To the Editor,
In reading the letter by Åsberg and Skadberg [1], we noted that the Authors seem to mix up the measurement uncertainty (MU) of the values assigned to the end-user calibrator (ucal) with the source of variation related to the calibration procedure of the in vitro diagnostic measurement procedure (IVD-MP) when the calibrator lot is changed. ucal belongs to the MU of the selected calibration hierarchy and related higher-order references together with the MU associated to the manufacturer’s protocol for assignment of the unbiased values to the IVD calibrator [2], 3], while, according to the ISO 20914:2019 Technical Specification, the latter (i.e., the experimental variability that is introduced during each calibration procedure of the IVD-MP in the daily practice) is accounted for in the estimate of ‘long-term imprecision’ (uRw – uncertainty component under conditions of within-laboratory precision, obtained under defined conditions in same laboratory for a period sufficient to include all routine changes to measuring conditions, e.g., different lots of reagents, calibrators, and instrument maintenance; clause 3.40), derived from the internal quality control (IQC) data [3], [4], [5], [6], [7]. These are of course not the same source of MU and the sentence by the Authors “If the effect of ucal is already accounted for in uRw, it should not be added again … as the square root of the sum of the relative variances” is mistaken.
What is demonstrated by the simulation provided by the Authors is not surprising, with the uRw increasing when more calibrator lots are used, confirming the experimental background of the ISO 20914 clause 3.40 asking for the monitoring of the measurement variability for an extended time period, which should include a suitable number of calibrator lots and calibration procedures, simulating what is currently happening in daily laboratory practice [8], 9]. If uRw can be derived from IQC (and the presented simulation shows the influence of one of the most important sources of variability, i.e., the change of calibrator lots), ucal must include all uncertainties introduced by the selected calibration hierarchy for the measurand, beginning with the highest available reference down to the assigned value of the calibrator for the commercial IVD-MP, including the uncertainty of bias correction (ubias), if a not negligible bias has been detected and corrected by the manufacturer when implementing metrological traceability (Figure 1).
![Figure 1:
Sources of measurement uncertainty (u) across the entire metrological traceability chain contributing to the estimate of u of clinical sample [u(y)]. uref, uncertainty of selected higher-order reference; ubias, uncertainty of the bias correction; ucal value assign, uncertainty associated to the manufacturer’s protocol for value assignment to the end-user calibrator; ucal, uncertainty of end-user calibrator; uRw, assay precision under intermediate reproducibility conditions obtained by an individual laboratory using the in vitro diagnostic measurement procedure.](/document/doi/10.1515/cclm-2026-0103/asset/graphic/j_cclm-2026-0103_fig_001.jpg)
Sources of measurement uncertainty (u) across the entire metrological traceability chain contributing to the estimate of u of clinical sample [u(y)]. uref, uncertainty of selected higher-order reference; ubias, uncertainty of the bias correction; ucal value assign, uncertainty associated to the manufacturer’s protocol for value assignment to the end-user calibrator; ucal, uncertainty of end-user calibrator; uRw, assay precision under intermediate reproducibility conditions obtained by an individual laboratory using the in vitro diagnostic measurement procedure.
In conclusion, the authors’ message as presented appears to convey a common misconception about the use of the reproducibility of a measurement result (even when important sources of measurement variability such as the change in calibrator lots are considered) as expression of the overall MU of a clinical sample measurement [10]. The underrated point is that this MU must always include (“combine”) the uncertainty contributions from each higher calibration step in the employed hierarchy. Particularly, the MU at the level of clinical sample results must be the combination of all MU contributions accumulated across the entire traceability chain (i.e., √[ucal2 + uRw2], where ucal=√[uref2 + ucal value assign2 + ubias2, if any]). Only the square root of the sum of the squares of all sources of MU gives the accurate evaluation of MU of clinical sample results. This is the main message of the ISO 20914:2019 guideline [3]. We understand that the information about ucal from IVD manufacturers is sometimes difficult to obtain by end-user laboratories and not rarely incomplete, because, in practice, few manufacturers provide ucal estimated as combined with uref of the selected higher-order reference, as indicated above. What IVD manufacturers are usually providing is just MU of value assignment of commercial calibrators (ucal value assign), so that the medical laboratory should independently retrieve the corresponding uref and combine it to the former to obtain the correct ucal estimate to be finally combined with uRw in the evaluation of MU of laboratory results [11].
-
Research ethics: Not applicable.
-
Informed consent: Not applicable.
-
Author contributions: The authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Use of Large Language Models, AI and Machine Learning Tools: None declared.
-
Conflict of interest: The authors state no conflict of interest.
-
Research funding: None declared.
-
Data availability: Not applicable.
References
1. Åsberg, A, Skadberg, Ø. Combining the calibrator uncertainty and the long-term measurement uncertainty? A comment to the ISO/TS 20914 guideline. Clin Chem Lab Med 2025;26. https://doi.org/10.1515/cclm-2025-1389. [Epub ahead of print].Search in Google Scholar PubMed
2. International Organization for Standardization. In vitro diagnostic medical devices — Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human sample. ISO 17511:2020.Search in Google Scholar
3. International Organization for Standardization. Medical laboratories – practical guidance for the estimation of measurement uncertainty. ISO/TS 20914:2019.Search in Google Scholar
4. Braga, F, Panteghini, M. The utility of measurement uncertainty in medical laboratories. Clin Chem Lab Med 2020;58:1407–13. https://doi.org/10.1515/cclm-2019-1336.Search in Google Scholar PubMed
5. Braga, F, Pasqualetti, S, Aloisio, E, Panteghini, M. The internal quality control in the traceability era. Clin Chem Lab Med 2021;59:291–300. https://doi.org/10.1515/cclm-2020-0371.Search in Google Scholar PubMed
6. Panteghini, M, Krintus, M. Establishing, evaluating and monitoring analytical quality in the traceability era. Crit Rev Clin Lab Sci 2025;62:148–81. https://doi.org/10.1080/10408363.2024.2434562.Search in Google Scholar PubMed
7. Panteghini, M. Not all biases are created equal: how to deal with bias on laboratory measurements. Clin Chem Lab Med 2025;63:916–22. https://doi.org/10.1515/cclm-2024-1208.Search in Google Scholar PubMed
8. Miller, GW. The role of analytical performance specifications in international guidelines and standards dealing with metrological traceability in laboratory medicine. Clin Chem Lab Med 2024;62:1462–9. https://doi.org/10.1515/cclm-2023-1201.Search in Google Scholar PubMed
9. Borrillo, F, Panteghini, M. State-of-the-art model for derivation of analytical performance specifications: how to define the highest level of analytical performance technically achievable. Clin Chem Lab Med 2024;62:1490–6. https://doi.org/10.1515/cclm-2023-1286.Search in Google Scholar PubMed
10. Panteghini, M. The simple reproducibility of a measurement result does not equal its overall measurement uncertainty. Clin Chem Lab Med 2022;60:e221–2. https://doi.org/10.1515/cclm-2022-0618.Search in Google Scholar PubMed
11. Capoferri, A, Pasqualetti, S, Borrillo, F, Dolci, A, Panteghini, M. The information about the metrological traceability pedigree of the in vitro diagnostic calibrators should be improved: the case of plasma ethanol. Clin Chem Lab Med 2025;63:e70–2. https://doi.org/10.1515/cclm-2024-1010.Search in Google Scholar PubMed
© 2026 the author(s), published by De Gruyter, Berlin/Boston
This work is licensed under the Creative Commons Attribution 4.0 International License.