Home Analysis of a 6-year pilot external quality assurance survey of free light chain using Sigma metrics
Article Open Access

Analysis of a 6-year pilot external quality assurance survey of free light chain using Sigma metrics

  • Hyojin Chae , Jaeeun Yoo , Joonhong Park , Kyoungho Cha , Jeong Joong Lee , Bongrae Cho , Myungshin Kim EMAIL logo and Yonggoo Kim EMAIL logo
Published/Copyright: September 14, 2019
Become an author with De Gruyter Brill

Abstract

Background

A pilot external quality assurance (EQA) survey for the free light chain (FLC) assay was developed and implemented in Korea.

Methods

Survey data over 6 years (2010–2015) were collected retrospectively and Sigma metrics were calculated for method-specific peer groups.

Results

Nineteen to 29 laboratories participated in the EQA survey, and nephelometric (20%) and turbidimetric (80%) methods were used. Using a previously published clinically relevant reference change value (RCV) of 54.5% as the tolerance limit, the method-specific median Sigma metrics of kappa (κ) and lambda (λ) FLC achieved greater than Three-Sigma for 86–97% of all EQA distributions, and Five-Sigma for 48–72% of all distributions.

Conclusions

This EQA analysis of FLC assay applied clinically relevant quality specifications using Sigma metrics. During the 6-year EQA survey, we found that most of the results from participating laboratories meet clinically relevant quality specifications. In addition, method-specific differences were noted for λ FLC, at FLC concentrations above the initial measuring range that require a sample dilution.

Introduction

Free light chain (FLC) assays quantitatively measure light chains that are not bound to immunoglobulin heavy chains, and detect clonal restriction through a skewed ratio of free kappa (κ) to free lambda (λ) light chains [1]. The assay uses polyclonal antibodies that bind exclusively to the hidden epitopes of FLC molecules located at the junction of the immunoglobulin heavy and light chains in an intact immunoglobulin. These epitopes are only accessible when the light chain molecules are not bound to the heavy chain, which makes the assay highly specific. The use of latex enhancement further enhances the sensitivity of the assay to a few mg/L, and the assay can be performed by turbidimetry or nephelometry on a number of automated laboratory instruments.

The FLC assay has become standard practice and is recommended in various guidelines for diagnosis, prognostic assessment and evaluation of patient response to treatment in multiple myeloma and related plasma cell disorders [2], [3], [4], [5], [6]. However, technical concerns with the FLC assay include non-linearity and antigen excess, FLC polymerization, possibility of undetected “private” epitopes, lack of international standards, lot-to-lot variability of the reagent and inter-instrument variability.

In this regard, an external quality assurance (EQA) survey for the quantitative FLC assay was developed and implemented in collaboration with the Quality Assurance Committee of the Korean Society for Laboratory Medicine. We analyzed the 6-year EQA survey data and applied Sigma metrics to evaluate the observed analytical characteristics of the FLC assay and to assess the quality of the FLC assay that is being achieved.

Materials and methods

Preparation of EQA materials and design of the EQA survey

The EQA material was derived from surplus serum from patients with monoclonal gammopathies and from healthy controls, and a serum pool was prepared as described in the Clinical and Laboratory Standards Institute document C37-A [7]. FLC concentrations were titrated to examine medical decision points, linearity, antigen excess and lower limit of quantitation. Serum material manufactured for quality assurance (QA) measures (Binding Site, Birmingham, UK) and commercial quality control (QC) materials purchased from Randox (Randox, Antrim, UK) were also used.

Three levels of EQA specimens were distributed bi-annually to the participating laboratories. The specimens were shipped frozen with ice pack and were transported by registered mail. Instructions regarding the correct processing of the specimens upon receipt, care needed for reconstruction, storage, and deadlines for sample analysis and reporting were provided with each dispatch of EQA materials. For each sample, the participating laboratory reported the ᴋ and λ FLC results and the ᴋ/λ ratios. The participating laboratories were given a unique laboratory code known only to the EQA coordinator, staff and the laboratory, for anonymization and confidentiality.

Two types of reports were provided for each level of EQA material. One was a summary report with the mean, standard deviation (SD) and standard deviation index (SDI) of all participant results. SDI expresses the reported result as the number of SDs it is from the mean value (SDI=[result–mean]/SD).

The other report was specific to each method peer group. Due to the lack of an international standard material or a reference measurement procedure for FLC, the FLC EQA was designed as a consensus program with major performance goals (criteria) based on the laboratory consensus. For each type of EQA report, the mean of all participants’ results and the mean of each method peer group results were regarded as the target values, respectively. Those results distributed more than 3 SD away from the mean were regarded as outliers and removed from statistical analyses [8].

Application of Sigma metrics and statistical analysis of FLC EQA survey data

Survey data from 32 EQA samples distributed in 11 EQA surveys over 6 years (from 2010 to 2015) were collected retrospectively. An Excel spreadsheet was used to tabulate the observed means, SDs and medians for the method subgroups to calculate observed bias, coefficient of variation (CV), total error (TE) and Sigma metrics, and to plot %bias and %CV for each subgroup for each specimen in each episode on the Sigma Proficiency Assessment Chart [9]. Sigma metrics were calculated as follows [9]:

Sigma metricpeer group=(TEa|%bias|)/CVpeer groupTE=Bias+1.65 SD or CV

Due to the small number of laboratories comprising specific instrument peer groups, and for a simplified analysis, only method-specific peer groups were defined. The method-specific bias was estimated as the difference from the weighted average. The weighted average was calculated to account for the number of laboratories in each method subgroup [9]. Due to the lack of a well-established consensus definition of quality requirements regarding the FLC assay, multiple surrogate definitions of quality requirements were assessed (Table 1). The Sigma Proficiency Assessment Chart was developed using a previously published reference change value (RCV) of 54.5% [12] derived from the estimate of monoclonal FLC variability in patients with plasma cell disorders as the tolerance [13] in calculating the Sigma metrics for κ FLC and λ FLC. Of note, the RCV value of 54.5% was chosen because it was comparable to the reported interlaboratory reproducibility values achieved during the UK National External Quality Assessment Service (UKNEQAS) FLC EQA scheme (CV of FLC assays was >50%, and often over 100%) [14].

Table 1:

Multiple surrogate definitions of quality requirements found in the literature.

Surrogate definitions of quality requirements
Desirable biological [10]Minimal [11]Desirable [11]Optimal [11]RCV [12]RCV (modular P) [1]RCV (BNII) [1]
κ FLC8.016.110.75.454.57769
λ FLC8.619.312.96.454.5121106
  1. FLC, free light chain; RCV, reference change value.

Continuous data are presented as mean (SD) for normally distributed data and median (range) for non-normally distributed data. Categorical data are presented as numbers (%). Normality was assessed using the D’Agostino and Pearson normality tests. Continuous variables were compared by the Mann-Whitney U-test and/or one-way analysis of variance for non-normally distributed data and categorical variables were compared by chi-squared (χ2) tests. MedCalc version 14.12.0 (MedCalc Software, Mariakerke, Belgium) was used for statistical analyses. A p-value<0.05 was considered statistically significant.

This study was conducted in accordance with the ethical guidelines of the Declaration of Helsinki and was approved by the Institutional Review Board (IRB)/Ethics Committee of Seoul St. Mary’s Hospital (IRB No. KC10SISI0037).

Results

Nineteen to 29 laboratories participated in the EQA survey. All participating laboratories used the FLC Freelite® kits manufactured by Binding Site. Nephelometric (20%) and turbidimetric (80%) methods were used for the FLC assay, and different instruments including Siemens BNII (Siemens Healthineers, Deerfield, IL, USA), Siemens BN-Prospec (Siemens Healthineers), Roche Modular P (Roche, Mannheim, Germany), Hitachi analyzers (Hitachi, Tokyo, Japan), Toshiba (Toshiba, Tokyo, Japan), Siemens Advia 1650 (Siemens Healthineers) and Binding Site SPAPLUS (Binding Site) were used (Table 2).

Table 2:

Distribution of methods and instruments used for the FLC assay during the 6-year EQA survey.

EQA event
Method/manufacturerS10-02S11-01S11-02S12-01S12-02S13-01S13-02S14-01S14-02S15-01S15-02
Nephelometry87655435433
 Siemens BNII76544435433
 Siemens BN-Prospec11111000000
Turbidimetric1212191921202122232425
 Roche Modular/Hitachi911161717171817181719
 Toshiba11324333343
 Siemens Advia11000000000
 Binding Site SPAplus00000012233
All methods2019252426242427272728
  1. EQA, external quality assurance; FLC, free light chain.

Method-specific total error and Sigma metrics

The method-specific median TE values of κ FLC across 29 EQA specimens were 26% for nephelometry and 25% for turbidimetry. The method-specific median TE values of λ FLC across 29 EQA specimens were 24% and 19% for nephelometry and turbidimetry, respectively.

As the Sigma metric is inversely proportional to CVpeer group and EQA distributions with low FLC concentrations (less than the lower limit of the reference interval) are associated with a disproportionately large CVpeer group, three κ FLC and three λ FLC EQA distributions with low FLC concentrations were not included in the Sigma metrics analysis. The method-specific median Sigma metrics of κ FLC across 29 EQA specimens were 4.9 for nephelometric methods and 4.9 for turbidimetric methods. The method-specific median Sigma metrics of λ FLC across 29 EQA specimens were 4.9 for nephelometric methods and 5.4 for turbidimetric methods. There were no statistically significant differences in TE and Sigma metrics between turbidimetric methods and nephelometric methods for κ FLC; however, for λ FLC, nephelometric methods showed larger TE and lower Sigma metrics compared with turbidimetric methods (p<0.05) (Figure 1).

Figure 1: Sigma Proficiency Assessment Chart for κ FLC (A) and λ FLC (B) with TEa=54.5%.Diagonal lines from top to bottom represent 2σ and 3σ. The x-coordinate represents the interlaboratory precision observed for the method-specific peer group. The y-coordinate represents the bias compared to the laboratory consensus value used as the major performance goal of the EQA survey.
Figure 1:

Sigma Proficiency Assessment Chart for κ FLC (A) and λ FLC (B) with TEa=54.5%.

Diagonal lines from top to bottom represent 2σ and 3σ. The x-coordinate represents the interlaboratory precision observed for the method-specific peer group. The y-coordinate represents the bias compared to the laboratory consensus value used as the major performance goal of the EQA survey.

Method harmonization

We compared the median reported results of the nephelometric and turbidimetric methods for each EQA specimen. Due to the small number of laboratories reporting nephelometric results, the Bland-Altman plot was used. The Bland-Altman plot showed that the average of the difference between the nephelometric and turbidimetric methods for κ FLC was 7.2 mg/L and for λ FLC was 42.1 mg/L (Figure 2). While the average difference between the two methods for κ FLC can be regarded as insignificant, for λ FLC there seems to be a significant difference between the two methods, and the difference seems to be due to higher λ FLC concentrations that would require a different sample dilution from that of the initial measuring range.

Figure 2: Bland-Altman plot of median reported results of nephelometry and turbidimetry for each EQA specimen.Shown are absolute difference plots for κ FLC (A) and λ FLC (B).
Figure 2:

Bland-Altman plot of median reported results of nephelometry and turbidimetry for each EQA specimen.

Shown are absolute difference plots for κ FLC (A) and λ FLC (B).

Interlaboratory reproducibility

The method-specific median interlaboratory reproducibility (CV) of κ FLC across 32 EQA specimens was 12% for both nephelometric and turbidimetric methods. The method-specific median interlaboratory reproducibility of λ FLC across 32 EQA specimens was 12% and 10% for the nephelometric and turbidimetric methods, respectively (Table 3).

Table 3:

Statistical summary of the κ and λ FLC EQA results for all specimens distributed.

EQA event
MethodE10-02E11-01E11-02E12-01E12-02E13-01E13-02E14-01E14-02E15-01E15-02
S10-02-1S10-02-2S10-02-3S11-01-1S11-01-2S11-02-1S11-02-2S11-02-3S12-01-1S12-01-2S12-01-3S12-02-1S12-02-2S13-01-1S13-01-2S13-02-1S13-02-2S13-02-3S14-01-1S14-01-2S14-01-3S14-02-2S14-02-3S15-01-1S15-01-2S15-01-3S15-02-1S15-02-2S15-02-3
Nephelometry
 Mean12.213.61.813.033.118.664.124.027.015.02793.3231.113.6647.318.02673.315.312.712.02925.212.414.9231.411.9559.78.4380.715.52.6
 %bias11565146–11–116–8–2115–36–3–5–106–11–6–10–39–7–3–5–21
 CV (%)59131315812177104101518153114411911828231912.57.73.2
 TEa1920272639193129132214182545285826122921291913164738241826
 Sigma10.96.04.24.23.66.84.53.27.85.413.65.43.63.03.61.83.913.64.96.04.96.827.26.82.42.94.47.117.0
Turbidimetric
 Mean10.312.51.812.026.016.678.624.326.613.63234.8241.113.4503.119.42347.816.414.114.72733.215.316.7283.812.7466.49.6405.817.44.0
 %bias–6–36–3–11–5900–472–1–104–74510–1106103–973621
 CV (%)16121282361757101017812131377101591112816915824
 TEa332326174815379122123301430262815172626252430163622281961
 Sigma3.44.54.56.82.49.13.210.97.85.45.43.26.84.54.24.27.87.85.43.66.04.94.56.83.46.03.66.82.3
All methods
 Weighted mean11.012.91.712.429.117.571.924.226.714.23031.0235.613.5560.918.62512.815.813.413.42765.513.915.8257.312.3513.29.0393.316.43.3
 CV (%)15111210238188710111791614177812141211148181114927
 Number of outliers00200010001020001100000000100
MethodE10-02E11-01E11-02E12-01E12-02E13-01E13-02E14-01E14-02E15-01E15-02
S10-02-1S10-02-2S10-02-3S11-01-1S11-01-2S11-02-1S11-02-2S12-01-1S12-01-2S12-02-1S12-02-2S12-02-3S13-01-1S13-01-2S13-01-3S13-02-1S13-02-2S13-02-3S14-01-1S14-01-3S14-02-1S14-02-2S14-02-3S15-01-1S15-01-2S15-01-3S15-02-1S15-02-2S15-02-3
Nephelometry
 Mean13.941.73160.413.739.917.114.713.755.414.314.01775.118.3247.74940.41.915.5357.113.3269.04297.214.813.413.825.81048.911.712.64028.4
 %bias–313–4–3–4–5–1–1–31–5–1–105–5–20–635–5–4–796–9–145
 CV (%)22152313111010101011182119181777121191311118713111013
 TEa3926412521202218172130403239331713202417272322212028273127
 Sigma2.53.62.44.25.05.45.45.45.45.03.02.62.93.03.27.87.84.54.96.14.25.05.06.87.84.24.95.44.2
Turbidimetric
 Mean14.744.22874.814.942.618.916.614.255.715.113.91932.518.7313.34726.72.016.0353.714.9256.03873.816.214.916.021.7924.313.916.73626.7
 %bias37–7546720303214001–16–2–5577–8–6914–5
 CV (%)96771110108611514191181511312101016626221512
 TEa18171816222324151021826333213251962519223117111810123825
 Sigma6.19.17.87.85.05.45.46.89.15.010.93.92.94.96.83.65.018.24.55.45.43.49.127.29.127.227.23.64.5
All methods
 Weighted mean14.341.33077.014.241.117.815.513.955.714.713.91872.218.4274.34718.62.015.8355.714.1262.44081.615.513.914.923.7986.712.814.73831.2
 CV (%)1912201211101110911162011191797111291312119913121513
 Number of outliers10000100000100120000000000100
  1. CV, coefficient of variation; EQA, external quality assurance; FLC, free light chain; TEa, total allowable error.

The κ FLC and λ FLC values in healthy individuals occur at the lower end of the measuring range, within the manufacturer’s reference intervals of 3–19 mg/L for κ FLC and 6–26 mg/L for λ FLC. FLC concentrations that border on the upper reference limit of the reference interval (medical decision point) could potentially be misclassified as normal or abnormal, and discordant results could occur by simply repeating the assay on those samples. Regarding interlaboratory reproducibility at values closest to the medical decision point, four EQA specimens had weighted average values near the κ or λ medical decision points. The all-method interlaboratory reproducibility values (CV) were 8%, 14% and 9% at κ FLC concentrations of 17.5, 18.6 and 16.4 mg/L, respectively, and 9% at a λ FLC concentration of 23.7 mg/L.

Discussion

The FLC assay is extensively used for the diagnosis, prognostic assessment and management of patients with monoclonal gammopathy. The assay can be used on several types of automated analyzers and has been adopted for use in many laboratories nationwide in Korea as an indispensable resource. However, as with any clinical laboratory test, EQA (also known as proficiency testing) is considered a mandatory component of a laboratory quality system [15]. National EQA schemes for the FLC assay have been developed and implemented in France, United Kingdom and Czech [16], and a number of international schemes also exist. The lack of a national EQA scheme in Korea prompted the development and implementation of an EQA survey for the FLC assay in collaboration with the Quality Assurance Committee of the Korean Society for Laboratory Medicine. There have been more than 30 distributions over a 6-year period, and over 35 laboratories nationwide participated in the scheme.

In this retrospective analysis of the EQA scheme, we applied clinically relevant quality specifications using Sigma metrics. Sigma metrics combine bias, imprecision and the allowable TE and convert them into an overall assessment of the analytical quality of the test [15]. For that purpose, the previously described Sigma Proficiency Assessment Chart was used [9].

However, due to the lack of a well-established consensus definition of quality requirement regarding the FLC assay, multiple surrogate definitions of quality requirements, including those based on biological variations [12], [17], [18] and/or measurement uncertainties [1], were assessed. We chose a reported RCV value of 54.5% derived from the estimate of monoclonal FLC variability in patients with plasma cell disorders [12] and applied as tolerance limits in calculating the Sigma metrics for κ FLC and λ FLC. We acknowledge the limitations of using an RCV value as a quality specification as it does not have component for bias and therefore may not be an appropriate substitute for an allowable TE. Moreover, an RCV value of 54.5%, which is derived from the estimate of monoclonal FLC variability in patients with plasma cell disorders, is much larger than reported biological variation of FLC in healthy individuals. However, FLC is clinically a tumor marker and typically in myeloma the monoclonal FLC levels are variably increased above the upper limit of the reference interval. Therefore, a biological variation-based quality requirement may not be a clinically relevant quality specification for the FLC assay. In this regard, the RCV value of 54.5% was chosen because it was comparable to the reported interlaboratory reproducibility values achieved during the UKNEQAS FLC EQA scheme [14] and was similar to a clinical cut-point of 50% FLC reduction defining a clinically relevant FLC response reported in a previous clinical recommendation [10].

The method-specific Sigma metrics of κ and λ FLC achieved greater than Three-Sigma for 86–97% of all EQA distributions, and the higher quality specifications consistent with Five-Sigma, beyond which little improvement to quality can be achieved [11], was achieved in 48% of all κ FLC EQA distributions, for the nephelometric and turbidimetric methods, respectively, and 48% and 72% of all λ FLC EQA distributions for the nephelometric and turbidimetric methods, respectively. When only EQA distributions with consensus κ and λ FLC values in the vicinity of κ or λ medical decision points were considered, the method-specific Sigma metrics of κ and λ FLC achieved greater than Three-Sigma for all EQA distributions for both methods, and Five-Sigma was achieved in 67% and 100% of all κ and λ FLC EQA distributions, for the nephelometric and turbidimetric methods, respectively.

We also assessed the harmonization of results between the nephelometric and turbidimetric methods. While the average difference between the two methods is not significant for κ FLC, for λ FLC, the Bland-Altman plot showed a significant difference between the two methods at higher λ FLC concentrations that require a different sample dilution from that of the initial measuring range. This is not surprising as significant method-specific differences in FLC concentrations have been previously reported [14]. Hence, despite the use of the same source of commercial reagent, values were not completely harmonized between assay systems that use different method principles. Possible causes of discordant results between methods include different reaction times, differences in the antigen (sample)/antibody (reagent) ratio in the reaction mixture, different concentrations of polyethylene glycol, different sample dilution protocol and different linearity. The positive difference in nephelometry at higher λ FLC concentrations demonstrated in this study should be investigated further in a larger number of samples.

The method-specific median interlaboratory reproducibility (CV) of κ FLC and λ FLC across 32 EQA specimens ranged from 10 to 12% for both the nephelometric and turbidimetric methods. Interlaboratory reproducibility values reported in this survey were considered acceptable for clinical practice and were better than those achieved during the UK NEQAS external control cycles. Moreover, EQA specimens that had consensus values in the vicinity of κ or λ medical decision points achieved an all-method interlaboratory reproducibility value (CV) of 8–14%, which may be larger than, but not significantly different from, the manufacturer’s predetermined acceptance criteria of total precision (%CV<8.5%) and between-instrument precision (%CV<8.5%), for a relatively narrow FLC concentration range without the need for serum dilution [19].

EQA provides an essential tool for medical laboratories for evaluating assay performances and establishing quality standards [20]. During the 6-year EQA survey, we found that most of the results from participating laboratories meet clinically relevant quality specifications. We could also provide an estimate of the interlaboratory variability in the FLC assay that participants and physicians should be aware of. In addition, we have highlighted the method-specific differences for λ FLC and also how they tend to increase with increasing FLC concentrations. This may have the potential to mislead clinical assessment of the FLC assay results when serial FLC results involving different methods are used.

There are acknowledged limitations of this study. First, the small number of participating laboratories comprising the nephelometric method subgroup may have introduced bias in the estimation of method-specific performance, despite the use of a weighted average. Second, due to the lack of a well-established consensus definition of quality requirement regarding the FLC assay, a reported RCV value of 54.5% derived from the estimate of monoclonal FLC variability in patients with plasma cell disorders was applied as the tolerance limit in calculating the Sigma metrics. Lastly, as a reference measurement procedure or certified reference material is not available for FLC, we could not assess the reported results of the EQA scheme with reference to trueness.

Despite these limitations, the data collected by the first national EQA scheme for the FLC assay in Korea represent a unique resource for clinical laboratories. We used Sigma metrics and a clinically relevant tolerance limit in analyzing the FLC EQA scheme. The study also shows that between-method differences still exit, and further standardization and harmonization efforts are necessary.


Corresponding authors: Myungshin Kim, MD and Yonggoo Kim, MD, Department of Laboratory Medicine, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea; and Catholic Laboratory Development and Evaluation Center, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea, Phone: +82 2 2258 1645 (M. Kim); +82 2 2258 1642 (Y. Kim)

  1. Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

  2. Research funding: None declared.

  3. Employment or leadership: None declared.

  4. Honorarium: None declared.

  5. Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

References

1. Wang L, Chan PC. Measurement uncertainty for serum free light chain assays: estimation and implication on result interpretation. Clin Biochem 2013;46:381–84.10.1016/j.clinbiochem.2012.11.015Search in Google Scholar

2. Dispenzieri A, Kyle R, Merlini G, Miguel JS, Ludwig H, Hajek R, et al. International myeloma working group guidelines for serum-free light chain analysis in multiple myeloma and related disorders. Leukemia 2009;23:215–24.10.1038/leu.2008.307Search in Google Scholar

3. Moreau P, San Miguel J, Ludwig H, Schouten H, Mohty M, Dimopoulos M, et al. Multiple myeloma: ESMO clinical practice guidelines for diagnosis, treatment and follow-up. Ann Oncol 2013;24(Suppl 6):vi133–7.10.1093/annonc/mdt297Search in Google Scholar

4. Tan D, Chng WJ, Chou T, Nawarawong W, Hwang SY, Chim CS, et al. Management of multiple myeloma in Asia: resource-stratified guidelines. Lancet Oncol 2013;14:e571–81.10.1016/S1470-2045(13)70404-2Search in Google Scholar

5. Rajkumar SV, Dimopoulos MA, Palumbo A, Blade J, Merlini G, Mateos MV, et al. International myeloma working group updated criteria for the diagnosis of multiple myeloma. Lancet Oncol 2014;15:e538–48.10.1016/S1470-2045(14)70442-5Search in Google Scholar

6. Pratt G, Jenner M, Owen R, Snowden JA, Ashcroft J, Yong K, et al. Updates to the guidelines for the diagnosis and management of multiple myeloma. Br J Haematol 2014;167:131–3.10.1111/bjh.12926Search in Google Scholar PubMed

7. Clinical and Laboratory Standards Institute. Preparation and validation of commutable frozen human serum pools as secondary reference materials for cholesterol measurement procedures. Approved Guideline. C37-A. Wayne, PA: Clinical and Laboratory Standards Institute, 1999.Search in Google Scholar

8. Healy MJ. Outliers in clinical chemistry quality-control schemes. Clin Chem 1979;25:675–7.10.1093/clinchem/25.5.675Search in Google Scholar

9. Westgard JO, Westgard SA. Assessing quality on the Sigma scale from proficiency testing and external quality assessment surveys. Clin Chem Lab Med 2015;53:1531–5.10.1515/cclm-2014-1241Search in Google Scholar PubMed

10. Dispenzieri A, Zhang L, Katzmann JA, Snyder M, Blood E, Degoey R, et al. Appraisal of immunoglobulin free light chain as a marker of response. Blood 2008;111:4908–15.10.1182/blood-2008-02-138602Search in Google Scholar PubMed PubMed Central

11. Jassam N, Yundt-Pacheco J, Jansen R, Thomas A, Barth JH. Can current analytical quality performance of UK clinical laboratories support evidence-based guidelines for diabetes and ischaemic heart disease? – a pilot study and a proposal. Clin Chem Lab Med 2013;51:1579–84.10.1515/cclm-2012-0840Search in Google Scholar PubMed

12. Katzmann JA, Snyder MR, Rajkumar SV, Kyle RA, Therneau TM, Benson JT, et al. Long-term biological variation of serum protein electrophoresis M-spike, urine M-spike, and monoclonal serum free light chain quantification: implications for monitoring monoclonal gammopathies. Clin Chem 2011;57:1687–92.10.1373/clinchem.2011.171314Search in Google Scholar PubMed PubMed Central

13. Westgard JO, Westgard SA. The quality of laboratory testing today: an assessment of Sigma metrics for analytic quality using performance data from proficiency testing surveys and the CLIA criteria for acceptable performance. Am J Clin Pathol 2006;125:343–54.10.1309/V50H4FRVVWX12C79Search in Google Scholar

14. Vercammen M, Meirlaen P, Broodtaerts L, Vande Broek I, Bossuyt X. Effect of sample dilution on serum free light chain concentration by immunonephelometric assay. Clin Chim Acta 2011;412:1798–804.10.1016/j.cca.2011.06.021Search in Google Scholar PubMed

15. Westgard SA. Utilizing global data to estimate analytical performance on the Sigma scale: a global comparative analysis of methods, instruments, and manufacturers through external quality assurance and proficiency testing programs. Clin Biochem 2016;49:699–707.10.1016/j.clinbiochem.2016.02.013Search in Google Scholar PubMed

16. Vávrová J, Maisnar V, Tichý M, Friedecký B, Čermáková Z, Dastych M, et al. Interlaboratory study of free monoclonal immunoglobulin light chain quantification. Clin Chem Lab Med 2011;49:89–92.10.1515/CCLM.2011.019Search in Google Scholar PubMed

17. Ricos C, Alvarez V, Cava F, García-Lario JV, Hernández A, Jiménez CV, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest 1999;59:491–500.10.1080/00365519950185229Search in Google Scholar PubMed

18. Braga F, Infusino I, Dolci A, Panteghini M. Biological variation of free light chains in serum. Clin Chim Acta 2013;415:10–1.10.1016/j.cca.2012.09.008Search in Google Scholar PubMed

19. U. S. Food & Drug Administration. 510(k) Substantial Equivalence in Determination Decision Memorandum. https://www.accessdata.fda.gov/cdrh_docs/reviews/k150658.pdf (updated on September 2015).Search in Google Scholar

20. Vlašić Tanasković J, Coucke W, Leniček Krleža J, Vuković Rodriguez J. Peer groups splitting in Croatian EQA scheme: a trade-off between homogeneity and sample size number. Clin Chem Lab Med 2017;55:539–45.10.1515/cclm-2016-0284Search in Google Scholar PubMed

Received: 2018-11-05
Accepted: 2019-08-19
Published Online: 2019-09-14
Published in Print: 2019-10-25

©2019 Walter de Gruyter GmbH, Berlin/Boston

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.