Analytical interference by monoclonal immunoglobulins on the direct bilirubin AU Beckman Coulter assay: the benefit of unsuspected diagnosis from spurious results

  • Elena García-González , Ricardo González-Tarancón , Maite Aramendía and Luis Rello EMAIL logo
Published/Copyright: December 17, 2015

Abstract

Background: Monoclonal (M) components can interfere with the direct bilirubin (D-Bil) assay on the AU Beckman Coulter instrumentation and produce spurious results, such as D-Bil values greater than total bilirubin (T-Bil) or very low/negative D-Bil values. If properly detected, this interference may uncover undiagnosed patients with monoclonal gammopathy (MG).

Methods: We investigated the interference rate on the D-Bil AU assay in serum samples known to contain M proteins along with their isotype and described the protocol set up in our laboratory to help with the diagnosis of MG based on D-Bil spurious results as first indication.

Results: During a period of 4 years, 15.4% (345 of 2235) of serum samples containing M immunoglobulins produced erroneous D-Bil results, although no clear relationship between the magnitude or isotype of the M component and interference could be found. In total 22 new patients were diagnosed with MG based on the analytical artefact with the D-Bil as first indication.

Conclusions: The D-Bil interference from MG on the Beckman AU analysers needs to be made known to laboratories in order to prevent clinical confusion and/or additional workup to explain the origin of anomalous results. Although this information may not add to the management of existing patients with serum paraproteins, it can benefit patients that have not been diagnosed with MG by triggering follow up testing to determine if M components are present.


Corresponding author: Luis Rello, Department of Clinical Biochemistry, “Miguel Servet” University Hospital, Paseo Isabel La Católica 1-3, 50009, Zaragoza, Spain, Phone: +34 976765544, Fax: +34 976765543, E-mail:

  1. Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

  2. Research funding: None declared.

  3. Employment or leadership: None declared.

  4. Honorarium: None declared.

  5. Competing interests: The funding organisation(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

References

1. Kroll MH, Elin RJ. Interference with clinical laboratory analyses. Clin Chem 1994;40:1996–2005.10.1093/clinchem/40.11.1996Search in Google Scholar

2. Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev 2004;25:105–20.Search in Google Scholar

3. Bakker AJ, Mücke M. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007;45:1240–3.10.1515/CCLM.2007.254Search in Google Scholar PubMed

4. Dalal BI, Brigden ML. Factitious biochemical measurements resulting from hematologic conditions. Am J Clin Pathol 2009;131:195–204.10.1309/AJCPY9RP5QYTYFWCSearch in Google Scholar PubMed

5. Pantanowitz L, Horowitz GL, Upalakalin JN, Beckwith BA. Artifactual hyperbilirubinemia due to paraprotein interference. Arch Pathol Lab Med 2003;127:55–9.10.5858/2003-127-55-AHDTPSearch in Google Scholar PubMed

6. Yang Y, Howanitz PJ, Howanitz JH, Gorfajn H, Wong K. Paraproteins are a common cause of interferences with automated chemistry methods. Arch Pathol Lab Med 2008;132:217–23.10.5858/2008-132-217-PAACCOSearch in Google Scholar PubMed

7. Nauti A, Barassi A, Merlini G, d’Eril GV. Paraprotein interference in an assay of conjugated bilirubin. Clin Chem 2005;51:1076–7.10.1373/clinchem.2005.048710Search in Google Scholar PubMed

8. Song L, Kelly KA, Butch AW. Monoclonal and polyclonal immunoglobulin interference in a conjugated bilirubin assay. Arch Pathol Lab Med 2014;138:950–4.10.5858/arpa.2013-0042-OASearch in Google Scholar PubMed

9. Seimiya M, Suzuki Y, Yoshida T, Sawabe Y, Matsushita K, Nomura F. The abnormal reaction data-detecting function of the automated biochemical analyzer was useful to prevent erroneous total-bilirubin measurement and to identify monoclonal proteins. Clin Chim Acta 2015;441:44–6.10.1016/j.cca.2014.12.013Search in Google Scholar PubMed

10. Jelinek AG, Bachmann LM. Unexpected test results in a patient with multiple myeloma. Clin Chem 2014;60:1375–8.10.1373/clinchem.2013.213884Search in Google Scholar PubMed

11. Wadhera RK, Rajkumar SV. Prevalence of monoclonal gammopathy of undetermined significance: a systematic review. Mayo Clin Proc 2010;85:933–42.10.4065/mcp.2010.0337Search in Google Scholar PubMed PubMed Central

12. Kyle RA, Rajkumar SV. Monoclonal gammopathies of undetermined significance. Best Pract Res Clin Haematol 2005;18:689–707.10.1016/j.beha.2005.01.025Search in Google Scholar PubMed

13. Monk C, Wallage M, Wassell J, Whiteway A, James J, Beetham R. A monoclonal protein identified by an anomalous lipaemia index. Ann Clin Biochem 2009;46:250–2.10.1258/acb.2008.008192Search in Google Scholar PubMed

Received: 2015-6-28
Accepted: 2015-11-11
Published Online: 2015-12-17
Published in Print: 2016-8-1

©2016 by De Gruyter

Articles in the same Issue

  1. Frontmatter
  2. Editorials
  3. Laboratory analytical quality – the process continues
  4. Measurement uncertainty – a revised understanding of its calculation and use
  5. Review
  6. Substrate-zymography: a still worthwhile method for gelatinases analysis in biological samples
  7. Opinion Papers
  8. Is the combination of trueness and precision in one expression meaningful? On the use of total error and uncertainty in clinical chemistry
  9. The problem with total error models in establishing performance specifications and a simple remedy
  10. Measurement uncertainty for clinical laboratories – a revision of the concept
  11. Uncertainty in measurement and total error – are they so incompatible?
  12. General Clinical Chemistry and Laboratory Medicine
  13. Using the hazard ratio to evaluate allowable total error in predictive measurands
  14. Performance of electrolyte measurements assessed by a trueness verification program
  15. Analytical interference by monoclonal immunoglobulins on the direct bilirubin AU Beckman Coulter assay: the benefit of unsuspected diagnosis from spurious results
  16. Preliminary probe of quality indicators and quality specification in total testing process in 5753 laboratories in China
  17. Analytical and clinical evaluation of the new Fujirebio Lumipulse®G non-competitive assay for 25(OH)-vitamin D and three immunoassays for 25(OH)D in healthy subjects, osteoporotic patients, third trimester pregnant women, healthy African subjects, hemodialyzed and intensive care patients
  18. Patient-performed extraction of faecal calprotectin
  19. Comparison of the clinical utility of the Elia CTD Screen to indirect immunofluorescence on Hep-2 cells
  20. Reference Values and Biological Variations
  21. Sex-related differences in the association of ghrelin levels with obesity in adolescents
  22. Gestation specific reference intervals for thyroid function tests in pregnancy
  23. Cancer Diagnostics
  24. SOX17 promoter methylation in plasma circulating tumor DNA of patients with non-small cell lung cancer
  25. Dopamine concentration in blood platelets is elevated in patients with head and neck paragangliomas
  26. Letters to the Editor
  27. Cancer dynamics and the success of cancer screening programs
  28. Mother’s instinct – a rare case of multiple test interferences due to heterophile antibodies
  29. Sigma metric or defects per million opportunities (DPMO): the performance of clinical laboratories should be evaluated by the Sigma metrics at decimal level with DPMOs
  30. A national survey of preanalytical handling of oral glucose tolerance tests in pregnancy
  31. Updating pregnancy diabetes guidelines: is (y)our laboratory ready?
  32. Low serum bilirubin values are associated with pulmonary embolism in a case-control study
  33. Effect of Hb H on HbA1c measurements as measured by IFCC reference method and affinity HPLC
  34. Adipocytes in venipunctures cause falsely elevated S-100B serum values
  35. Earlier detection of sepsis by Candida parapsilosis using three-dimensional cytographic anomalies on the Mindray BC-6800 hematological analyzer
  36. Theranos phenomenon – part 4: Theranos at an International Conference
Downloaded on 7.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2015-0608/pdf
Scroll to top button