Home Evaluation of a novel, stand-alone system to measure interleukin-6
Article Open Access

Evaluation of a novel, stand-alone system to measure interleukin-6

  • Sascha Dierks , Mechthild Wiederhold , Julie Schanz , Andreas Fischer and Moritz Schnelle EMAIL logo
Published/Copyright: January 27, 2023

To the Editor,

Early detection of sepsis and its subsequent treatment are crucial to prevent life-threatening situations in patient care. Interleukin-6 (IL-6) is a pro-inflammatory cytokine released in response to inflammation by a variety of cell types, including T lymphocytes and endothelial cells [1, 2]. IL-6 plasma concentrations are often used to assess acute inflammatory reactions due to trauma and/or infection, including severe COVID19 [3, 4], and especially in the early detection of neonatal sepsis [5]. In addition, IL-6 plasma levels correlate with the severity of sepsis and may predict outcome [6, 7]. Therefore, assessment of IL-6 plasma concentrations represents a clinically relevant diagnostic tool for monitoring critically ill individuals, including infants. Although well established, options for quantitative IL-6 measurements are often limited to large laboratories with expensive high-throughput devices. In addition, not all companies that offer automated laboratory systems have IL-6 in their own current parameter portfolio (e.g. Abbott, USA). Given the diagnostic and clinical importance of IL-6, however, more options for measuring its concentration in plasma and/or serum are necessary.

In this context, smaller stand-alone systems such as the novel iChroma 50 (Boditech Med Inc., Republic of Korea; distributed by WS Labor, Germany) might represent a useful alternative to the fully automated, large laboratory systems. In order to evaluate this, we compared the IL-6 assay using this novel system (iChroma 50) with the widely used, state-of-the-art method on the cobas e411 (Roche Diagnostics, Germany). Briefly, the iChroma 50 system is an automated, straightforward device for a test card-based fluorescent immunoassay (FIA), as compared to the well-known IL-6 electro-chemiluminescence immunoassay (ECLIA) on the cobas e411. Specific assay details can be found in Supplementary Table 1. In our comparison, 100 lithium heparinate plasma samples were identified based on their initial, medically justified IL-6 measurements on the cobas e411 in the central laboratory of the University Medical Center Goettingen (UMG), covering a wide concentration range (2.9–12,333.0 pg/mL). Samples were then stored at 4 °C before they were simultaneously measured on the iChroma 50 and cobas e411 for this comparative analysis.

The plasma used in this study was leftover material, and all patient samples were processed in an anonymized manner without consideration of further patient-related data. The study was conducted in accordance with the World Medical Declaration of Helsinki and approved by the UMG ethics committee (No.: 20/3/22). For statistical analysis and visualization, Microsoft Excel (Microsoft, USA) and RStudio (RStudio, USA) with the ggplot2 package were used.

IL-6 concentrations ranged from 2.0 to 13,193.6 pg/mL (median=30.9 pg/mL) on the iChroma 50, and 2.9–12,333.0 pg/mL (median= 43.9 pg/mL) on the cobas e411. Of note, ten samples had to be excluded from the following quantitative comparison as they showed negative results, i.e. below the limit of detection (LoD), on both devices. Passing-Bablok regression analysis of the remaining 90 samples revealed good comparability between both systems with a slope of 1.085 and a Y-intercept of 3.079 (Figure 1). The correlation was excellent with r>0.99 (Figure 1). Our data further indicate that IL-6 concentrations below ∼100 pg/mL are systematically measured slightly lower on the iChroma 50 vs. cobas e411, whereas this is the opposite for higher IL-6 concentrations (Figure 1). Agreement of qualitative results (positive/negative) - according to the proposed cut-off of 7 pg/mL - between both systems was very good as 95.0% of all samples (95/100) were equally interpreted using both systems, with a Cohen’s kappa of 0.86 (Supplementary Table 2).

Figure 1: 
Passing-Bablok regression analysis. Passing-Bablok regression analysis between the iChroma 50 and the cobas e411 was performed from IL-6 measurements in 90 lithium heparinate plasma samples. The regression line is shown in red, the bisecting line is presented as a dashed black line. Regression equations with respective 95% confidence intervals (CI) and Pearson’s correlation coefficients (r) are presented in the respective figures. The magnified insert shows the regression line for IL-6 measurements up to 250 pg/mL.
Figure 1:

Passing-Bablok regression analysis. Passing-Bablok regression analysis between the iChroma 50 and the cobas e411 was performed from IL-6 measurements in 90 lithium heparinate plasma samples. The regression line is shown in red, the bisecting line is presented as a dashed black line. Regression equations with respective 95% confidence intervals (CI) and Pearson’s correlation coefficients (r) are presented in the respective figures. The magnified insert shows the regression line for IL-6 measurements up to 250 pg/mL.

Additionally, we assessed the imprecision according to the CLSI EP15 A2 protocol. This was found to be much higher on the iChroma 50 as compared to the cobas e411, with total imprecision coefficients of variation (CVs) ranging from 6.9 to 14.5% on the iChroma 50 vs. 2.7 and 2.4% on the cobas e411 (Table 1). A similar pattern could be observed for the within-run imprecision (CVs of 7.6 and 13.5% on the iChroma 50 vs. 2.8 and 2.6% on the cobas e411) (Table 1). According to our analysis, CVs were particularly high for the high control level (target value 876.81 pg/mL).

Table 1:

Imprecision.

Target conc., pg/mL Mean conc., pg/mL Total imprecision Within-run imprecision
pg/mL CV, % pg/mL CV, %
iChroma 50 41.5 38.3 2.6 6.9 2.9 7.6
876.8 1,011.0 146.8 14.5 136.1 13.5

cobas e411 40.1 41.1 1.1 2.7 1.2 2.8
255.0 264.1 6.3 2.4 6.7 2.6
  1. Imprecision, i.e. total and within-run imprecision, was assessed for IL-6 measurements on the iChroma 50 and cobas e411. For this, two controls each with different target values were measured on the iChroma 50 and cobas e411 three times a day on five consecutive days according to the CLSI EP15 A2 protocol. CV, coefficient of variation.

As mentioned above, IL-6 assessment is of great diagnostic importance in neonatal sepsis. Due to the limited amount of plasma in neonatal samples, the following algorithm is often used in laboratories to obtain as much information as possible from one single plasma sample containing ethylenediamine tetraacetic acid (EDTA) as anticoagulant: (i) blood count before centrifugation (this analysis requires EDTA as anticoagulant), (ii) centrifugation, (iii) measurements of the C-reactive protein (CRP) and IL-6 in the remaining EDTA plasma. Therefore, we tested whether IL-6 measurements on the iChroma 50 were also comparable with the cobas e411 in 30 EDTA samples (instead of lithium heparinate). Passing-Bablok analysis revealed moderate comparability with a slope of 1.234 and a Y-intercept of 16.122, indicating a relevant systematic bias (Supplementary Figure 1). In addition, this suggests that the type of anticoagulant may affect the results. This analysis had the limitation that EDTA samples were from adults rather than neonates due to the limited sample volume in neonatal specimens, which did not allow us to perform multiple measurements. This has to be evaluated separately in future studies.

Taken together, our data suggest that the iChroma 50 may serve as a suitable, stand-alone alternative to allow sufficient assessment of IL-6. This is particularly due to the good comparability with the state-of-the-art assay on the cobas e411 and the excellent correlation when using lithium heparinate samples. However, our data demonstrate relevant differences in comparability of both systems for EDTA plasma samples. Therefore, care should be taken when measuring IL-6 concentrations on the iChroma 50 in EDTA plasma, which is of particular clinical relevance with respect to neonatal samples. Another weakness is the relatively high imprecision, especially at high IL-6 concentrations, with CVs up to 14.5%. This is not in concordance with the precision data provided by the manufacturer (CV of 6.5% for an IL-6 concentration of 1,274 pg/mL) and should be investigated in further studies. It appears to be an assay-specific problem rather than a general issue with the iChroma 50, as a recent study demonstrated excellent precision for procalcitonin (PCT) measurements on the iChroma, even for high PCT concentrations [8]. The imprecision of IL-6 measurements clearly must be taken into account when using the iChroma 50 in regular patient care; false clinical conclusions could be made when patients are followed up longitudinally, e.g. a reduction in IL-6 could be interpreted as therapeutic success although it may simply be based on the imprecise measurements. This consequential problem needs to be addressed.


Corresponding author: Priv. Doz. Dr. med. Dr. Moritz Schnelle, MD, PhD, Department of Clinical Chemistry, University Medical Center Goettingen, Robert-Koch-Str. 40, 37075 Goettingen, Germany, Phone: +49 551-3965510, E-mail:

Acknowledgments

Reagents for the iChroma 50 were kindly provided by Abbott.

  1. Research funding: None declared.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Direct informed consent was not necessary as only left over plasma material was used in this study and no further patient related data were included.

  5. Ethical approval: Research involving human subjects complied with all relevant national regulations, institutional policies and is in accordance with the tenets of the Helsinki Declaration (as revised in 2013), and has been approved by the authors’ Institutional Review Board (No.: 20/03/22).

References

1. Tanaka, T, Narazaki, M, Kishimoto, T. IL-6 in inflammation, immunity, and disease. Cold Spring Harbor Perspect Biol 2014;6:a016295. https://doi.org/10.1101/cshperspect.a016295.Search in Google Scholar PubMed PubMed Central

2. Thompson, DK, Huffman, KM, Kraus, WE, Kraus, VB. Critical appraisal of four IL-6 immunoassays. PLoS One 2012;7:e30659. https://doi.org/10.1371/journal.pone.0030659.Search in Google Scholar PubMed PubMed Central

3. Liu, T, Zhang, J, Yang, Y, Ma, H, Li, Z, Zhang, J, et al.. The role of interleukin-6 in monitoring severe case of coronavirus disease 2019. EMBO Mol Med 2020;12:e12421. https://doi.org/10.15252/emmm.202012421.Search in Google Scholar PubMed PubMed Central

4. Henry, BM, Oliveira, MHSde, Benoit, S, Plebani, M, Lippi, G. Hematologic, biochemical and immune biomarker abnormalities associated with severe illness and mortality in coronavirus disease 2019 (COVID-19): a meta-analysis. Clin Chem Lab Med 2020;58:1021–8. https://doi.org/10.1515/cclm-2020-0369.Search in Google Scholar PubMed

5. Küster, H, Weiss, M, Willeitner, AE, Detlefsen, S, Jeremias, I, Zbojan, J, et al.. Interleukin-1 receptor antagonist and interleukin-6 for early diagnosis of neonatal sepsis 2 days before clinical manifestation. Lancet 1998;352:1271–7. https://doi.org/10.1016/s0140-6736(98)08148-3.Search in Google Scholar

6. Damas, P, Ledoux, D, Nys, M, Vrindts, Y, Groote, Dde, Franchimont, P, et al.. Cytokine serum level during severe sepsis in human IL-6 as a marker of severity. Ann Surg 1992;215:356–62. https://doi.org/10.1097/00000658-199204000-00009.Search in Google Scholar PubMed PubMed Central

7. Hack, CE, Groot, ERde, Felt-Bersma, RJ, Nuijens, JH, van Strack Schijndel, RJ, Eerenberg-Belmer, AJ, et al.. Increased plasma levels of interleukin-6 in sepsis. Blood 1989;74:1704–10. https://doi.org/10.1182/blood.v74.5.1704.1704.Search in Google Scholar

8. Stenner, E, Barbati, G, West, N, Del Ben, F, Martin, F, Ruscio, M. Agreement between procalcitonin measurements using the new point-of-care testing ichroma™ reader and the automated Kryptor instrument. J Lab Med 2019;43:169–72. https://doi.org/10.1515/labmed-2018-0179.Search in Google Scholar


Supplementary Material

This article contains supplementary material (https://doi.org/10.1515/cclm-2022-1320).


Received: 2022-12-29
Accepted: 2023-01-19
Published Online: 2023-01-27
Published in Print: 2023-06-27

© 2023 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. ChatGPT: Angel or Demond? Critical thinking is still needed
  4. Mini Review
  5. Diagnostic accuracy of Siemens SARS-CoV-2 Antigen (CoV2Ag) chemiluminescent immunoassay for diagnosing acute SARS-CoV-2 infection: a pooled analysis
  6. Opinion Papers
  7. Neurofilament light chain as neuronal injury marker – what is needed to facilitate implementation in clinical laboratory practice?
  8. Clinical evidence requirements according to the IVDR 2017/746: practical tools and references for underpinning clinical evidence of IVD-MDs
  9. EFLM Papers
  10. Potentials and pitfalls of ChatGPT and natural-language artificial intelligence models for the understanding of laboratory medicine test results. An assessment by the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group on Artificial Intelligence (WG-AI)
  11. Detection of antinuclear antibodies: recommendations from EFLM, EASI and ICAP
  12. Analytical aspects of the antinuclear antibody test by HEp-2 indirect immunofluorescence: EFLM report on an international survey
  13. Guidelines and Recommendations
  14. Variability of cardiac troponin levels in normal subjects and in patients with cardiovascular diseases: analytical considerations and clinical relevance
  15. General Clinical Chemistry and Laboratory Medicine
  16. Pre-analytical long-term stability of neopterin and neurofilament light in stored cerebrospinal fluid samples
  17. Development of a candidate reference measurement procedure by ID-LC-MS/MS for total tau protein measurement in cerebrospinal fluid (CSF)
  18. Assessing the commutability of candidate reference materials for the harmonization of neurofilament light measurements in blood
  19. Surface plasmon resonance assays for the therapeutic drug monitoring of infliximab indicate clinical relevance of anti-infliximab antibody binding properties
  20. An antibody-free LC-MS/MS method for the quantification of sex hormone binding globulin in human serum and plasma
  21. Accurate stratification between VEXAS syndrome and differential diagnoses by deep learning analysis of peripheral blood smears
  22. Establishing quality indicators for point of care glucose testing: recommendations from the Canadian Society for Clinical Chemists Point of Care Testing and Quality Indicators Special Interest Groups
  23. Clinical implication by differential analytical performances of serum free light chain quantitation analysis using fully automated analyzers
  24. Simultaneous analysis of antihyperglycemic small molecule drugs and peptide drugs by means of dual liquid chromatography high-resolution mass spectrometry
  25. Reference Values and Biological Variations
  26. Reference intervals for thyroid biomarkers to enhance the assessment of thyroid status in childhood and adolescence
  27. Biological variation of CA 15-3, CA 125 and HE 4 on lithium heparinate plasma in apparently healthy Caucasian volunteers
  28. Cancer Diagnostics
  29. Individual risk prediction of high grade prostate cancer based on the combination between total prostate-specific antigen (PSA) and free to total PSA ratio
  30. Cardiovascular Diseases
  31. Using logistic regression models to investigate the effects of high-sensitivity cardiac troponin T confounders on ruling in acute myocardial infarction
  32. Infectious Diseases
  33. Assessment of humoral and cellular immunity after bivalent BNT162b2 vaccination and potential association with reactogenicity
  34. Three rounds of a national external quality assessment reveal a link between disharmonic anti-SARS-CoV-2 antibody quantifications and the infection stage
  35. Corrigendum
  36. Where is laboratory medicine headed in the next decade? Partnership model for efficient integration and adoption of artificial intelligence into medical laboratories
  37. Letters to the Editor
  38. Please do not call it Theranos
  39. Lot-to-lot bias for high-sensitivity cardiac troponin I concentrations ≥1000 ng/L
  40. Lot-to-lot reagent changes and commutability of quality testing materials for total bile acid measurements
  41. On the importance of sampling interval in studies of biological variation in thyroid function
  42. Stability of plasma renin concentration based on plasma freezing time, as an adjunct to the stability data reported in the paper by Hepburn and others
  43. Falsely low beta-hCG results in pregnant woman on Siemens Atellica: don’t forget the “hook effect”
  44. Proenkephalin A 119–159 (penKid) – a novel biomarker and its quantification on the Nexus IB10 POC system for assessing kidney function
  45. Investigating the polyuria-polydipsia syndrome: the “PP” Shiny app
  46. Evaluation of a novel, stand-alone system to measure interleukin-6
  47. Spurious low WBC count in the WNR channel of Sysmex XN-9000 hematology analyzer in a case with leukocyte aggregation
Downloaded on 10.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2022-1320/html?lang=en
Scroll to top button