Analytical validation of serum bone alkaline phosphatase (BAP OSTASE) on Liaison

  • Etienne Cavalier , Eric Rozet , Agnès Carlisi , Anne-Catherine Bekaert , Olivier Rousselle , Philippe Hubert , Jean-Paul Chapelle and Pierre Delanaye
Published/Copyright: November 30, 2009

Abstract

Background: The goal of this study was to validate the DiaSorin Liaison BAP OSTASE, a new method for measurement of bone alkaline phosphatase (BAP), and to compare this method with the Beckman-Coulter Access Ostase. We also wanted to establish the reference range for BAP in adults and children.

Methods: We determined the precision, functional sensitivity, recovery, linearity and measurement uncertainty, accuracy profile and β-expectation limits. We defined an adult reference interval using individuals with 25-OH vitamin D >80 nmol/L, parathormone <58 ng/L, and normal calcium, phosphorous and estimated glomerular filtration rate. Each adult subclass (men/non-menopausal women/menopause women) contained 120 individuals. We also determined the 2.5th and 97.5th percentiles from a population of 450 children, stratified according to age and gender.

Results: The results of the validation showed: precision <6%, functional sensitivity <0.74 μg/L, mean recovery 98.8±4.2% and good linearity. Relative uncertainty ranged from 9.0% to 12.9%, and the risk of one result falling out of the ±15% acceptance limits was <5% for concentrations between 7 and 94 μg/L. The Bland-Altman plot showed no systematic bias between the two methods. In adults, we did not find any statistical difference between the different subclasses. The upper limit of normality observed in the entire population (n=360) was 21.3 μg/L (90% CI: 18.3–24.2 μg/L).

Conclusions: The Liaison BAP OSTASE is a robust method, and is completely validated between 7 and 93 μg/L: in this range, 95% of the values obtained will be within ±15% of the true value.

Clin Chem Lab Med 2010;48:67–72.


Corresponding author: Dr. Etienne Cavalier, Department of Clinical Chemistry, University Hospital of Liège, Domaine du Sart-Tilman, 4000 Liège, Belgium Phone: +32 4 3667692, Fax: +32 4 3667691,

Received: 2009-8-1
Accepted: 2009-9-2
Published Online: 2009-11-30
Published in Print: 2010-01-01

©2010 by Walter de Gruyter Berlin New York

Articles in the same Issue

  1. Editorials
  2. Rapid diagnosis of swine-origin influenza A (H1N1) virus (S-OIV): useful or hype?
  3. Application of traceability concepts to analytical quality control may reconcile total error with uncertainty of measurement
  4. Reviews
  5. Pandemic influenza A (H1N1) 2009: the experience of the first six months
  6. The role of anti-core antibody response in the detection of occult hepatitis B virus infection
  7. Opinion Papers
  8. Managing quality vs. measuring uncertainty in the medical laboratory
  9. Collective opinion paper on findings of the 2009 convocation of experts on quality control
  10. General Clinical Chemistry and Laboratory Medicine
  11. Diagnostic performance of rapid influenza antigen assays in patients infected with the new influenza A (H1N1) virus
  12. Achieving continuous improvement in laboratory organization through performance measurements: a seven-year experience
  13. Evaluation of errors in a clinical laboratory: a one-year experience
  14. Validation and Outcome Studies
  15. Analytical validation of serum bone alkaline phosphatase (BAP OSTASE) on Liaison
  16. Analytical performance of the serum free light chain assay
  17. A study on the stability of urinary free catecholamines and free methyl-derivatives at different pH, temperature and time of storage
  18. Plasma nitrite/nitrate concentrations in patients with schizophrenia
  19. Inverse relationship between serum high density lipoprotein and negative syndrome in antipsychotic-naive schizophrenia
  20. Reference Values and Biological Variations
  21. Biological variation of free plasma amino acids in healthy individuals
  22. Differences in serum potassium concentrations between Chinese, Indians and Malays
  23. Determination of median levels of the free β subunit of human chorionic gonadotropin in women from mainland China using a new time-resolved fluoroimmunoassay
  24. Age- and gender-specific reference ranges for serum insulin-like growth factor I (IGF-I) and IGF-binding protein-3 concentrations on the Immulite 2500: results of the Study of Health in Pomerania (SHIP)
  25. Cardiovascular Diseases
  26. Improvement of the cardiac marker N-terminal-pro brain natriuretic peptide through adjustment for renal function: a stratified multicenter trial
  27. Plasma homocysteine, apolipoprotein E status and vascular disease in elderly patients with mental illness
  28. Letters to the Editor
  29. Serum total and free sialic acid in patients with chronic liver disease
  30. Measurement of nitrotyrosine in plasma by immunoassays is fraught with danger: commercial availability is no guarantee of analytical reliability
  31. Problems with the determination of nitrotyrosine in human plasma are not solved in a satisfactory way and conclusions from clinical trials need to be carefully revised
Downloaded on 6.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2010.019/html
Scroll to top button