451–8.10.1515/cclm-2014-0598Search in Google Scholar PubMed

18. Little RR, Rohlfing CR, Hanson S, Connolly S, Higgins T, Weykamp C, et al. Effects of hemoglobin E and D traits on glycated hemoglobin (HbA1c) Measurements by twenty-three methods. Clin Chem 2008;54:1277–82.10.1373/clinchem.2008.103580Search in Google Scholar PubMed

19. Lenters-Westra E. Common Hb-variants show no longer interference on the Tosoh G8 after an update of the software. Clin Chim Acta 2016;463:73–4.10.1016/j.cca.2016.10.014Search in Google Scholar PubMed

20. Li YQ, Huang HP, Chen ZZ, Zhao L, Liang, L, Qin, GF, et al. Comparison of capillary electrophoresis and high performance liquid chromatography for detection and quantification of hemoglobin New York. Clin Chem Lab Med 2016;54:91–5.Search in Google Scholar

Received: 2017-12-27
Accepted: 2018-02-19
Published Online: 2018-04-07
Published in Print: 2018-07-26

©2018 Walter de Gruyter GmbH, Berlin/Boston

Articles in the same Issue

  1. Frontmatter
  2. Editorials
  3. Procalcitonin for diagnosing and monitoring bacterial infections: for or against?
  4. Is it time to abandon the Nobel Prize?
  5. In Memoriam
  6. Norbert Tietz, 13th November 1926–23rd May 2018
  7. Reviews
  8. Procalcitonin guidance in patients with lower respiratory tract infections: a systematic review and meta-analysis
  9. Telomere biology and age-related diseases
  10. Opinion Paper
  11. Procalcitonin-guided antibiotic therapy: an expert consensus
  12. Genetics and Molecular Diagnostics
  13. mRNA expression profile in peripheral blood mononuclear cells based on ADRB1 Ser49Gly and Arg389Gly polymorphisms in essential hypertension — a case-control pilot investigation in South Indian population
  14. General Clinical Chemistry and Laboratory Medicine
  15. External quality assessment schemes for glucose measurements in Germany: factors for successful participation, analytical performance and medical impact
  16. Long-term stability of glucose: glycolysis inhibitor vs. gel barrier tubes
  17. Observational studies on macroprolactin in a routine clinical laboratory
  18. Anti-streptavidin IgG antibody interference in anti-cyclic citrullinated peptide (CCP) IgG antibody assays is a rare but important cause of false-positive anti-CCP results
  19. Point-of-care creatinine testing for kidney function measurement prior to contrast-enhanced diagnostic imaging: evaluation of the performance of three systems for clinical utility
  20. Hematology and Coagulation
  21. Analytical performance of an automated volumetric flow cytometer for quantitation of T, B and natural killer lymphocytes
  22. Lupus anticoagulant testing using two parallel methods detects additional cases and predicts persistent positivity
  23. Reference Values and Biological Variations
  24. Within-subject biological variation of activated partial thromboplastin time, prothrombin time, fibrinogen, factor VIII and von Willebrand factor in pregnant women
  25. Within-subject and between-subject biological variation estimates of 21 hematological parameters in 30 healthy subjects
  26. Immunoglobulin G (IgG) and IgG subclass reference intervals in children, using Optilite® reagents
  27. Cancer Diagnostics
  28. Serum carbohydrate sulfotransferase 7 in lung cancer and non-malignant pulmonary inflammations
  29. Cardiovascular Diseases
  30. Clinical performance of a new point-of-care cardiac troponin I test
  31. Diabetes
  32. Testing for HbA1c, in addition to the oral glucose tolerance test, in screening for abnormal glucose regulation helps to reveal patients with early β-cell function impairment
  33. Effects of common hemoglobin variants on HbA1c measurements in China: results for α- and β-globin variants measured by six methods
  34. Infectious Diseases
  35. Serum procalcitonin concentration within 2 days postoperatively accurately predicts outcome after liver resection
  36. Predictive value of serum gelsolin and Gc globulin in sepsis – a pilot study
  37. Cerebrospinal fluid free light chains as diagnostic biomarker in neuroborreliosis
  38. Letter to the Editor
  39. Glucose and total protein: unacceptable interference on Jaffe creatinine assays in patients
  40. Interference of glucose and total protein on Jaffe-based creatinine methods: mind the covolume
  41. Interference of glucose and total protein on Jaffe based creatinine methods: mind the covolume – reply
  42. Measuring procalcitonin to overcome heterophilic-antibody-induced spurious hypercalcitoninemia
  43. Paraprotein interference in the diagnosis of hepatitis C infection
  44. Timeo apis mellifera and dona ferens: bee sting-induced Kounis syndrome
  45. Bacterial contamination of reusable venipuncture tourniquets in tertiary-care hospital
  46. Detection of EDTA-induced pseudo-leukopenia using automated hematology analyzer with VCS technology
  47. Early adjustment of antimicrobial therapy after PCR/electrospray ionization mass spectrometry-based pathogen detection in critically ill patients with suspected sepsis
  48. Thromboelastometry reveals similar hemostatic properties of purified fibrinogen and a mixture of purified cryoprecipitate protein components
  49. Bicarbonate interference on cobas 6000 c501 chloride ion-selective electrodes
Downloaded on 24.1.2026 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2017-1211/html
Scroll to top button