Abstract

Background: We demonstrated that lyophilised EQA/PT control materials, under certain circumstances, provide equal information about bias values as pools of native patient sera, and that in some cases, long-term reliable work with such materials is possible.

Methods: Bias values, estimated from results of routine surveys of EQA SEKK (Czech Republic) programmes for eight basic blood serum analytes using lyophilised control materials, were compared with bias values reached in the CAP (USA) programme where pools of native and lyophilised patient sera were used.

Results: Results for the components Na, K, Mg, Cl, P, urea, glucose, and uric acid were assessed. No significant differences were found between the bias values estimated by CAP using native sera, and those estimated by SEKK using lyophilised sera.

Conclusions: It can be concluded that well-prepared lyophilised control materials may show the required commutability level. If so, they constitute a practical and cost-efficient alternative to native or fresh frozen sera when assessing bias.

Clin Chem Lab Med 2009;47:489–90.


Corresponding author: Josef Kratochvila, SEKK Pardubice, Bartolomejska 90, 53002 Pardubice, Czech Republic Phone: +420-325516388, Fax: +420-325516388,

Received: 2008-9-11
Accepted: 2009-1-9
Published Online: 2009-03-06
Published in Print: 2009-04-01

©2009 by Walter de Gruyter Berlin New York

Articles in the same Issue

  1. Minireview
  2. Risk loci for type 2 diabetes – Quo vadis?
  3. Genetics and Molecular Diagnostics
  4. A common variant in the FTO gene is associated with body mass index in males and postmenopausal females but not in premenopausal females. Czech post-MONICA and 3PMFs studies
  5. Genotype distribution of estrogen receptor α polymorphisms in pregnant women from healthy and preeclampsia populations and its relation to blood pressure levels
  6. Do common genetic variants in endotoxin signaling pathway contribute to predisposition to alcoholic liver cirrhosis?
  7. The selection and application of ssDNA aptamers against MPT64 protein in Mycobacterium tuberculosis
  8. Stromal cell-derived factor-1 but not its receptor, CXCR4, gene variants increase susceptibility and pathological development of hepatocellular carcinoma
  9. Expression of lineage markers using real-time quantitative polymerase chain reaction (RT-qPCR) in normal and in leukemia bone marrow
  10. Translation termination and protein folding pathway genes are not correlated in gastric cancer
  11. General Clincial Chemistry and Laboratory Medicine
  12. Novel serum paraoxonase activity assays are associated with coronary artery disease
  13. Elevated erythrocyte carbonic anhydrase activity is a novel clinical marker in hyperventilation syndrome
  14. Elevation of the glycoxidation product Nε-(carboxymethyl)lysine in patients presenting with acute myocardial infarction
  15. Lipoprotein(a) level and its association with tumor stage in male patients with primary lung cancer
  16. A high-performance liquid chromatography method for the determination of carbamazepine and carbamazepine-10,11-epoxide and its comparison with chemiluminescent immunoassay
  17. Reference Values and Biological Variations
  18. Minimizing the effects of multicollinearity in the polynomial regression of age relationships and sex differences in serum levels of pregnenolone sulfate in healthy subjects
  19. Determination of physiological plasma pentraxin 3 (PTX3) levels in healthy populations
  20. Validation and Outcome Studies
  21. Analytical precision of the Urolizer® for the determination of the BONN-Risk-Index (BRI) for calcium oxalate urolithiasis and evaluation of the influence of 24-h urine storage at moderate temperatures on BRI
  22. Determination of nitrotyrosine concentrations in plasma samples of diabetes mellitus patients by four different immunoassays leads to contradictive results and disqualifies the majority of the tests
  23. Estimation of trueness of measurement results obtained in external quality assessment
  24. Letters to the Editor
  25. A new PCR-RFLP assay for –1123 G>C polymorphism in the PTPN22 gene: allele and genotype frequencies in a western Mexican population
  26. Proposed classification of various limit values (guide values) used in assisting the interpretation of quantitative laboratory test results
  27. Case report: Over-substitution of thyroxine due to interference in serum thyroid-stimulating hormone measurement
  28. Rv2629 191A/C nucleotide change is not associated with rifampicin resistance in Mycobacterium tuberculosis
Downloaded on 5.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2009.098/html
Scroll to top button