Impact of ultra-low temperature long-term storage on the preanalytical variability of twenty-one common biochemical analytes

  • , , , and EMAIL logo
Published/Copyright: April 26, 2022

Abstract

Objectives

Retrospective studies frequently assume analytes long-term stability at ultra-low temperatures. However, these storage conditions, common among biobanks and research, may increase the preanalytical variability, adding a potential uncertainty to the measurements. This study is aimed to evaluate long-term storage stability of different analytes at <−70 °C and to assess its impact on the reference change value formula.

Methods

Twenty-one analytes commonly measured in clinical laboratories were quantified in 60 serum samples. Samples were immediately aliquoted and frozen at <−70 °C, and reanalyzed after 11 ± 3.9 years of storage. A change in concentration after storage was considered relevant if the percent deviation from the baseline measurement was significant and higher than the analytical performance specifications.

Results

Preanalytical variability (CVP) due to storage, determined by the percentage deviation, showed a noticeable dispersion. Changes were relevant for alanine aminotransferase, creatinine, glucose, magnesium, potassium, sodium, total bilirubin and urate. No significant differences were found in aspartate aminotransferase, calcium, carcinoembryonic antigen, cholesterol, C-reactive protein, direct bilirubin, free thryroxine, gamma-glutamyltransferase, lactate dehydrogenase, prostate-specific antigen, triglycerides, thyrotropin, and urea. As nonnegligible, CVP must remain included in reference change value formula, which was modified to consider whether one or two samples were frozen.

Conclusions

After long-term storage at ultra-low temperatures, there was a significant variation in some analytes that should be considered. We propose that reference change value formula should include the CVP when analyzing samples stored in these conditions.


Corresponding author: Álvaro González, PhD, Service of Biochemistry, Clínica Universidad de Navarra, Avenida Pío XII 36, 31008 Pamplona, Spain, Phone: +34 948 255400, Fax: +34 948 296 500, E-mail:
Estibaliz Alegre and Nerea Varo contributed equally to this work.

Funding source: Roche Diagnostics

Acknowledgments

We thank Sonia Irisarri, Ruben del Campo, and Dra. María Romero for their technical assistance.

  1. Research funding: This study was partially supported by Roche Diagnostics, Barcelona, Spain.

  2. Author contribution: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Not applicable.

  5. Ethical approval: Sample collection and storage was approved by the local Ethics Committee (project 14/2004 and 111/2010 and collection ISCIII C.0003132).

References

1. Brinc, D, Chan, MK, Venner, AA, Pasic, MD, Colantonio, D, Kyriakopolou, L, et al.. Long-term stability of biochemical markers in pediatric serum specimens stored at −80 degrees C: a CALIPER Substudy. Clin Biochem 2012;45:816–26, https://doi.org/10.1016/j.clinbiochem.2012.03.029.Search in Google Scholar PubMed

2. Schuh, MP, Nehus, E, Ma, Q, Haffner, C, Bennett, M, Krawczeski, CD, et al.. Long-term stability of urinary biomarkers of acute kidney injury in children. Am J Kidney Dis 2016;67:56–61, https://doi.org/10.1053/j.ajkd.2015.04.040.Search in Google Scholar PubMed PubMed Central

3. Gislefoss, RE, Grimsrud, TK, Morkrid, L. Stability of selected serum proteins after long-term storage in the Janus Serum Bank. Clin Chem Lab Med 2009;47:596–603, https://doi.org/10.1515/CCLM.2009.121.Search in Google Scholar PubMed

4. Elliott, P, Peakman, TC, Biobank, U. The UK Biobank sample handling and storage protocol for the collection, processing and archiving of human blood and urine. Int J Epidemiol 2008;37:234–44, https://doi.org/10.1093/ije/dym276.Search in Google Scholar PubMed

5. Haslacher, H, Szekeres, T, Gerner, M, Ponweiser, E, Repl, M, Wagner, OF, et al.. The effect of storage temperature fluctuations on the stability of biochemical analytes in blood serum. Clin Chem Lab Med 2017;55:974–83, https://doi.org/10.1515/cclm-2016-0608.Search in Google Scholar PubMed

6. Hubel, A, Spindler, R, Skubitz, AP. Storage of human biospecimens: selection of the optimal storage temperature. Biopreserv Biobanking 2014;12:165–75, https://doi.org/10.1089/bio.2013.0084.Search in Google Scholar PubMed

7. Ellervik, C, Vaught, J. Preanalytical variables affecting the integrity of human biospecimens in biobanking. Clin Chem 2015;61:914–34, https://doi.org/10.1373/clinchem.2014.228783.Search in Google Scholar PubMed

8. Vermeersch, P, Frans, G, von Meyer, A, Costelloe, S, Lippi, G, Simundic, AM. How to meet ISO15189:2012 pre-analytical requirements in clinical laboratories? A consensus document by the EFLM WG-PRE. Clin Chem Lab Med 2021;59:1047–61, https://doi.org/10.1515/cclm-2020-1859.Search in Google Scholar PubMed

9. Mouttham, L, Garrison, SJ, Archer, DL, Castelhano, MG. A biobank’s journey: implementation of a quality management system and accreditation to ISO 20387. Biopreserv Biobanking 2021;19:163–70.10.1089/bio.2020.0068Search in Google Scholar PubMed

10. Kachhawa, K, Kachhawa, P, Varma, M, Behera, R, Agrawal, D, Kumar, S. Study of the stability of various biochemical analytes in samples stored at different predefined storage conditions at an accredited laboratory of India. J Lab Phys 2017;9:11–5, https://doi.org/10.4103/0974-2727.187928.Search in Google Scholar PubMed PubMed Central

11. Veronesi, A, Rota, C, Carone, C, Trenti, T, Cariani, E. Long-term stability of serum samples positive for carbohydrate deficient transferrin (CDT) routinely stored at −20 degrees C. Clin Chem Lab Med 2016;54:e285–7, https://doi.org/10.1515/cclm-2016-0074.Search in Google Scholar PubMed

12. Jansen, E, Beekhof, P, Viezeliene, D, Muzakova, V, Skalicky, J. Long-term stability of cancer biomarkers in human serum: biomarkers of oxidative stress and redox status, homocysteine, CRP and the enzymes ALT and GGT. Biomarkers Med 2015;9:425–32, https://doi.org/10.2217/bmm.15.14.Search in Google Scholar PubMed

13. Reed, AB, Ankerst, DP, Leach, RJ, Vipraio, G, Thompson, IM, Parekh, DJ. Total prostate specific antigen stability confirmed after long-term storage of serum at -80C. J Urol 2008;180:534–7, https://doi.org/10.1016/j.juro.2008.04.021.Search in Google Scholar PubMed

14. Egger, M, Dieplinger, B, Mueller, T. One-year in vitro stability of cardiac troponins and galectin-3 in different sample types. Clin Chim Acta 2018;476:117–22, https://doi.org/10.1016/j.cca.2017.11.018.Search in Google Scholar PubMed

15. Milevoj Kopcinovic, L, Brcic, M, Vrtaric, A, Unic, A, Bozovic, M, Nikolac Gabaj, N, et al.. Long-term stability of clinically relevant chemistry analytes in pleural and peritoneal fluid. Biochem Med 2020;30:020701, https://doi.org/10.11613/BM.2020.020701.Search in Google Scholar PubMed PubMed Central

16. Fraser, CG. Improved monitoring of differences in serial laboratory results. Clin Chem 2011;57:1635–7, https://doi.org/10.1373/clinchem.2011.175026.Search in Google Scholar PubMed

17. Lippi, G, Chance, JJ, Church, S, Dazzi, P, Fontana, R, Giavarina, D, et al.. Preanalytical quality improvement: from dream to reality. Clin Chem Lab Med 2011;49:1113–26, https://doi.org/10.1515/CCLM.2011.600.Search in Google Scholar PubMed

18. Carobene, A, Marino, I, Coşkun, A, Serteser, M, Unsal, I, Guerra, E, et al.. The EuBIVAS project: within- and between-subject biological variation data for serum creatinine using enzymatic and alkaline picrate methods and implications for monitoring. Clin Chem 2017;63:1527–36, https://doi.org/10.1373/clinchem.2017.275115.Search in Google Scholar PubMed

19. Harris, EK, Yasaka, T. On the calculation of a “reference change” for comparing two consecutive measurements. Clin Chem 1983;29:25–30, https://doi.org/10.1093/clinchem/29.1.25.Search in Google Scholar

20. Fraser, CG. Reference change values. Clin Chem Lab Med 2011;50:807–12, https://doi.org/10.1515/CCLM.2011.733.Search in Google Scholar PubMed

21. Gomez-Rioja, R, Segovia Amaro, M, Diaz-Garzon, J, Bauca, JM, Martinez Espartosa, D, Fernandez-Calle, P, et al.. A protocol for testing the stability of biochemical analytes. Technical document. Clin Chem Lab Med 2019;57:1829–36, https://doi.org/10.1515/cclm-2019-0586.Search in Google Scholar PubMed

22. Gomez Rioja, R, Martinez Espartosa, D, Segovia, M, Ibarz, M, Llopis, MA, Bauca, JM, et al.. Laboratory sample stability. Is it possible to define a consensus stability function? An example of five blood magnitudes. Clin Chem Lab Med 2018;56:1806–18, https://doi.org/10.1515/cclm-2017-1189.Search in Google Scholar PubMed

23. Shimizu, Y, Ichihara, K. Elucidation of stability profiles of common chemistry analytes in serum stored at six graded temperatures. Clin Chem Lab Med 2019;57:1388–96, https://doi.org/10.1515/cclm-2018-1109.Search in Google Scholar PubMed

24. EFLM biological variation. Available from: https://biologicalvariation.eu/ [Accessed 14 Jan 2022].Search in Google Scholar

25. Muzakova, V, Beekhof, PK, Jansen, E. Very long-term stability of lipid biomarkers in human serum. Anal Biochem 2020;597:113695, https://doi.org/10.1016/j.ab.2020.113695.Search in Google Scholar PubMed

26. Yu, R, Dan, Y, Xiang, X, Zhou, Y, Kuang, X, Yang, G, et al.. Stability of chronic hepatitis-related parameters in serum samples after long-term storage. Biopreserv Biobanking 2017;15:211–9, https://doi.org/10.1089/bio.2016.0043.Search in Google Scholar PubMed

27. Drake, SK, Bowen, RA, Remaley, AT, Hortin, GL. Potential interferences from blood collection tubes in mass spectrometric analyses of serum polypeptides. Clin Chem 2004;50:2398–401, https://doi.org/10.1373/clinchem.2004.040303.Search in Google Scholar PubMed

28. Enroth, S, Hallmans, G, Gyllensten, U. Effects of long-term storage time and original sampling month on biobank plasma protein concentrations. EBioMedicine 2016;12:309–14, https://doi.org/10.1016/j.ebiom.2016.08.038.Search in Google Scholar PubMed PubMed Central

29. Flores, CFY, de Las Mercedes Hurtado Pineda, A, Bonilla, VMC, Saenz-Flor, K. Sample management: stability of plasma and serum on different storage conditions. EJIFCC 2020;31:46–55.Search in Google Scholar

30. Fokkema, MR, Herrmann, Z, Muskiet, FA, Moecks, J. Reference change values for brain natriuretic peptides revisited. Clin Chem 2006;52:1602–3, https://doi.org/10.1373/clinchem.2006.069369.Search in Google Scholar PubMed

31. Paal, M, Habler, K, Vogeser, M. Estimation of inter-laboratory reference change values from external quality assessment data. Biochem Med 2021;31:030902, https://doi.org/10.11613/bm.2021.030902.Search in Google Scholar PubMed PubMed Central


Supplementary Material

The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2022-0063).


Received: 2022-01-22
Accepted: 2022-04-11
Published Online: 2022-04-26
Published in Print: 2022-06-27

© 2022 Walter de Gruyter GmbH, Berlin/Boston

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. The never-ending quest for antibody assays standardization and appropriate measurement units
  4. Review
  5. Glycated albumin in diabetes mellitus: a meta-analysis of diagnostic test accuracy
  6. Opinion Paper
  7. Ad interim recommendations for diagnosing SARS-CoV-2 infection by the IFCC SARS-CoV-2 variants working group
  8. Guidelines and Recommendations
  9. Recommendations for IVDR compliant in-house software development in clinical practice: a how-to paper with three use cases
  10. General Clinical Chemistry and Laboratory Medicine
  11. Comparison between polynomial regression and weighted least squares regression analysis for verification of analytical measurement range
  12. Transport stability profiling – a proposed generic protocol
  13. Impact of ultra-low temperature long-term storage on the preanalytical variability of twenty-one common biochemical analytes
  14. Development of a liquid chromatography mass spectrometry method for the determination of vitamin K1, menaquinone-4, menaquinone-7 and vitamin K1-2,3 epoxide in serum of individuals without vitamin K supplements
  15. TSH-receptor autoantibodies in patients with chronic thyroiditis and hypothyroidism
  16. Evaluation of the AFIAS-1 thyroid-stimulating hormone point of care test and comparison with laboratory-based devices
  17. Lack of analytical interference of dydrogesterone in progesterone immunoassays
  18. A comprehensive comparison between ISAC and ALEX2 multiplex test systems
  19. Detection of subarachnoid haemorrhage with spectrophotometry of cerebrospinal fluid – a comparison of two methods
  20. Importance of cerebrospinal fluid storage conditions for the Alzheimer’s disease diagnostics on an automated platform
  21. Estimating urine albumin to creatinine ratio from protein to creatinine ratio using same day measurement: validation of equations
  22. An automated, rapid fluorescent immunoassay to quantify serum soluble programmed death-1 (PD-1) protein using testing-on-a-probe biosensors
  23. Hematology and Coagulation
  24. The VES-Matic 5 system: performance of a novel instrument for measuring erythrocyte sedimentation rate
  25. Cancer Diagnostics
  26. Is thyroglobulin a reliable biomarker of differentiated thyroid cancer in patients treated by lobectomy? A systematic review and meta-analysis
  27. Cardiovascular Diseases
  28. The intra-individual variation of cardiac troponin I: the effects of sex, age, climatic season, and time between samples
  29. Infectious Diseases
  30. A cohort analysis of SARS-CoV-2 anti-spike protein receptor binding domain (RBD) IgG levels and neutralizing antibodies in fully vaccinated healthcare workers
  31. Patients with severe COVID-19 do not have elevated autoantibodies against common diagnostic autoantigens
  32. Cerebrospinal fluid markers of inflammation and brain injury in Lyme neuroborreliosis – a prospective follow-up study
  33. Letters to the Editors
  34. Misleading nomenclature of units of WHO materials used for standardization of SARS COv-2 serology
  35. Assessment of the humoral response in Omicron breakthrough cases in healthcare workers who received the BNT162b2 booster
  36. COVID-19 related mortality and religious denomination vs. genetics
  37. Unexpectedly low tacrolimus concentrations attributed to inappropriately labeled water container from the instrument manufacturer
  38. Effects of different pepsinogen cut offs in the screening of apparently healthy people
  39. Low levels of G17 and Barrett esophagus: a clinical relationship
  40. (In)direct chloride ISE measurements, room for improvement
  41. New concept of interactive academic poster used in medical scientific congress presented in the EuroMedLab 2021 Munich Session
Downloaded on 23.3.2026 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2022-0063/html
Scroll to top button