Evaluation of the blood gas analyzer GEM(r) PREMIER(tm) 3000

  • Bénédicte Bénéteau-Burnat , Marie-Chantal Bocque , Anne Lorin , Catherine Martin and Michel Vaubourdolle
Published/Copyright: June 1, 2005

Abstract

During the past few decades, new technologies have allowed the fabrication of miniaturized sensors and the development of analyzers well designed for point-of-care testing (POCT). They combined the ease-of-use and portability required for POC with the accuracy and the reliability of traditional systems. Instrumentation Laboratory introduced the GEM®Premier™3000, a compact and portable system to rapidly analyze pH, pCO2, pO2, Ca2+, Na2+, K+, lactate, glucose and hematocrit, from 135 μl of whole blood with a simple component: a cartridge, the GEM®Premier Pak. The analytical performance was compared with the Radiometer® ABL 725 instrument for blood gas/electrolytes/metabolites, and for hematocrit with the centrifuged microhematocrit method. Aqueous quality control materials (Instrumentation Laboratory) were used for precision and recovery studies. The evaluation was performed according to the Valtec protocol, designed by the French Society of Clinical Biology, for imprecision, linearity and accuracy. Within-run and between-run imprecisions for all the parameters gave good CV values. Linearity was good and results comparison showed close agreement with correlation coefficients between 0.91 and 0.99. The analyzer is very easy to use; the simplicity and the convenience of only one consumable make the GEM®Premier™ 3000 very suitable for a STAT laboratory or for POCT.

Published Online: 2005-6-1
Published in Print: 2004-2-16

Copyright (c) 2004 by Walter de Gruyter GmbH & Co. KG

Articles in the same Issue

  1. International collaboration in clinical chemistry and laboratory medicine: the Human Proteome Organization (HUPO) Plasma Proteome Project
  2. Current concepts in standardization of cardiac marker immunoassays
  3. Collagen, ageing and nutrition
  4. Evaluation of a new efficient procedure for single-nucleotide polymorphism genotyping: tetra-primer amplification refractory mutation system-polymerase chain reaction
  5. High prevalence of non-HFE gene-associated haemochromatosis in patients from southern Italy
  6. Limited usefulness of the PFA-100™ for the monitoring of ADP receptor antagonists–in vitro experience
  7. Isolation of plasma small-dense low-density lipoprotein using a simple air-driven ultracentrifuge and quantification using immunassay of apolipoprotein B
  8. Analytical performance and diagnostic accuracy of a fully-automated electrochemiluminescent assay for the N-terminal fragment of the pro-peptide of brain natriuretic peptide in patients with cardiomyopathy: comparison with immunoradiometric assay methods for brain natriuretic peptide and atrial natriuretic peptide
  9. Calcium oxalate crystallization in untreated urine, centrifuged and filtered urine and ultrafiltered urine
  10. Matrix effects in clinical immunoassays and the effect of preheating and cooling analytical samples
  11. Blood alcohol testing: comparison of the performance obtained with the different methods used in the Belgian external quality assessment schemes
  12. Determination of lipase catalytic activity in two reference materials: BCR 693 and BCR 694 by titrimetry at constant pH
  13. Evaluation of a new generation of plastic evacuated blood-collection tubes in clinical chemistry, therapeutic drug monitoring, hormone and trace metal analysis
  14. Association between the CYP2C9 polymorphism and the drug metabolism phenotype
  15. Analytical goals for coagulation tests based on biological variation
  16. In vivo total antioxidant capacity: comparison of two different analytical methods
  17. Performance evaluation of automated assays for β-CrossLaps, N-MID-Osteocalcin and intact parathyroidhormone (BIOROSE Multicenter Study)
  18. Evaluation of the blood gas analyzer GEM(r) PREMIER(tm) 3000
  19. Lactate dehydrogenase (LD) extra isoenzyme electrophoretic band between LD1 and LD2 caused by a complex with α1-lipoprotein. A case report
  20. Relationship between cryoglobulins and hepatitis C virus (HCV) core antigen or antibody titers
  21. Spurious hyperphosphatemia due to sample contamination with heparinized saline from an indwelling catheter
  22. Biological variation of plasma chromogranin A
  23. Verification, validation and evaluation of analytical procedures in laboratory medicine
  24. DACH-LIGA Homocystein (German, Austrian and Swiss Homocysteine Society): Consensus Paper on the Rational Clinical Use of Homocysteine, Folic Acid and B-Vitamins in Cardiovascular and Thrombotic Diseases: Guidelines and Recommendations
  25. Meetings and Awards
Downloaded on 13.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2004.018/html
Scroll to top button