Multicentre performance evaluation of the E170 Module for MODULAR ANALYTICS

  • Christian Bieglmayer , Daniel W. Chan , Lori Sokoll , Roland Imdahl , Masaji Kobayashi , Erike Yamada , Diana J. Lilje , Hilmar Luthe , Jochen Meissner , Gianni Messeri , Alessandra Celli , Paola Tozzi , Heinz-Jürgen Roth , Frank-Peter Schmidt , Marie-Luise Mächler , Peter Schuff-Werner , Christiana Zingler , Johan Smitz , Johan Schiettecatte , Dieter J. Vonderschmitt , Patrick Pei , Katherine Ng , Christoph Ebert , Peter Kirch , Michael Wanger , Margaret McGovern , Wolfgang Stockmann and Albert Kunst
Published/Copyright: September 21, 2011

Abstract

The E170 module was evaluated at 13 sites in an international multicentre study. The objective of the study was to assess the analytical performance of 49 analytes, and to collect feedback on the system’s reliability and practicability. The typical, within-run coefficients of variation (CVs) for most of the quantitative assays ranged between 1 and 2% while a range of 2–4% was achieved with the infectious disease methods. Total precision CVs were found to be within the manufacturer’s expected performance ranges, demonstrating good concordance of the system’s measuring channels and a high reproducibility during the 2–4-week trial period. The functional sensitivity of 11 selected assays met the clinical requirements (e.g., thyreotroponin (TSH) 0.008 mU/l, troponin T 0.02 µg/l, total prostate-specific antigen (PSA) 0.03 µg/l). The E170 showed no drift during an 8-hour period and no relevant reagent carryover. Accuracy was confirmed by ring trial experiments and method comparisons vs. Elecsys® 2010. The reliability and practicability of the system’s hardware and software met with, or even exceeded, the evaluator’s requirements. Workflow studies showed that E170 can cover the combined workload of various routine analysers in a variety of laboratory environment. Throughput and sample processing time requirements were achieved while personnel ‘hands-on-time’ could be reduced.

References

1 Stockmann W, McGovern M, Ng K. The analytical performance of the Roche/Hitachi MODULAR ANALYTICS [abstract]. Clin Chem Lab Med 1999; 37;Suppl:S400.Search in Google Scholar

2 Stockmann W, Engeldinger W, Kunst A, McGovern M, Ng K. Assessment of the Roche/Hitachi MODULAR ANALYTICS under simulated routine conditions in 16 European and US laboratories [abstract]. Clin Chem Lab Med 1999; 37;Suppl:S400.Search in Google Scholar

3 Neumeier D, Chan DW. Multicentre evaluation of the Boehringer Mannheim Elecsy® 2010 immunoassay system. Wien Klin Wochenschr 1998; 110:Suppl 3:5–68.Search in Google Scholar

4 Myers MB. Elecsys® immunoassay systems. In: Wild D, editor. The immunoassay handbook. London: Macmillan Reference Ltd., 2000:341–6.Search in Google Scholar

5 Guidelines for the evaluation of analysers in clinical chemistry. ECCLS document, 1986;3:no. 2.Search in Google Scholar

6 NCCLS evaluation protocols, ISBN-1-56238-146-6, Order Code SC 1, National Committee for Clinical Laboratory Standards. Villanova, PA, 1992.Search in Google Scholar

7 Bablok W, Stockmann W. An alternative approach to a system evaluation in the field [abstract]. Quim Clin 1995; 14:239.Search in Google Scholar

8 Bablok W, Barembruch R, Stockmann W, Brauer P, Graber P, Michel R, et al. CAEv – a program for computer aided evaluation. J Autom Chem 1991; 13:167–79.10.1155/S1463924691000317Search in Google Scholar PubMed PubMed Central

9 Passing H, Bablok W. A new biometrical procedure for testing the equality of measurements from two different analytical methods. J Clin Chem Clin Biochem 1983; 21:709–20.Search in Google Scholar

10 Stockmann W, Bablok W, Poppe W, Bayer PM, Keller F, Schweiger CR. Criteria of practicability. In: Haeckel H, editor. Evaluation methods in laboratory medicine. Weinheim: VCH, 1993:185–201.Search in Google Scholar

11 Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest 1999; 59:491–500.10.1080/00365519950185229Search in Google Scholar PubMed

12 Apple FS, Wu AHB. Myocardial infarction redefined: role of cardiac troponin testing. Clin Chem 2001; 47:377–9.10.1093/clinchem/47.3.377Search in Google Scholar

Received: 2004-3-19
Accepted: 2004-8-20
Published Online: 2011-9-21
Published in Print: 2004-10-1

© Walter de Gruyter

Articles in the same Issue

  1. The significance of serum γ-glutamyltransferase in cardiovascular diseases
  2. Anti-tissue transglutaminase antibodies in inflammatory bowel disease: new evidence
  3. Expression level of MDR1 message in peripheral blood leukocytes from healthy adults: a competitive nucleic acid sequence-based amplification assay for its determination
  4. Characteristics of prolinase against various iminodipeptides in erythrocyte lysates from a normal human and a patient with prolidase deficiency
  5. Alteration of homocysteine catabolism in pre-eclampsia, HELLP syndrome and placental insufficiency
  6. Immune system-mediated endothelial damage is associated with NO and antioxidant system disorders
  7. Bone marrow microvascular density and angiogenic growth factors in multiple myeloma
  8. The insulin-like growth factor system in the circulation of patients with viral infections
  9. Time course of systemic markers of inflammation in patients presenting with acute coronary syndromes
  10. Age- and sex-related reference values for serum insulin concentration and its biological determinants in a French healthy population. The STANISLAS cohort
  11. Potential pitfalls of comparative measurements of reticulocytes with flow cytometry and microscopy in prematures and infants
  12. Screening of antinuclear antibodies: comparison between enzyme immunoassay based on nuclear homogenates, purified or recombinant antigens and immunofluorescence assay
  13. Analytical and diagnostic accuracy of the EliA™ automated enzyme fluoroimmunoassay for antineutrophil cytoplasmic autoantibody detection
  14. A particle-enhanced turbidimetric immunoassay for quantitative determination of orosomucoid in urine: development, validation and reference values
  15. Analytical performance and clinical results of a fully automated MEIA system for brain natriuretic peptide assay: comparison with a point of care testing method
  16. Multicentre performance evaluation of the E170 Module for MODULAR ANALYTICS
  17. Congress of Clinical Chemistry and Laboratory Medicine Annual meeting of the German United Society for Clinical Chemistry and Laboratory Medicine (DGKL), Düsseldorf, Germany, November 22-24, 2004
Downloaded on 11.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2004.239/html
Scroll to top button