Home A case of falsely elevated D-dimer result
Article Open Access

A case of falsely elevated D-dimer result

  • Tülay Çevlik ORCID logo EMAIL logo , Rana Turkal ORCID logo , Goncagül Haklar ORCID logo and Önder Şirikçi ORCID logo
Published/Copyright: August 24, 2022

Abstract

Objectives

Heterophile antibodies can cause interferences in immunometric assays. While many tests are shown to be affected by interference from heterophilic antibodies, the D-dimer test has rarely been reported to be affected. With this case, we report an elevated D-dimer measurement which was not compatible with the clinical presentation.

Case presentation

A 41-year-old patient who was admitted to hospital with heart palpitations had a D-dimer elevation irrelevant to his clinical condition. D-dimer measurements were repeated in new samples directly and after being treated in heterophilic blocking tube with two different reagent lots of a latex-based automated immunoturbidimetric assay and an immunoturbidometric assay. D-dimer values were normalized (0–0.5 mg/L) when we used a new lot of reagent on the same instrument or measured by an immunoturbidometric method on the chemistry analyzer. After treatment with HBT, all samples revealed D-dimer results within the reference ranges.

Conclusions

The presence of heterophile antibodies in a sample should be considered when an elevated D-dimer value that is not compatible with the clinical presentation is encountered. Apart from the patient’s HA’s causing false results, sporadic susceptibility of the reagents should also be kept in mind as a possibility.

Introduction

D-dimer is a product of fibrin degradation which forms during fibrinolysis. It is most used for the diagnosis of deep vein thrombosis (DVT), pulmonary embolism (PE), dissection of the aorta, sepsis, infection, and disseminated intravascular coagulation (DIC). D-Dimer can also be measured in low amounts in healthy individuals [1]. Whole blood, plasma, and serum samples can be used for analysis. At least 30 commercial D-dimer assays based on three other methods are available to determine D-dimer qualitatively and quantitatively. These methods are enzyme-linked immunosorbent assays (ELISA), immunofluorescent assays and latex-based automated assays with immunoturbidometric readings. The common point of these methods is the use of monoclonal antibodies against epitopes on D-dimer fragments. Each of these methods has its specific considerations and limitations [2].

Paraproteinemia, icterus, lipemia, and hemolysis are common factors interfering with coagulation tests at the preanalytical phase. The interference will also depend on the analytical method used [3].

Heterophile antibodies (HA) are a source of interference in immunoassays. HAs interact weakly or strongly with the antibody present in the reagent or antigen-antibody complex formed during the immunoassay analysis, which may cause false-positive or false-negative results. HAs interact weakly or strongly with the antibody or antigen-antibody complex at the end of the reaction and cause false-positive or false-negative results. Many reports are available regarding the effect of HA on different immunometric assays but reports of interference with D-dimer assays are rare [4].

In this case report, we investigated the reason behind an elevated D-dimer result which was not compatible with the clinical presentation of the patient who was admitted to our hospital with heart palpitations.

Case presentation

A 41-year-old male patient was admitted to Marmara University Pendik E&R Hospital with a complaint of heart palpitations. The patient’s history revealed sleep disturbances, and he had stated to be passing through a stressful period. The patient had no medical history of illness, except for high triglycerides and cholesterol levels, and the patient was on a lipid-lowering diet. There was no history of any drug use. His physical examination and routine laboratory tests were normal except for an elevated triglyceride level and an unexpected D-dimer elevation (1.21 mg/L; reference range 0–0.5 mg/L) which had persisted on a repeated sample on the same day (1.13 mg/L). His ECG, Holter monitoring records and echocardiography were normal. COVID-19 PCR (Polymerase Chain Reaction) and antibodies against SARS-CoV-2 were all negative. CT (Computed Tomography) angiography was normal. Heart palpitations were attributed to the patient’s stressful period. His physician consulted the laboratory for the elevated D-dimer levels. An elevated D-dimer result in a 41 years old patient without compatible clinical presentation on admission and status on follow-up has led us to search for other causes than thrombosis.

New samples were drawn to determine D-dimer levels directly and after being treated in a heterophilic blocking tube (HBT, Scantibodies Laboratories, Santee, CA) with two different reagent lots of a latex-based automated immunoturbidimetric assay (STA-R, Diagnostica Stago, USA) and an immunoturbidometric assay (Olympus AU400 Chemistry Analyzer, Beckman Coulter, USA). Additionally, triglyceride and anti-rheumatoid factor (RF) tests were ordered reflectively.

For HA interference screening, 500 µL of patient’s plasma was added to the HBT that contains a pellet of blocking reagent. The tube was then incubated for 1 h at room temperature. The untreated and treated plasma samples were assayed with two different lots of reagents on the hemostasis analyzer STA-R and Olympus AU400 Chemistry Analyzer, simultaneously.

Results

The preanalytical and analytical conditions of previous samples were requestioned for any factor that could account for the elevated D-dimer results, and all operations were found to be in accordance with the procedures. The internal and external quality control results to pertaining to days of elevated results were reviewed and were found to be within acceptable limits. Except for triglycerides, all test results were within the normal reference range (Table 1). His RF was negative, and his triglyceride level was 334 mg/dL (reference range <200 mg/dL). The samples’ hemolysis, icterus, and lipemia indices were within normal limits.

Table 1:

Laboratory test results.

Patient results Reference range
At admission (30.12.2020) Follow-up (30.12.2020)
Leukocytes, ×10³/µL 8.3 ND 4.0–10
Red blood cell, ×106/µL 5.6 ND 3.5–5.7
Hemoglobin, g/dL 16.8 ND 12–17
Hematocrit, % 48.9 ND 36–50
Platelets, ×10³/µL 231 ND 150–440
D-dimer, mg/L 1.21 1.13 0–0.5
Troponin T-hs, ng/L 4 ND 0–14
Fibrinojen, mg/dL 306 ND 200–400
Glucose, mg/dL 95 ND 65–110
Triglycerides, mg/dL 334 30–200
Bilirubin, mg/dL 0.31 ND 0–1.13
Creatinine, mg/dL 0.73 ND 0–1.12
ALT, U/L 26 ND 10–40
AST, U/L 21 ND 10–37
COVID-19 PCR Negative
Rheumatoid factor, IU/mL ND 13 0–14
  1. ND, not determined.

The new sample still revealed an elevated D-dimer result with the current reagent lot used in the STA-R hemostasis analyzer. However, D-dimer results were within reference ranges when measured with a new lot of reagents on the same instrument and with the immunoturbidometric method on the chemistry analyzer. After treatment with HBT, all samples’ D-dimer results were within reference ranges (Table 2).

Table 2:

D-dimera results obtained from the new sample before and after HBT treatment.

Analyzer Untreated sample After HBT treatment
Hemostasis analyzer STA-R (new sample, current lot) 0.94 mg/L 0.34 mg/L
Hemostasis analyzer STA-R (new sample, new lot) 0.30 mg/L 0.13 mg/L
Chemistry analyzer olympus AU400 0.22 mg/L 0.31 mg/L
  1. aReference range, 0–0.5 mg/L.

Discussion

D-Dimer is a fibrin degradation product, and elevated levels could be an alarming sign indicating thrombus formation, a potentially fatal situation, which can also be encountered in COVID-19 patients. Elevated D-dimer concentrations may also be observed in many conditions unrelated to thrombosis, including preanalytical and analytical interactions, and differentiation of these conditions is a challenge [3]. In our case, the lack of compatible clinical presentation with a high D-dimer result upon admission, lack of any indicator of thrombosis in cardiovascular tests and radiological investigations, a negative COVID-19 PCR, lack of clinical signs of thrombosis during follow-up led us to non-thrombotic reasons, mainly to preanalytical and analytical reasons interfering with the test result.

D-Dimer measurements might be affected by hemolysis, icterus, and lipemia, but the repeating samples’ hemolysis, icterus, and lipemia indices were all within acceptable limits. The patient’s triglyceride concentration was elevated (334 mg/dL), but it was still below the threshold of lipemia interference of 670 mg/dL on Stago hemostasis analyzers as stated by the manufacturer [5]. Elevated RF is another molecule that can interfere with D-dimer assays, but the patients’ RF was found to be negative.

Another increasingly recognized cause of interference in immunoassays is the presence of HA, which are multi-specific immunoglobulins. The formation of HA is triggered by known or unknown antigenic stimulants encountered as a result of exposure to animals and/or products containing antigenic molecules of animal source, after immunization by vaccines, following blood transfusions, autoimmune diseases, and maternal metastases [4]. Depending on the design of the assay, interference due to HA may lead to falsely low or high analyte levels in several immunoassays. When suspected, the presence of HA can be verified by following the methods, checking for linearity of results in serially diluted samples, re-analyzing the sample by a different method or by using heterophilic antibody blocking reagents/tubes [6]. Wu et al. [4] and Sun et al. [7] have also reported heterophilic antibody interference with very high D-dimer results. In both cases, the patients were over 70 years of age with underlying diseases (including hypertension, diabetes, bronchial asthma and chronic heart failure). In both cases, the heterophilic antibody interference was shown by treating the samples with heterophilic antibody blockers; the D-dimer levels had decreased after adding heterophilic antibody blockers but were still above the reference ranges. Whereas, our patient was a young and healthy individual, who had a slightly elevated result above the reference range but well within the measuring range. After treating with heterophilic antibody blockers, the D-dimer level had returned to within the reference range interval. Getting normalized D-dimer results with HBT treated samples led us to conclude that there was an HA interference in our patients’ samples, causing a false elevation.

The prevalence of HA in the general population can be as high as 40% [8]. False results due to HA interference has the potential to have medical and legal consequences. Therefore, IVD companies try to decrease the frequency of interferences by adding HA blocking agents to reagents [9]. These agents are usually nonimmune (irrelevant) globulins or normal animal sera in nature. This practice has dramatically decreased the frequency of interference observed in current assays to 0.05% (10 per 20,000) [10]. In addition to the demonstrated HA in patient’s plasma, the fact that normal results had been obtained both with another lot of reagent on the same hemostasis analyzer and with an immunoturbidometric assay on the chemistry analyzer has shown us that the current reagent lot used on the hemostasis analyzer was more susceptible to interference with HA, leading to the elevated D-dimer results.

Conclusions

Although D-dimer is a marker for fibrinolysis and the test is run on hemostasis analyzers in many laboratories, it should be remembered that the assays are immunoassays in principle. The presence of HA in a sample should be considered when an elevated D-dimer value that is not compatible with the clinical presentation is encountered. Apart from the patient’s HA’s causing false results, sporadic susceptibility of the reagents should also be kept in mind as a possibility.

Points to remember:

  1. It should be noted that D-dimer measurements are also antibody-based methods.

  2. Keep in mind the possibility of interference by heterophile antibodies.

  3. Maintain open communication with clinicians so they can freely reflect on inappropriate test results.


Corresponding author: Tülay Çevlik, Biohemistry Laboratory, Marmara University Pendik Education and Research Hospital, Istanbul, Turkey, E-mail:

  1. Research funding: This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

  2. Author contributions: ÖŞ and GH conceived the study. TÇ and RT abstracted the data and participated in the data interpretation. RT and TÇ performed data interpretation, statistical analysis and reported the results. TÇ wrote the original draft of the paper. ÖŞ and GH proof-read and corrected the initial manuscript. All authors have read and approved the final version.

  3. Conflict of interest: The authors declare no conflicts of interest regarding this article.

  4. Informed consent: Not applicable.

  5. Ethics declarations: Ethics approval and consent to participate. All methods were carried out in accordance with relevant guidelines and regulations.

  6. Data availability: The dataset analyzed during the current study are available from the corresponding author on reasonable request.

References

1. Weitz, JI, Fredenburgh, JC, Eikelboom, JW. A test in context: D-dimer. J Am Coll Cardiol 2017;70:2411–20. https://doi.org/10.1016/j.jacc.2017.09.024.Search in Google Scholar

2. Johnson, ED, Schell, JC, Rodgers, GM. The D-dimer assay. Am J Hematol 2019;94:833–9.10.1002/ajh.25482Search in Google Scholar

3. Favressea, J, Lippib, G, Roy, PM, Chatelaina, B, Jacqmina, H, Cate, H, et al.. Preanalytical, analytical, post-analytical variables, and clinical applications. Crit Rev Clin Lab Sci 2018;55:548–77. https://doi.org/10.1080/10408363.2018.1529734.Search in Google Scholar

4. Wu, Y, Xiao, YX, Huang, TY, Zhang, XY, Zhou, HB, Zhang, XX, et al.. What makes D-dimer assays suspicious–heterophilic antibodies? J Clin Lab Anal 2019;33:22687. https://doi.org/10.1002/jcla.22687.Search in Google Scholar

5. Ramzi, M, Esteve, F, Grimaux, M. D-dimer measurements and Lipemia Interference of STA-Liatest D-Di on diagnostica stago analyzers. A new approach for interference testing and lipid removal by high-speed centrifugation. Diagnostica Stago; 2018: 7 p. Report No. 301478.Search in Google Scholar

6. Çokluk, E, Tuncer, FB, Şekeroğlu, MR, Gözükara, C, Ağaç, S, Gözükara, Sİ. Effect of heterophyll antibody interference on CA 19-9 measurement: a case report. Sakarya Tip Dergisi 2019;9:558–64. https://doi.org/10.31832/smj.587512.Search in Google Scholar

7. Sun, HX, Ge, H, Xu, ZQ, Sheng, HM. Clinical laboratory investigation of a patient with an extremely high D-dimer level: a case report. World J Clin Cases 2020;8:3560–6. https://doi.org/10.12998/wjcc.v8.i16.3560.Search in Google Scholar

8. Bolstad, N, Warren, DJ, Nusta, K. Heterophilic antibody interference in immunometric assays. Best Pract Res Clin Endocrinol Metab 2013;27:647–61.10.1016/j.beem.2013.05.011Search in Google Scholar

9. STA®-Liatest® D-Di Plus, less interferences for even more confidence. Available from: https://www.stago.com/products-services/new-products/detail/article.Search in Google Scholar

10. Levinson, SS, Miller, JJJ. Towards a better understanding of heterophile (and the like) antibody interference with modern immunoassays. Clin Chim Acta 2002;325:1–15. https://doi.org/10.1016/s0009-8981(02)00275-9.Search in Google Scholar

Received: 2021-12-07
Accepted: 2022-06-09
Published Online: 2022-08-24

© 2022 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. COP27 climate change conference: urgent action needed for Africa and the world
  4. Review Article
  5. Clinical laboratory use of liquid chromatography mass spectrometry
  6. Research Articles
  7. Verification of enzymatic ethanol analysis method and method comparison with headspace gas chromatography
  8. Investigation of GHRL (rs4684677), FTO (rs8044769) and PGC1Α (rs8192678) polymorphisms in type 2 diabetic Turkish population
  9. Roles of OLR1 and IL17A variants on clinical phenotypes of Turkish patients undergoing coronary artery bypass surgery
  10. Cytokine gene polymorphism frequencies in Turkish population living in Marmara region
  11. Evaluation of BRCA1/2 gene mutations in patients with high-risk breast and/or ovarian cancer in Turkey
  12. Impacts of long noncoding RNA MALAT1 on LPS-induced periodontitis via modulating miR-155/SIRT1 axis
  13. Proteomics characterization of the adenovirus VA1 non-coding RNA on the landscape of cellular proteome
  14. Modulatory effect of pomegranate extract on TRPA1, TRPM2 and caspase-3 expressions in colorectal cancer induction of mice
  15. EpCAM is critical for tumor proliferation and oxaliplatin chemoresistance in EpCAMhigh/CD44+ colorectal cancer stem cells
  16. Prognostic factors in thrombotic thrombocytopenic purpura
  17. Possible effects of clinoptilolite on small intestinal ischemia-reperfusion injury caused by experimental mesenteric artery occlusion
  18. The relationship between pregnancies complicated with fetal growth restriction and umbilical cord blood endocan concentrations
  19. Sildenafil for the treatment of necrotizing enterocolitis: an experimental study
  20. The importance of LDH/Albumin, LDH/Lymphocyte, and LDH/Platelet ratios in the evaluation of COVID-19 B.1.1.7 variant
  21. The relationship between serum vitamin D and antibody response following two doses of inactivated COVID-19 vaccine
  22. The relationship between plasminogen activator inhibitor-1 levels and the course of disease in COVID-19 patients
  23. Could tear be an alternative specimen for SARS-CoV-2 detection?
  24. Case Report
  25. A case of falsely elevated D-dimer result
Downloaded on 30.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/tjb-2021-0262/html?lang=en&srsltid=AfmBOoqJrjHxyoszLMbUNyyYqbeprMzitRQjPD1EoE4_O-aQLg2M0BVs
Scroll to top button