A new statistical method for evaluating long-term analytical performance of laboratories applied to an external quality assessment scheme for flow cytometry
-
Wim Coucke
Abstract
Background: The Belgian External Quality Assessment Scheme for Flow Cytometry evaluates the long-term analytical performance of participating laboratories by calculating a regression line between the target and reported values of each parameter for each laboratory during the past 3 years. This study aims to develop a method to find laboratories with aberrant variability or bias using robust techniques and to obtain robust estimates of the variability.
Methods: A method is proposed to find outliers with respect to the individual regression line, followed by a step to find regression lines with excessive variability and finally a step to find regression lines with high bias.
Results: The model was applied to the results obtained by 52 laboratories for CD4%. From the 1340 data points, 35 were determined to be regression outliers. The second step revealed one regression line with excessive variability; the third step detected three regression lines with exceeding bias.
Conclusions: The methodology allows assessment of the long-term performance of laboratories, taking into account samples with different target values. Outliers in the first step indicate accidental mistakes, outliers in the second and third step point to high analytical variability or bias.
Clin Chem Lab Med 2010;48:645–50.
©2010 by Walter de Gruyter Berlin New York
Articles in the same Issue
- Editorials
- Laboratory medicine and natural disasters: are we ready for the challenge?
- Reference values: still a critical issue in laboratory medicine
- Review
- Contemporary platelet function testing
- Minireviews
- Disasters: where they find us
- Inter-laboratory variation in cerebrospinal fluid biomarkers for Alzheimer's disease: united we stand, divided we fall
- Opinion Paper
- Glycated hemoglobin (HbA1c): old dogmas, a new perspective?
- Guidelines and Recommendations
- IFCC reference procedures for measurement of the catalytic concentrations of enzymes: corrigendum, notes and useful advice
- Recommendations for the implementation of international standardization of glycated hemoglobin in Italy
- Genetics and Molecular Diagnostics
- UGT2B17 and SULT1A1 gene copy number variation (CNV) detection by LabChip microfluidic technology
- Genetic polymorphisms of CYP1A2, CYP3A4, CYP3A5, pregnane/steroid X receptor and constitutive androstane receptor in 207 healthy Spanish volunteers
- Automated genomic DNA extraction from saliva using the QIAxtractor
- General Clinical Chemistry and Laboratory Medicine
- A new statistical method for evaluating long-term analytical performance of laboratories applied to an external quality assessment scheme for flow cytometry
- Evaluation of the Becton-Dickinson rapid serum tube: does it provide a suitable alternative to lithium heparin plasma tubes?
- Immunoglobulin M antibodies to aquaporin-4 in neuromyelitis optica and related disorders
- Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids
- Extracellular haemoglobin, oxidative stress and quality of red blood cells relative to perioperative blood salvage
- Positive predictive value of serological diagnostic measures in celiac disease
- Retinol potentiates the inhibitory effect of ascorbic acid on uric acid assay
- Reference Values and Biological Variations
- Reference intervals for brain natriuretic peptide in healthy newborns and infants measured with an automated immunoassay platform
- Factors influencing serum total tryptase concentrations in a general adult population
- Leptin and adiponectin levels in pubertal children: relationship with anthropometric variables and body composition
- Cardiovascular risk factors and γ-glutamyltransferase fractions in healthy individuals
- Cancer Diagnostics
- Combined analysis of survivin autoantibody and carcinoembryonic antigen biomarkers for improved detection of colorectal cancer
- Oligoclonal bands and immunoglobulin isotype switch during monitoring of patients with multiple myeloma and autologous hematopoietic cell transplantation: a 16-year experience
- Letters to the Editor
- Methodological weakness in using correlation coefficients for assessing the interchangeability of analyte data between samples collected under different sampling conditions – the example of matrix metalloproteinase 9 determined in serum and plasma samples
- Intralipid causes falsely increased glucose concentrations with the Hemocue glucose analyzer
- Serum cholesterol: required and actual bias values
- Abstracts
- ISMD2010 Eighth International Symposium on Molecular Diagnostics
Articles in the same Issue
- Editorials
- Laboratory medicine and natural disasters: are we ready for the challenge?
- Reference values: still a critical issue in laboratory medicine
- Review
- Contemporary platelet function testing
- Minireviews
- Disasters: where they find us
- Inter-laboratory variation in cerebrospinal fluid biomarkers for Alzheimer's disease: united we stand, divided we fall
- Opinion Paper
- Glycated hemoglobin (HbA1c): old dogmas, a new perspective?
- Guidelines and Recommendations
- IFCC reference procedures for measurement of the catalytic concentrations of enzymes: corrigendum, notes and useful advice
- Recommendations for the implementation of international standardization of glycated hemoglobin in Italy
- Genetics and Molecular Diagnostics
- UGT2B17 and SULT1A1 gene copy number variation (CNV) detection by LabChip microfluidic technology
- Genetic polymorphisms of CYP1A2, CYP3A4, CYP3A5, pregnane/steroid X receptor and constitutive androstane receptor in 207 healthy Spanish volunteers
- Automated genomic DNA extraction from saliva using the QIAxtractor
- General Clinical Chemistry and Laboratory Medicine
- A new statistical method for evaluating long-term analytical performance of laboratories applied to an external quality assessment scheme for flow cytometry
- Evaluation of the Becton-Dickinson rapid serum tube: does it provide a suitable alternative to lithium heparin plasma tubes?
- Immunoglobulin M antibodies to aquaporin-4 in neuromyelitis optica and related disorders
- Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids
- Extracellular haemoglobin, oxidative stress and quality of red blood cells relative to perioperative blood salvage
- Positive predictive value of serological diagnostic measures in celiac disease
- Retinol potentiates the inhibitory effect of ascorbic acid on uric acid assay
- Reference Values and Biological Variations
- Reference intervals for brain natriuretic peptide in healthy newborns and infants measured with an automated immunoassay platform
- Factors influencing serum total tryptase concentrations in a general adult population
- Leptin and adiponectin levels in pubertal children: relationship with anthropometric variables and body composition
- Cardiovascular risk factors and γ-glutamyltransferase fractions in healthy individuals
- Cancer Diagnostics
- Combined analysis of survivin autoantibody and carcinoembryonic antigen biomarkers for improved detection of colorectal cancer
- Oligoclonal bands and immunoglobulin isotype switch during monitoring of patients with multiple myeloma and autologous hematopoietic cell transplantation: a 16-year experience
- Letters to the Editor
- Methodological weakness in using correlation coefficients for assessing the interchangeability of analyte data between samples collected under different sampling conditions – the example of matrix metalloproteinase 9 determined in serum and plasma samples
- Intralipid causes falsely increased glucose concentrations with the Hemocue glucose analyzer
- Serum cholesterol: required and actual bias values
- Abstracts
- ISMD2010 Eighth International Symposium on Molecular Diagnostics