Home Determination of reference change values for thyroid-related biomarkers: TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO
Article Open Access

Determination of reference change values for thyroid-related biomarkers: TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO

  • Mehmet Akif Bildirici ORCID logo EMAIL logo , Yüksel Aliyazıcıoğlu ORCID logo , Hüseyin Yaman ORCID logo , Süleyman Caner Karahan ORCID logo and Asım Örem ORCID logo
Published/Copyright: April 4, 2025

Abstract

Objectives

Reference change value (RCV) is an objective tool that indicates whether the change between consecutive measurements is clinically significant. This study aimed to determine the instrument-specific RCVs of thyroid-stimulating hormone (TSH), free triiodothyronine (fT3), free thyroxine (fT4), thyroglobulin (Tg), thyroglobulin antibody (Anti-Tg), and thyroperoxidase antibody (Anti-TPO) tests, which are routinely analyzed for the diagnosis, case-finding, and screening of thyroid diseases.

Methods

This cross-sectional study was conducted between January 1 and December 31, 2018. The twelve-month mean analytical imprecision (CVA) was calculated from internal quality control data for each test on three different instruments. All tests’ individuality index (II) was calculated using the biological variation data. RCVs were determined by both conventional and logarithmic approaches.

Results

It was shown that the II of all tests was 0.6 and below. Upward logarithmic RCVs were slightly higher than conventional RCVs, while downward logarithmic RCVs were lower (e.g., conventional RCV for TSH 39 %, whereas upward and downward RCVs 48 and 32 %, respectively).

Conclusions

Due to high individuality, the results of thyroid-related biomarkers would be more appropriate to evaluate with RCV instead of reference interval. Instrument-specific RCVs should be determined because the analytical imprecision of analyzers may vary. Most presentations to the hospital are not healthy individuals, so logarithmic RCV could provide a more accurate assessment of significant change.

Introduction

Clinical laboratory test results are frequently evaluated with population-based reference intervals (expected values for healthy individuals) [1]. Individuality index (II) is a parameter that gives information about the suitability of evaluating test results with the reference interval and is determined by the ratio of within-subject biological variation (CVI) to between-subject biological variation (CVG). Most of the tests in the laboratory have CVI considerably lower than CVG, and II values lower than 0.6 mean the high individuality of the test. When evaluating the serial results of these tests, it is recommended to use the reference change value (RCV) instead of the reference interval, that is, to interpret it according to the previous result [2]. The usefulness of evaluating test results with reference intervals remains limited due to this high individuality [3]. Even if the serial measurement results of many individuals are within the reference interval, significant changes may be observed in the individual, or an abnormal measurement result may not indicate pathology. It is especially valid when the result is close to the critical threshold. Thus, interpreting the results with these traditional intervals can sometimes lead to the wrong assessment. Variations in some pre-analytical, analytical, and post-analytical processes, and most importantly, biological variation (BV) cause this misinterpretation [4].

The RCV, which covers the main natural sources of variation – CVI and analytical imprecision (CVA) – is an objective tool that shows whether the change between consecutive measurements is clinically significant. A clinically significant change has occurred if the difference between the individual’s serial results is greater than the RCV [5]. The RCV was formulated by Fraser as follows: 21/2×Z×(CVA 2+CVI 2)1/2 [3]. Z–score shows the number of the standard deviation appropriate to the desired confidence interval. When evaluating the test results, an increase or a decrease occurs compared to the previous result. Therefore, one-tailed (unidirectional) Z-scores should be used as there will be no concomitant increase and decrease in clinical decision management (e.g., 1.65 for p<0.05, 2.33 for p<0.01) [6], 7].

In the study of Fokkema et al., it was found that serial results in an individual were not normally distributed. Therefore, different from the traditional Fraser method, an approach involving logarithmic transformation was developed for RCV calculation [8]. Lund et al., in their study with a computer simulation model, made this logarithmic approach more practical for tests where the total variation (CVT) (CVT=(CVA 2+CVI 2)1/2) is below 30 % [9]. The CVI is a random fluctuation that occurs in the individual’s homeostasis process, normally (Gaussian) distributed. However, disease, abnormal metabolic cases, and medication cause abnormal fluctuations of biological variation. In such cases, the relevant analytes should not be expected to show a normal distribution [9], and the RCVs should be determined using the logarithmic transformation method [10], 11].

Thyroid dysfunction is one of the most common clinical presentations in the community and may cause many health problems, including mood changes, hair loss, fatigue, and obesity [12]. To make a clinical decision about a patient’s thyroid functions, it is necessary to evaluate the complaints, physical examination findings, and thyroid function tests together. Thyroid-stimulating hormone (TSH), free triiodothyronine (fT3), and free tetraiodothyronine/thyroxine (fT4) levels are most commonly used in the diagnosis of diseases that cause thyroid dysfunctions [13], 14]. In addition to these thyroid function tests, thyroperoxidase antibodies (Anti-TPO) and thyroglobulin antibodies (Anti-Tg) are helpful in the investigation of autoimmune thyroid diseases. Besides, thyroglobulin (Tg), a tumor marker, is widely used with the anti-Tg test in the follow-up of patients with differentiated thyroid carcinoma [14], [15], [16], [17].

Evaluating the results of thyroid-related biomarkers is essential since thyroid dysfunction and the thyroid diseases that cause it are common [12], [13], [14]. Our hospital is a tertiary care center and serves a large population in the Eastern Black Sea Region, with its location as a regional hospital. Most of those who apply to our hospital are individuals with the disease, medication, or abnormal metabolic conditions. Therefore, their thyroid biomarkers may not show a normal distribution [9]. Our clinical laboratory has three different systems for analyzing these tests. Since analytical imprecision can vary, it may be essential to determine RCVs on a device-specific basis. Consequently, this study aimed to determine instrument-specific RCVs of TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO tests, which are routinely examined in our laboratory for the diagnosis, treatment, and follow-up of thyroid diseases, using classical and logarithmic methods.

Materials and methods

Study design

This cross-sectional study was conducted in Farabi Hospital Clinical Biochemistry Laboratory (Karadeniz Technical University Faculty of Medicine) between January 1 and December 31, 2018. It was approved by the Scientific Research Ethics Committee of Karadeniz Technical University Faculty of Medicine in accordance with the Declaration of Helsinki (protocol dated 05.13.2019 and numbered 2019–157).

Materials and methods

Thyroid-related biomarkers (TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO) were analyzed on two different Beckman Coulter Unicel DXI 800 (Beckman Coulter Inc., Minnesota, USA) immunochemistry auto-analyzers (Routine 1: R1 and Routine 2: R2) in the routine unit. Furthermore, thyroid function tests (TSH, fT3, and fT4) analyzed in the Roche Cobas e411 (Roche Diagnostics GmbH, Mito, Japan) immunochemistry auto-analyzer (E) in the emergency unit were also included in the study. Manufacturers’ original brand reagents (Beckman Coulter Inc. and Roche Diagnostics GmbH) were used in the study, and thyroid-related biomarkers were analyzed with chemiluminescence and electrochemiluminescence technologies on Unicel DXI 800 and Cobas e411 model instruments, respectively.

The tests’ 2018 internal quality control (IQC) data were obtained from the laboratory information management system, and the CVA values were calculated for each test. IQC samples were analyzed for all tests at two levels (clinically normal and pathological, or levels close to decision limits for Anti-Tg and Anti-TPO) per day. For Unicel DXI 800 instruments, lyophilized Sero Autonorm Immunoassay (Sero As, Billingstad, Norway) brand control materials with lot numbers 1,604,200 and 1,604,201 were used in the first three months of the year, and lot numbers 1,608,805 and 1,609,806 were used in the remaining nine months. For the Cobas e411 analyzer, liquid PreciControl Universal (Roche Diagnostics GmbH, Mannheim, Germany) brand control materials with lot numbers 147,130 and 147,140 were used in the first three months of the year, 249,617 and 249,618 in April-September, and lot numbers 270,758 and 270,761 in the remaining three months. IQC results were evaluated according to Westgard rules and accepted or rejected [18]. Standard deviation (SD) and mean were calculated monthly for the two-level controls of each test. The CVs of the two control levels were determined with the SD/mean×100 formula. The monthly CVA values of the tests were obtained from the average CV of the two levels. Then, the mean CVA for the 12 months of 2018 was calculated for each test.

The biological variation data (CVI and CVG) of the tests were obtained from The European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) BV database [19]. The BV data of Anti-Tg and Anti-TPO tests, which were not in this database, were obtained from the BV study of Ricós et al. [20]. The II was calculated using the CVI/CVG formula for each parameter [1].

Statistical analysis

Instrument-specific RCVs of TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO tests were determined by conventional (Fraser) and logarithmic approaches using CVI and CVA values. The RCV % values were determined using the classical approach 21/2×Z×(CVA 2+CVI 2)1/2 formula [21]. The Lund approach was considered more appropriate if the CVT values ​​calculated with the formula (CVA 2+CVI 2)1/2 were below 30 %. Therefore, the reference change factor limits (RCFup and RCFdown) suggested by Lund et al. were determined (RCFup=exp (Z×21/2×CVT/100) and RCFdown=1/RCFup) [9]. These factor limits were converted to RCVs to make these results comparable. For this, the distance of the RCF from the value of one was calculated proportionally, and the upward/downward RCVs were determined as a percentage. For example, the RCFup value for TSH was 1.48, and the RCFdown value was 0.68. These values, calculated as factor limits, were converted to percentages. For upward change, the RCV value was 48 % with the identity ‘(1.48–1)/1’; for downward change, it was found to be 32 % with the identity ‘(1–0.68)/1’. The Z constant in the formulas was used as 1.65 one-sided for the 95 % confidence interval (p<0.05) [6]. All analyses were performed using the Microsoft Excel 2013 program (Microsoft Corporation, Washington, USA).

Results

The average CVA values ​​of each test were determined on an instrument-specific basis because the analytical performances of the analyzers differed. While the analytical imprecision of the tests was within the limits recommended by the manufacturers, the CVA values ​​of fT3, fT4 (for R1 and R2), Anti-Tg, and Anti-TPO tests were higher than the desirable CV. The number of quality-control measurements run for all three instruments, outliers (number of samples rejected according to Westgard rules [18]), the average CV of the two levels, the twelve-month average CVA values, ​​and the desirable CV data are presented in Table 1. All the II values calculated using BV data of TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO parameters were 0.6 and below. The CVT values were found in the range of 6.9–16.8 %. Lund et al.’s approach was considered appropriate for the logarithmic RCV calculation as all CVT values were below 30 %. The biological variation (CVI and CVG), II, CVA, and CVT data of thyroid parameters are presented in Table 2. The instrument-specific RCVs calculated with the conventional and logarithmic approaches using these data are given in Table 3. Since Tg, Anti-Tg, and Anti-TPO parameters were not analyzed on the Roche Cobas e411 instrument, their RCVs were not presented. The upward logarithmic RCVs were higher than the conventional approach, while logarithmic RCVs in the downward change were lower than the conventional RCVs.

Table 1:

The number of quality-control measurements, outliers, analytical imprecision, and desirable CV data of thyroid parameters.

Test Analyzera Measurem, n Outlierb n (%) CVLevel 1, % CVLevel 2, % CVA, % CVDesirable, %
TSH R1 555 25 (5) 3.9 4 4 8.2
R2 536 0 (0) 3.5 3.8 3.7
E 972 201 (21) 3.1 3.1 3.1
fT3 R1 614 67 (11) 6.6 5.9 6.3 3
R2 586 48 (8) 6.4 5 5.8
E 931 100 (11) 3.3 3.4 3.4
fT4 R1 642 115 (18) 5.5 5.4 5.5 3.9
R2 586 50 (9) 4.7 4.2 4.5
E 969 183 (19) 3.1 3.1 3.1
Tg R1 548 53 (10) 4.3 4.3 4.3 6.4
R2 526 23 (4) 6.4 6.2 6.3
Anti-Tg R1 682 225 (33) 6.7 5.8 6.4 4.3c
R2 574 81 (14) 6.7 9 7.9
Anti-TPO R1 641 110 (17) 6.7 7.4 7 5.7c
R2 625 108 (17) 6.7 6.8 6.7
  1. Measurem, measurement; CV, coefficient of variation; CVA, analytical imprecision; TSH, thyroid stimulating hormone; fT3, free triiodothyronine; fT4, free thyroxine; Tg, thyroglobulin; Anti-Tg, thyroglobulin antibody; Anti-TPO, thyroperoxidase antibody. aR1 and R2 analyzers – two different Beckman Coulter Unicel DXI 800 auto-analyzers in the routine unit, E analyzer – Roche Cobas e411 auto-analyzer in the emergency unit. bOutlier is the number of samples rejected according to Westgard rules [18]. cDesirable CV, data for the Anti-Tg and Anti-TPO parameters was from the study of Ricós et al., and others from the EFLM BV database [19], 20].

Table 2:

Biological variation, individuality index, analytical imprecision, and total variation data of thyroid parameters.

Test CVI CVG II CVA CVT
R1 R2 E R1 R2 E
TSH 16.3 29.2 0.6 4 3.7 3.1 16.8 16.7 16.6
fT3 6 16.5 0.4 6.3 5.8 3.4 8.7 8.4 6.9
fT4 7.7 12 0.6 5.5 4.5 3.1 9.4 8.9 8.3
Tg 12.8 29.2 0.4 4.3 6.3 13.5 14.3
Anti-Tga 8.5 82 0.1 6.4 7.9 10.7 11.6
Anti-TPOa 11.3 147 0.1 7 6.7 13.3 13.2
  1. CVI, within-subject biological variation; CVG, between-subject biological variation; II, individuality index; CVA, analytical imprecision; CVT, total variation; TSH, thyroid stimulating hormone; fT3, free triiodothyronine; fT4, free thyroxine; Tg, thyroglobulin; Anti-Tg, thyroglobulin antibody; Anti-TPO, thyroperoxidase antibody. R1 and R2 – two different Beckman Coulter Unicel DXI 800 auto-analyzers in the routine unit, E – Roche Cobas e411 auto-analyzer in the emergency unit. aBV data (CVI/CVG) for the Anti-Tg and Anti-TPO parameters was from the study of Ricós et al., and others from the EFLM BV database [19], 20].

Table 3:

Reference change values calculated with the conventional and logarithmic approaches.

Test Analyzera Conventional RCV, % Logarithmic RCV (+), % Logarithmic RCV (−), %
TSH R1 39 48 32
R2 39 48 32
E 39 47 32
fT3 R1 20 22 18
R2 20 22 18
E 16 17 15
fT4 R1 22 25 20
R2 21 23 19
E 19 21 18
Tg R1 32 37 27
R2 33 39 28
Anti-Tg R1 25 28 22
R2 27 31 24
Anti-TPO R1 31 36 27
R2 31 36 26
  1. RCV, reference change value; TSH, thyroid stimulating hormone; fT3, free triiodothyronine; fT4, free thyroxine; Tg, thyroglobulin; Anti-Tg, thyroglobulin antibody; Anti-TPO, thyroperoxidase antibody. aAnalyzers: R1 and R2 – two different Beckman Coulter Unicel DXI 800 auto-analyzers in the routine unit, E – Roche Cobas e411 auto-analyzer in the emergency unit.

Discussion

In this study, the RCVs of thyroid-related biomarkers (TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO) were determined by both traditional and logarithmic approaches based on the analytical performance of three different immunochemistry auto-analyzers in our laboratory. Instrument-specific RCVs were calculated since the analytical performances of the instruments differ (Table 1, 3). RCVs were determined for all three instruments’ TSH, fT3, and fT4 tests. However, since Tg, Anti-Tg, and Anti-TPO biomarkers were studied in the routine unit, RCVs were determined on R1 and R2 instruments. It was observed that the II of TSH and fT4, of which RCV was calculated, was 0.6, and the other tests were <0.6 (Table 2). Therefore, since the individuality of the tests was quite high, it was found that it would be more appropriate to evaluate the results of thyroid-related biomarkers with RCV instead of the population-based reference interval [2]. Eventually, if the change between consecutive results for thyroid-related biomarkers is greater than the RCV, it may be a clinically significant change [5]. An illustrative example: in a routine outpatient clinic control of an adult stable patient (Hashimoto thyroiditis) on levothyroxine treatment, serum TSH is measured as 1.6 mIU/L (the previous visit was 1.0 mIU/L). The increase is 0.6 mIU/L (60 %), and this result, when evaluated with RCV, shows a significant change even though the result is within the reference interval (classical and logarithmic RCVs of TSH: 39 and 48 %, respectively). This assessment may indicate that the drug dose needs to be increased.

This study showed that the upward logarithmic RCVs were higher than the conventional approach; the logarithmic RCVs in the downward change were lower than the conventional RCVs (e.g., conventional RCV for TSH 39 %, upward and downward RCVs 48 and 32 %, respectively) (Table 3). The fact that the logarithmic RCV is higher in the upward change compared to the conventional approach can be explained by the Fraser approach’s higher false positive rates, as demonstrated in Lund et al.’s study. In their study with a computer simulation model, the conventional approach’s false positive rate (caused by the confidence interval) increased even more at values where CVT exceeded 10 %. In contrast, the rate did not change in the logarithmic approach [22]. In our study, the CVT values of TSH, Tg, Anti-Tg, and Anti-TPO tests were greater than 10 %. Therefore, using logarithmic RCV for these tests may provide a more accurate assessment. In addition, the EFLM BV database and the study by Ricós et al. provide data from studies on healthy individuals [19], 20]. The validity of these data for monitoring disease has been questioned, but in general, it has been shown that CVI values are indeed similar in health and chronic stable disease. Therefore, the current available BV data are ubiquitously applicable [6]. However, disease, medication, and abnormal metabolic status cause abnormal fluctuations of BV [10], 11]. Therefore, logarithmic RCVs may be more beneficial for evaluating thyroid-related biomarkers of individuals with these conditions.

When instrument-specific RCVs were compared, the RCVs of the TSH test were quite similar, while the RCVs of the fT3 and fT4 parameters in the E instrument were slightly lower than the R1 and R2 instruments (e.g., conventional RCVs of fT4 for R1, R2, and E are 22 , 21, 19 %, respectively) (Table 3). This difference was due to the better analytical performance of fT3 and fT4 hormones in the E analyzer. The RCV results were also quite similar, as the CVA values of the parameters in the R1 and R2 auto-analyzers were close to each other (Table 2, 3). Since the CVI and Z values in the RCV formula are independent of the auto-analyzer or the methodology used [23], CVA is the only variable that affects the calculation when making inter-analyzer RCV comparisons for the same test. Therefore, similar RCVs are determined for auto-analyzers with similar analytical performance. However, since the analytical imprecision of each system is different, it is important to calculate device-specific RCV.

In the study by Erden and colleagues, in which they determined the BV and RCV values ​​of thyroid function tests, RCVs were determined bidirectionally with the classical approach within a 95 % confidence interval. They determined the RCVs as 104 % for TSH, 63.1 % for fT3, and 38.8 % for fT4 [24]. These results were remarkably higher than the RCVs calculated in our study. This difference occurred because the CVI values ​​determined by Erden and colleagues (37.2 % for TSH, 22.3 % for fT3, and 13.3 % for fT4) were considerably higher than the data in the EFLM database, and they used a two-sided Z-score.

Cinpolat et al. calculated the RCVs of 24 different immunoassay parameters unidirectionally with the classical approach within a 95 % confidence interval. They determined RCVs as 45.2 % for TSH, 19.4 % for fT3, 14.6 % for fT4, 24.6 % for Anti-Tg, and 35.8 % for Anti-TPO [25]. Similar RCV results were obtained for the Anti-Tg and Anti-TPO tests, in which we used the same BV data (Ricós et al. study) in our study. There was a difference in TSH and fT4 tests. The different BV databases used and our possible analytical imprecision differences have caused these results.

Ağca, in his thesis on measurement uncertainty and RCV in biochemistry and hormone tests, RCVs were calculated bidirectionally with the traditional method within a 95 % confidence interval. His results were 53.6 % for TSH and 17 % for fT4 [26]. The difference in the results was thought to be due to the differences in the Z-scores and BV databases we used.

In their study, Walz and Fierz determined the RCVs of many analytes, and they calculated 99 % for the increase and −50 % for the decrease (one-sided with the logarithmic approach) for the TSH test [27]. It was determined that these values, which were considerably higher than our results, were caused by the different databases we used (EFLM/Ricós).

Bottani et al. estimated the BV data of thyroid biomarkers in the European BV study (EuBIVAS). They unidirectionally determined the RCVs of TSH, fT3, fT4, and Tg tests with the logarithmic approach within a 95 % confidence interval. RCVs were calculated as 50.7 , 13.1, 12.6, 29.0 % in the upward change and −33.6 %, −11.6 %, −11.2 %, −22.5 % in the downward change, respectively [28]. Our calculated RCVs for TSH were similar, but our results were higher for other parameters. Our study’s higher CVA and CVI values for parameters other than TSH caused these results.

In the meta-analysis study by Fernández-Calle et al., in which the BVs of thyroid-related measurements were estimated, RCVs of TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO parameters were also determined. This study determined RCVs unidirectionally with the logarithmic approach within a 95 % confidence interval. RCVs were calculated as 55.1 , 14.9, 16.0, 32.2, 33.3, 43.2 % on the increase and −35.5 %, −13.0 %, −13.8 %, −24.4 %, −25.0 %, −30.2 % on the decrease, respectively [29]. While the RCVs calculated for TSH, Anti-TPO, and Anti-Tg tests were higher than our study, it was observed to be lower for other parameters. These differences were due to different BV data and analytical imprecision.

In the literature, different RCVs have been observed in studies on BV and RCV of thyroid-related markers [24], [25], [26], [27], [28], [29]. The analytical imprecision, BV data, and bi- or unidirectional Z scores resulted in these differences. Even if the BV data used in the studies are taken from the up-to-date database, and the Z values are used the same, the calculated RCV results will be different because the analytical performance of each laboratory will be different. In short, each laboratory needs to determine and use its own RCVs for the analyzed parameters. In addition, reporting the RCV together with the reference interval in the clinical laboratory report is recommended [30], 31]. In this way, biological and analytical variation can be considered and RCV can be beneficial as an objective tool in evaluating patient results [5]. However, reporting RCV results as a percentage exposes clinicians to mathematical calculations and can make it difficult for clinicians to interpret the results. Therefore, RCV can be reported using warning flags for test results that show significant changes [23], 31].

This study has some limitations. First, the lot information of the reagents used throughout 2018 was not recorded. Second, there was no data on the preparation of lyophilized IQC. The other was that since the BV data was obtained from databases simultaneously with the study, it may not overlap with current data, especially the EFLM database. Moreover, the BV data used may only partially reflect the variations in our population. Therefore, further studies may need to be conducted in which preanalytical standardization is provided, and RCVs are calculated by determining BV data. Another critical issue is the use of instrument-specific RCV, which requires that patient results be analyzed on the same instrument. However, patient samples can be analyzed on any instrument in laboratories like ours with multiple instruments. Therefore, in practice, a single RCV can be determined by calculating a pooled CVA that includes all analyzers [32]. Lastly, the potential disadvantages of RCV are that not every change exceeding the RCV may be clinically significant: a) some biological variations may be affected by health status, b) diurnal variation, and c) differences in the timing of sample collection [6].

Evaluating the results of thyroid-related biomarkers is very important since thyroid dysfunction and the thyroid diseases that cause it are quite common in the community [12], [13], [14]. The II values of all tests in the study were 0.6 and below, indicating a high individuality. Therefore, TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO test results should be evaluated using the RCV instead of the population-based reference interval [2]. With the use of RCV in the clinical laboratory, analytical and biological variations will be considered, and it will benefit clinicians as an objective tool in assessing thyroid biomarker results and in clinical decision management of thyroid dysfunction and diseases. Thus, it will accelerate clinical decisions in the treatment processes of thyroid diseases, shorten the discharge times of inpatients, and reduce unnecessary test repetitions, which will be very useful in terms of workload, time, and cost. In addition, a more accurate evaluation will be provided by using RCV to follow up on adverse effects [23], 25], 33], 34]. Significantly, instrument-specific RCVs should be determined because the analytical imprecision of analyzers may vary. Moreover, since most of those who apply to the hospital, like us, have a disease, medication, or abnormal metabolic cases, it may be more accurate to evaluate thyroid test results with logarithmic RCVs. In conclusion, RCVs should be presented with the reference interval when reporting thyroid-related test results in clinical laboratories.


Corresponding author: Mehmet Akif Bildirici, Department of Medical Biochemistry, Faculty of Medicine, Kastamonu University, Kuzeykent Neighbourhood, 57.Alay Boulevard, No:29, TR37200, Kastamonu, Türkiye, E-mail:

Acknowledgments

Many thanks to Enver S., Ph.D., from Vocational School Of Technical Sciences, Muş Alparslan University, Muş, Turkey, for helping to improve the manuscript’s readability.

  1. Research ethics: This cross-sectional study was approved by the Scientific Research Ethics Committee of Karadeniz Technical University Faculty of Medicine in accordance with the Declaration of Helsinki (protocol dated 05.13.2019 and numbered 2019–157).

  2. Informed consent: Not applicable.

  3. Author contributions: The authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  4. Use of Large Language Models, AI and Machine Learning Tools: None declared.

  5. Conflict of interest: The authors state no conflict of interest.

  6. Research funding: None declared.

  7. Data availability: The raw data can be obtained on request from the corresponding author.

References

1. Fraser, CG. Inherent biological variation and reference values. Clin Chem Lab Med 2004;42:758–64. https://doi.org/10.1515/cclm.2004.128.Search in Google Scholar PubMed

2. Fraser, CG, Harris, EK. Generation and application of data on biological variation in clinical Chemistry. Crit Rev Clin Lab Sci 1989;27:409–37. https://doi.org/10.3109/10408368909106595.Search in Google Scholar PubMed

3. Fraser, CG. Reference change values: the way forward in monitoring. Ann Clin Biochem 2009;46:264–5. https://doi.org/10.1258/acb.2009.009006.Search in Google Scholar PubMed

4. Regis, M, Postma, TA, van den Heuvel, ER. A note on the calculation of reference change values for two consecutive normally distributed laboratory results. Chemometr Intell Lab Syst 2017;171:102–11. https://doi.org/10.1016/j.chemolab.2017.10.008.Search in Google Scholar

5. Harris, EK, Yasaka, T. On the calculation of a “reference change’’ for comparing two consecutive measurements. Clin Chem 1983;29:25–30. https://doi.org/10.1093/clinchem/29.1.25.Search in Google Scholar

6. Fraser, CG. Reference change values. Clin Chem Lab Med 2012;50:807–12. https://doi.org/10.1515/cclm.2011.733.Search in Google Scholar PubMed

7. Cooper, G, DeJonge, N, Ehrmeyer, S, Yundt-Pacheco, J, Jansen, R, Ricós, C, et al.. Collective opinion paper on findings of the 2010 convocation of experts on laboratory quality. Clin Chem Lab Med 2011;49:793–802. https://doi.org/10.1515/cclm.2011.149.Search in Google Scholar PubMed

8. Fokkema, MR, Herrmann, Z, Muskiet, FAJ, Moecks, J. Reference change values for brain natriuretic peptides revisited. Clin Chem 2006;52:1602–3. https://doi.org/10.1373/clinchem.2006.069369.Search in Google Scholar PubMed

9. Lund, F, Petersen, PH, Fraser, CG, Sölétormos, G. Calculation of limits for significant unidirectional changes in two or more serial results of a biomarker based on a computer simulation model. Ann Clin Biochem 2015;52:237–44. https://doi.org/10.1177/0004563214534636.Search in Google Scholar PubMed

10. Røraas, T, Støve, B, Petersen, PH, Sandberg, S. Biological variation: the effect of different distributions on estimated within-person variation and reference change values. Clin Chem 2016;62:725–36. https://doi.org/10.1373/clinchem.2015.252296.Search in Google Scholar PubMed

11. Ricós, C, Álvarez, V, Minchinela, J, Fernández-Calle, P, Perich, C, Boned, B, et al.. Biologic variation approach to daily laboratory. Clin Lab Med 2017;37:47–56. https://doi.org/10.1016/j.cll.2016.09.005.Search in Google Scholar PubMed

12. Vanderpump, MPJ. Epidemiology of thyroid disorders. In: Luster, M, Duntas, LH, Wartofsky, L, editors. The thyroid and its diseases: a comprehensive guide for the clinician. Cham: Springer International Publishing; 2019:75–85 pp.10.1007/978-3-319-72102-6_6Search in Google Scholar

13. Surks, MI, Chopra, IJ, Mariash, CN, Nicoloff, JT, Solomon, DH. American Thyroid Association guidelines for use of laboratory tests in thyroid disorders. JAMA 1990;263:1529. https://doi.org/10.1001/jama.1990.03440110095035.Search in Google Scholar

14. Soh, SB, Aw, TC. Laboratory testing in thyroid conditions – pitfalls and clinical utility. Ann Lab Med 2019;39:3–14. https://doi.org/10.3343/alm.2019.39.1.3.Search in Google Scholar PubMed PubMed Central

15. Vanderpump, MP, Tunbridge, WM, French, JM, Appleton, D, Bates, D, Clark, F, et al.. The incidence of thyroid disorders in the community; a twenty-year follow up of the Whickham survey. Clin Endocrinol 1995;43:55–68. https://doi.org/10.1111/j.1365-2265.1995.tb01894.x.Search in Google Scholar PubMed

16. Society of Endocrinology and Metabolism of Turkish. Guideline of diagnosis and treatment of thyroid diseases. Ankara: Türkiye Clinics; 2020.Search in Google Scholar

17. Algeciras-Schimnich, A. Thyroglobulin measurement in the management of patients with differentiated thyroid cancer. Crit Rev Clin Lab Sci 2018;55:205–18. https://doi.org/10.1080/10408363.2018.1450830.Search in Google Scholar PubMed

18. Westgard, JO, Groth, T. Design and evaluation of statistical control procedures: applications of a computer “quality control simulator” program. Clin Chem 1981;27:1536–45. https://doi.org/10.1093/clinchem/27.9.1536.Search in Google Scholar

19. Aarsand, AK, Fernandez-Calle, P, Webster, C, Coskun, A, Gonzales-Lao, E, Diaz-Garzón, J, et al.. EFLM biological variation database; 2019. [cited 2019 June 11]. Available from: https://biologicalvariation.eu/meta_calculations.Search in Google Scholar

20. Ricós, C, García-Lario, JV, Alvares, V, Cava, F, Domenech, M, Hernández, A, et al.. Biological variation database, and quality specifications for imprecision, bias and total error (desirable and minimum). The 2014 update. [cited 2019 June 10]. Available from: https://www.westgard.com/biodatabase1.html.Search in Google Scholar

21. Fraser, CG, Hyltoft Petersen, P, Lytken Larsen, M. Setting analytical goals for random analytical error in specific clinical monitoring situations. Clin Chem 1990;36:1625–8.10.1093/clinchem/36.9.1625Search in Google Scholar

22. Lund, F, Petersen, PH, Fraser, CG, Sölétormos, G. Calculation of limits for significant bidirectional changes in two or more serial results of a biomarker based on a computer simulation model. Ann Clin Biochem 2015;52:434–40. https://doi.org/10.1177/0004563214555163.Search in Google Scholar PubMed

23. Fraser, CG. Biological variation: from principles to practice, 1st ed. Washington DC: AACC Press; 2001.Search in Google Scholar

24. Erden, G, Barazi, AÖ, Tezcan, G, Yıldırımkaya, M. Biological variation and reference change values of TSH, free T3, and free T4 levels in serum of healthy Turkish individuals. Turk J Med Sci 2008;38:153–8.Search in Google Scholar

25. Cinpolat, HY, Bugdayci, G, Oguzman, H, Yis, MO. Calculation of reference change values of immun analysis parameters. Gazi Med J 2015;26:85–7.Search in Google Scholar

26. Ağca, E. Determination of measurement uncertainty and reference change values of biochemistry and hormone tests. [thesis for specialist]. Mersin: Mersin University, Faculty of Medicine, Department of Medical Biochemistry; 2016.Search in Google Scholar

27. Walz, B, Fierz, W. Der Referenzbereich ist tot–es lebe der Reference Change Value. Ther Umsch 2015;72:130–5. https://doi.org/10.1024/0040-5930/a000655.Search in Google Scholar PubMed

28. Bottani, M, Aarsand, AK, Banfi, G, Locatelli, M, Coşkun, A, Díaz-Garzón, J, et al.. European Biological Variation Study (EuBIVAS): within- and between-subject biological variation estimates for serum thyroid biomarkers based on weekly samplings from 91 healthy participants. Clin Chem Lab Med 2021;60:523–32. https://doi.org/10.1515/cclm-2020-1885.Search in Google Scholar PubMed

29. Fernández-Calle, P, Díaz-Garzón, J, Bartlett, W, Sandberg, S, Braga, F, Beatriz, B, et al.. EFLM working group on biological variation and task group for the biological variation database. Biological variation estimates of thyroid related measurands - meta-analysis of BIVAC compliant studies. Clin Chem Lab Med 2021;60:483–93.10.1515/cclm-2021-0904Search in Google Scholar PubMed

30. Biosca, C, Ricós, C, Lauzurica, R, Galimany, R, Petersen, PH. Reference change value concept combining two delta values to predict crises in renal posttransplantation. Clin Chem 2001;47:2146–8. https://doi.org/10.1093/clinchem/47.12.2146.Search in Google Scholar

31. Ricós, C, Perich, C, Minchinela, J, Álvarez, V, Simón, M, Biosca, C, et al.. Application of biological variation – a review. Biochem Med (Zagreb) 2009;19:250–9. https://doi.org/10.11613/bm.2009.023.Search in Google Scholar

32. Paal, M, Habler, K, Vogeser, M. Estimation of inter-laboratory reference change values from external quality assessment data. Biochem Med (Zagreb). 2021;31:030902. https://doi.org/10.11613/bm.2021.030902.Search in Google Scholar PubMed PubMed Central

33. Bugdayci, G, Polat, M, Oguzman, H, Cinpolat, HY. Interpretation of biochemical tests using the reference change value in monitoring adverse effects of oral isotretinoin in 102 ethnic Turkish patients. Lab Med 2016;47:213–9. https://doi.org/10.1093/labmed/lmw024.Search in Google Scholar PubMed PubMed Central

34. Zorbozan, N, Akarken, İ. Unnecessary repeated total PSA tests; evaluation with minimum retest interval and reference change value. J Turkish Clin Biochem 2019;17:42–7.Search in Google Scholar

Received: 2024-08-15
Accepted: 2024-12-06
Published Online: 2025-04-04

© 2025 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Articles in the same Issue

  1. Frontmatter
  2. Review
  3. Interaction of fetuin-A with obesity related insulin resistance and diabetes mellitus
  4. Research Articles
  5. Preanalytical errors in pediatric blood sampling: a systematic review of common challenges and risks
  6. The promotive role of reticulocalbin 3 (RCN3) in the pathogenesis of keloid via TGFβ1/Smad2/Smad7 signaling pathway in vitro
  7. Investigation into drug resistance to cisplatin in cancer stem cell-enriched population in non-small cell lung cancer
  8. The cytotoxic and antiproliferative effect of Polygala saponin XLIV on the human colorectal carcinoma cell line
  9. Development of HEK293T cell reference materials for β-thalassemia genetic testing using prime editing
  10. Investigation of propofol, fentanyl, and midazolam-related toxicity and the protective effect of midazolam on THLE-2 cell lines
  11. Leucine-rich α-2-glycoprotein 1 can be a novel angiogenic mediator in autosomal dominant polycystic kidney disease
  12. Asiaticoside reverses the inhibition effect of miR-184 on proliferation, migration and AKT phosphorylation of HTR-8/Svneo cells
  13. Perioperative D-dimer levels and head and neck cancer surgery: a prospective observational study
  14. Comprehensive analysis of KRT15 in pan-cancer and verification in lung adenocarcinoma
  15. Age and sex-dependent effects of nitrosative stress on the osmotic fragility of human red blood cells
  16. Evaluation of oxidative stress parameters in older patients with urinary incontinence
  17. Determination of reference change values for thyroid-related biomarkers: TSH, fT3, fT4, Tg, Anti-Tg, and Anti-TPO
  18. An evaluation of serum boron level in pregnancies with severe pre-eclampsia
  19. Involvement of SIRT3/FOXO1 and TLR4/NF-κB/NLRP3 pathways in protective effects of Δ(9)-tetrahydrocannabinol on diabetic cardiomyopathy in rats
  20. Evaluating the potential therapeutic effect of Rosa damascena Mill. essential oil on acetic acid-induced ulcerative colitis in rats
  21. Assessment of blood collection competency in phlebotomists: a survey study
Downloaded on 10.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/tjb-2024-0200/html
Scroll to top button