Two-site evaluation of the Roche Elecsys Vitamin D total III assay

  • Julien Favresse EMAIL logo , Johan Schiettecatte , Fleur Wolff , Frederic Cotton , Marc Elsen , Christine Eucher , Catherine Rosseels and Ellen Anckaert
Published/Copyright: July 15, 2022

Abstract

Objectives

The high request for vitamin D testing in the last decades has led manufacturers to develop assays on automated immunoassay platforms. The objective of this study was to evaluate the performance of the new Elecsys Vitamin D total III assay for the measurement of total 25(OH)D.

Methods

A total of 844 serum samples collected in two clinical laboratories were used to evaluate the new Roche Elecsys Vitamin D total III assay. Comparisons with Roche Elecsys Vitamin D total II and liquid chromatography tandem mass spectrometry (LC-MS/MS) were carried out. Additionally, assay imprecision, linearity, matrix effects, biotin interference, cross-reactivity with 24,25(OH)2D3 and 3-epi-25(OH)D3, and outlier rate were evaluated for the Elecsys Vitamin D total III assay.

Results

Only the comparison between LC-MS/MS and Roche Elecsys Vitamin D total III achieved the optimal specification for bias (i.e., <3.4%). Imprecision, linearity and matrix effects showed acceptable results. The biotin interference threshold was increased up to 1,200 ng/mL and the outlier rate was low (0.26%). The cross-reactivity with 24,25(OH)2D3 and 3-epi-25(OH)D3 was weak or modest in available patient samples. However, using SRM972a with a high level of 3-epi-25(OH)D3 (enriched) revealed an important cross-reactivity with both Roche Elecsys Vitamin D total II and III assays (+74.7% and +73.7%).

Conclusions

In conclusion, the Roche Elecsys Vitamin D total III assay presents several advantages compared to the previous assay generation: higher biotin interference threshold, broader measuring range, and better comparability with LC-MS/MS. However, the cross-reactivity toward 3-epi-25(OH)D3 is still problematic in high titer samples.


Corresponding author: Julien Favresse, Department of Laboratory Medicine, Clinique Saint-Luc Bouge, Namur, Belgium; and Department of Pharmacy, Namur Research Institute for Life Sciences, University of Namur, Namur, 5004 Bouge, Belgium, Phone: +32 81 20 91 44, E-mail:

  1. Research funding: None declared.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Informed consent is not required for the evaluation of a new generation of assay.

  5. Ethical approval: The local Institutional Review Board deemed the study exempt from review.

References

1. Rifai, N, Horvath, AR, Wittwer, C. Tietz textbook of clinical chemistry and molecular diagnostics, 6th ed. St. Louis, Missouri: Elsevier; 2018.Search in Google Scholar

2. Farrell, CJ, Herrmann, M. Determination of vitamin D and its metabolites. Best Pract Res Clin Endocrinol Metabol 2013;27:675–88. https://doi.org/10.1016/j.beem.2013.06.001.Search in Google Scholar PubMed

3. Souberbielle, JC, Body, JJ, Lappe, JM, Plebani, M, Shoenfeld, Y, Wang, TJ, et al.. Vitamin D and musculoskeletal health, cardiovascular disease, autoimmunity and cancer: recommendations for clinical practice. Autoimmun Rev 2010;9:709–15. https://doi.org/10.1016/j.autrev.2010.06.009.Search in Google Scholar PubMed

4. Altieri, B, Cavalier, E, Bhattoa, HP, Perez-Lopez, FR, Lopez-Baena, MT, Perez-Roncero, GR, et al.. Vitamin D testing: advantages and limits of the current assays. Eur J Clin Nutr 2020;74:231–47. https://doi.org/10.1038/s41430-019-0553-3.Search in Google Scholar PubMed

5. Bjerg, LN, Halgreen, JR, Hansen, SH, Morris, HA, Jorgensen, NR. An evaluation of total 25-hydroxyvitamin D assay standardization: where are we today? J Steroid Biochem Mol Biol 2019;190:224–33. https://doi.org/10.1016/j.jsbmb.2019.03.015.Search in Google Scholar PubMed

6. Fabregat-Cabello, N, Farre-Segura, J, Huyghebaert, L, Peeters, S, Le Goff, C, Souberbielle, JC, et al.. A fast and simple method for simultaneous measurements of 25(OH)D, 24,25(OH)2D and the Vitamin D metabolite ratio (VMR) in serum samples by LC-MS/MS. Clin Chim Acta 2017;473:116–23. https://doi.org/10.1016/j.cca.2017.08.024.Search in Google Scholar PubMed

7. CDC. Vitamin D Standardization-Certification Program (CDC VDSCP); 2022. Available from: https://www.cdc.gov/labstandards/pdf/hs/CDC_Certified_Vitamin_D_Assays-508.pdf [Accessed 25 May 2022].Search in Google Scholar

8. Wise, SA, Tai, SS, Nelson, MA, Burdette, CQ, Camara, JE, Hoofnagle, AN, et al.. Interlaboratory comparison for the determination of 24,25-dihydroxyvitamin D(3) in human serum using liquid chromatography with tandem mass spectrometry. J AOAC Int 2017;100:1308–17. https://doi.org/10.5740/jaoacint.17-0183.Search in Google Scholar PubMed

9. Favresse, J, Bayart, JL, Gruson, D, Bernardini, S, Clerico, A, Perrone, M. The underestimated issue of non-reproducible cardiac troponin I and T results: case series and systematic review of the literature. Clin Chem Lab Med 2021;59:1201–11. https://doi.org/10.1515/cclm-2020-1564.Search in Google Scholar PubMed

10. Favresse, J, Cadrobbi, J, Eucher, C, Laffineur, K, Rosseels, C, Pieters, D, et al.. Non-reproducible cardiac troponin results occurring with a particular reagent lot. Clin Chem Lab Med 2020;59:e9–12. https://doi.org/10.1515/cclm-2020-0562.Search in Google Scholar PubMed

11. Trambas, C, Lu, Z, Yen, T, Sikaris, K. Depletion of biotin using streptavidin-coated microparticles: a validated solution to the problem of biotin interference in streptavidin-biotin immunoassays. Ann Clin Biochem 2018;55:216–26. https://doi.org/10.1177/0004563217707783.Search in Google Scholar PubMed

12. Aarsand, AK, Fernandez-Calle, P, Webster, C, Coskun, A, Gonzales-Lao, E, Diaz-Garzon, J, et al.. The EFLM biological variation database. Available from: https://biologicalvariation.eu/[Accessed 2 July 2022].Search in Google Scholar

13. Enko, D, Kriegshauser, G, Stolba, R, Worf, E, Halwachs-Baumann, G. Method evaluation study of a new generation of vitamin D assays. Biochem Med 2015;25:203–12. https://doi.org/10.11613/bm.2015.020.Search in Google Scholar

14. Asif, M, Groboske, SE, Leung, EKY, Yeo, KJ, van Wijk, XMR. Evaluation of a new generation automated assay for 25-hydroxy vitamin D based on competitive protein binding. J Appl Lab Med 2019;4:247–53. https://doi.org/10.1373/jalm.2018.028415.Search in Google Scholar PubMed

15. Batista, MC, Menegat, FD, Ferreira, CES, Faulhaber, ACL, Campos, D, Mangueira, CLP. Analytical and clinical validation of the new Roche Elecsys vitamin D total II assay. Clin Chem Lab Med 2018;56:e298–301. https://doi.org/10.1515/cclm-2018-0406.Search in Google Scholar PubMed

16. van den Ouweland, JM, Beijers, AM, van Daal, H, Elisen, MG, Steen, G, Wielders, JP. Evaluation of 3-epi-25-hydroxyvitamin D3 cross-reactivity in the Roche Elecsys vitamin D total protein binding assay. Clin Chem Lab Med 2014;52:373–80. https://doi.org/10.1515/cclm-2013-0702.Search in Google Scholar PubMed

17. Berger, SE, Van Rompay, MI, Gordon, CM, Goodman, E, Eliasziw, M, Holick, MF, et al.. Investigation of the C-3-epi-25(OH)D3 of 25-hydroxyvitamin D3 in urban schoolchildren. Appl Physiol Nutr Metabol 2018;43:259–65. https://doi.org/10.1139/apnm-2017-0334.Search in Google Scholar PubMed PubMed Central

18. Cavalier, E, Lukas, P, Bekaert, AC, Peeters, S, Le Goff, C, Yayo, E, et al.. Analytical and clinical evaluation of the new Fujirebio Lumipulse(R)G non-competitive assay for 25(OH)-vitamin D and three immunoassays for 25(OH)D in healthy subjects, osteoporotic patients, third trimester pregnant women, healthy African subjects, hemodialyzed and intensive care patients. Clin Chem Lab Med 2016;54:1347–55. https://doi.org/10.1515/cclm-2015-0923.Search in Google Scholar PubMed

19. Bayart, JL, Favresse, J, Stoefs, A, Closset, M, Roy, T, Fillee, C, et al.. Biotin interferences: have we neglected the impact on serological markers? Clin Chim Acta 2020;503:107–12. https://doi.org/10.1016/j.cca.2020.01.012.Search in Google Scholar PubMed

20. Favresse, J, Burlacu, MC, Maiter, D, Gruson, D. Interferences with thyroid function immunoassays: clinical implications and detection algorithm. Endocr Rev 2018;39:830–50. https://doi.org/10.1210/er.2018-00119.Search in Google Scholar PubMed

21. Samarasinghe, S, Meah, F, Singh, V, Basit, A, Emanuele, N, Emanuele, MA, et al.. Biotin interference with routine clinical immunoassays: understand the causes and mitigate the risks. Endocr Pract 2017;23:989–98. https://doi.org/10.4158/ep171761.ra.Search in Google Scholar PubMed

22. Nevraumont, A, Deltombe, M, Favresse, J, Guillaume, L, Chapelle, V, Twerenbold, R, et al.. Interferences with cardiac biomarker assays: understanding the clinical impact. Eur Heart J 2022;43:2286–88. https://doi.org/10.1093/eurheartj/ehab924.Search in Google Scholar PubMed

23. Favresse, J, Bayart, JL, Burlacu, MC, Maiter, D, Dogne, JM, Douxfils, J, et al.. Usefulness of a non-streptavidin bead technology to overcome biotin interference: proof of principle with 25-OH vitamin D, TSH, and FT4. J Appl Lab Med 2021;6:1072–7. https://doi.org/10.1093/jalm/jfaa203.Search in Google Scholar PubMed

24. Carter, GD, Berry, J, Cavalier, E, Durazo-Arvizu, R, Gunter, E, Jones, G, et al.. Biotin supplementation causes erroneous elevations of results in some commercial serum 25-hydroxyvitamin d (25OHD) assays. J Steroid Biochem Mol Biol 2020;200:105639. https://doi.org/10.1016/j.jsbmb.2020.105639.Search in Google Scholar PubMed

25. Evenepoel, A, Schiettecatte, J, Van Dalem, A, Anckaert, E. Biotin interference: evaluation of an updated thyroglobulin electrochemiluminescent immunoassay. Clin Chem Lab Med 2022;60:e46–9. https://doi.org/10.1515/cclm-2021-0969.Search in Google Scholar PubMed

26. Mzougui, S, Favresse, J, Soleimani, R, Fillee, C, Gruson, D. Biotin interference: evaluation of a new generation of electrochemiluminescent immunoassays for high-sensitive troponin T and thyroid-stimulating hormone testing. Clin Chem Lab Med 2020;58:2037–45. https://doi.org/10.1515/cclm-2020-0214.Search in Google Scholar PubMed

27. Findeisen, PLM, Bendig, G, Grimme, J, Moser, E, Christenson, R. Multicenter performance evaluation of the new Elecsys Vitamin D total III assay. In: EuroMedLab congress in Munich, April 10–14 2022, poster M056. https://doi.org/10.1515/cclm-2021-5004.10.1515/cclm-2021-5004Search in Google Scholar PubMed


Supplementary Material

The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2022-0177).


Received: 2022-02-25
Accepted: 2022-07-03
Published Online: 2022-07-15
Published in Print: 2022-09-27

© 2022 Walter de Gruyter GmbH, Berlin/Boston

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. The relevance of establishing method-dependent decision thresholds of serum folate in pregnancy and lactation: when the laboratory stewardship meets the health-care needs
  4. Reviews
  5. Strategies to minimise the current disadvantages experienced by women in faecal immunochemical test-based colorectal cancer screening
  6. What is the impact of circulating histones in COVID-19: a systematic review
  7. Mini Review
  8. Diagnostic utility of pleural cell-free nucleic acids in undiagnosed pleural effusions
  9. Guidelines and Recommendations
  10. Evaluation of the cardiovascular risk in patients undergoing major non-cardiac surgery: role of cardiac-specific biomarkers
  11. Genetics and Molecular Diagnostics
  12. Detection of DNA copy number alterations by matrix-assisted laser desorption/ionization time-of-flight mass spectrometric analysis of single nucleotide polymorphisms
  13. General Clinical Chemistry and Laboratory Medicine
  14. Fifty years of newborn screening for congenital hypothyroidism: current status in Australasia and the case for harmonisation
  15. Commutability assessment of reference materials for homocysteine
  16. External quality assessment for detection of methylated Syndecan 2 (SDC2) in China
  17. Patient-based real-time quality control for quantitative hepatitis B virus DNA test using moving rate of positive and negative patient results
  18. Performance of HDL-C measurements assessed by a 4-year trueness-based EQA/PT program in China
  19. Two-site evaluation of the Roche Elecsys Vitamin D total III assay
  20. A novel point-of-care device accurately measures thyrotropin in whole blood, capillary blood and serum
  21. IgA rheumatoid factor in rheumatoid arthritis
  22. Reference Values and Biological Variations
  23. Reference intervals of 14 biochemical markers for children and adolescence in China: the PRINCE study
  24. Chinese normotensive and essential hypertensive reference intervals for plasma aldosterone and renin activity by liquid chromatography-tandem mass spectrometry
  25. Biological variation estimates obtained from Chinese subjects for 32 biochemical measurands in serum
  26. Cancer Diagnostics
  27. Testosterone analysis in castrated prostate cancer patients: suitability of the castration cut-off and analytical accuracy of the present-day clinical immunoassays
  28. Cardiovascular Diseases
  29. Point-of-care high-sensitivity troponin-I analysis in capillary blood for acute coronary syndrome diagnostics
  30. Platelet count and clinical outcomes among ischemic stroke patients with endovascular thrombectomy in DIRECT-MT
  31. Infectious Diseases
  32. Measurement of anti SARS-CoV-2 RBD IgG in saliva: validation of a highly sensitive assay and effects of the sampling collection method and correction by protein
  33. Letters to the Editors
  34. The simple reproducibility of a measurement result does not equal its overall measurement uncertainty
  35. Complexity of serological assays and misunderstandings of WHO International Units
  36. Ct values of different SARS CoV2 variants: a single center observational study from Innsbruck, Austria
  37. Alignment of the new generation of Abbott Alinity γ-glutamyltransferase assay to the IFCC reference measurement system should be improved
  38. Thermal and chronological stability of monocyte distribution width (MDW), the new biomarker for sepsis
  39. Two comments for the current commutability assessment
  40. Plasma neurofilament light chain in Chinese children with later-onset spinal muscular atrophy
Downloaded on 23.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2022-0177/html
Scroll to top button