Home The effect of automated hemolysis index measurement on sample and test rejection rates
Article Publicly Available

The effect of automated hemolysis index measurement on sample and test rejection rates

  • Fazıla Atakan Erkal ORCID logo EMAIL logo , Güzin Aykal , Hayriye Melek Yalçınkaya , Nihal Aksoy and Murat Özdemir
Published/Copyright: January 26, 2019

Abstract

Objective

Vast majority of laboratory errors occurs in preanalytical phase and in vitro hemolysis is the most common among preanalytical errors. Automated serum index measurement is being used in routine biochemical analysis in Antalya Public Health Care Laboratory, since June 2014. Our aim in this study is to reveal the impact of serum index usage on rejected samples and rejected test rates due to hemolysis.

Materials and methods

Hemolysis, icterus and lipemia (HIL) spectral interference reagent and program have been used in our laboratory since June 2014. In the current study, the number of samples and tests that were rejected due to hemolysis in June–August 2014 were compared with those rejected in the same period of 2013.

Results

In 2014, the sample rejection rate was 2.53% and the rejected test rate was 0.48%. In 2013, the sample rejection rate was 0.56% and the rejected test rate was 0.55%. When compared two periods, statistically significant increase in rejected sample number due to hemolysis in 2014 is result of, visually undetectable hemolyzed samples previously can be identified by HIL method (p<0.05).

Conclusion

Usage of hemolysis index program in automated systems for detecting hemolysis was evaluated as a method which is standardized, semi-quantitative, with high reproducibility and allows test based rejection.

Öz

Amaç

Laboratuvar hatalarının büyük kısmı preanalitik fazda oluşmakta ve in vitro hemoliz en sık rastlanan preanalitik hatalar arasında yer almaktadır. Haziran 2014 tarihinden itibaren, Antalya Halk Sağlığı Laboratuvarında rutin biyokimyasal analizlerde otomatize serum indeks ölçümü kullanılmaya başlanmıştır. Bu çalışmanın amacı serum indeks kullanımının hemoliz nedeniyle reddedilen numuneler ve testlere etkisini belirlemektir.

Gereç ve Yöntemler

Laboratuvarımızda Haziran 2014 tarihinden itibaren Hemoliz, İkter, Lipemi (HIL) spektral interferans reaktifi ve programı kullanılmaya başlanmıştır. Mevcut çalışmada 2014 yılı Haziran-Ağustos aylarında hemoliz nedeniyle reddedilen numune ve test sayıları 2013 yılı aynı döneminde reddedilenler ile karşılaştırıldı.

Sonuçlar

2014 yılında numune red oranı %2.53 ve reddedilen test oranı %0.48 olarak saptanmıştır. 2013 yılında ise numune red oranı %0.56 ve reddedilen test oranı %0.55 olarak bulunmuştur. İki dönem karşılaştırıldığında 2014 yılında hemoliz nedeniyle reddedilen numune sayısındaki istatistiksel olarak anlamlı artışın (p<0.05) nedeni daha önce gözle tespit edilemeyen hemolizli numunelerin HIL yöntemi ile saptanması olarak değerlendirilmiştir.

Tartışma

Otomatize biyokimya sistemlerinde hemolizin saptanması amacıyla, hemoliz indeks programının kullanılması, standardize, semi-kantitatif, tekrarlanabilirliği yüksek ve test bazında numune reddini sağlayan bir yöntem olarak değerlendirilmiştir.

Introduction

Laboratory test results affect clinical decision-making, patient safety and outcomes [1]. Therefore, data given by laboratories have to be accurate, reliable, cost-effective and time-saving. Preanalytical errors account for up to 60% of total testing process and among preanalytical errors hemolyzed samples are the most frequent ones, approximately 40–70% of the unsuitable samples [2], [3], [4]. Hemolysis is pathological break down of red blood cells (RBC’s) and the release of hemoglobin and other components of RBC’s into serum or plasma, which can occur either in vitro or in vivo [5]. In vitro hemolysis may be result of inappropriate blood collection, specimen handling, processing, storage and transportation [6] and referred as a preanalytical quality marker [7].

Hemolysis can interfere many laboratory tests and alter test results via biological, chemical or spectrophotometric interference [8], [9]. Therefore, it is essential to observe and identify hemolyzed samples for accurate results which are traditionally noticed visually by laboratory staff after centrifugation [10]. But many studies revealed that visual determination of hemolysis is unreliable, subjective and highly variable according to laboratory staff judgment. Automated measurement of serum indices are one of the technological interventions introduced to laboratory use by manufacturers. Serum indices also called as hemolysis, icterus and lipemia (HIL) indices on clinical chemistry analyzers measures free hemoglobin, bilirubin and lipemia semi-quantitative or quantitatively. The calculation of HIL index is easy, cost effective, fast and reliable especially identifying hemolysis [11], [12], [13], [14]. Thus, automated integration of HIL parameters are recommended to laboratories by Clinical and Laboratory Standards Institute in CLSI C56A guideline [15].

In this study we intended to investigate the impact of hemolysis index implementation on sample rejection rates and suppressed test results in current clinical chemistry analyzers.

Materials and methods

This retrospective study was performed in Antalya Public Health Care Laboratory which serves to family practitioners in Antalya in 194 distinct centers. As a routine procedure samples are collected by family health staff into 5 mL serum gel separator tubes with clot activator (BD Vacutainer; Becton Dickinson and Co., Franklin Lakes, NJ, USA). Family Health Centers (FHC) in Antalya province can be categorized by means of location as central and district FHCs. Samples are collected by family health staff between 8:30 and 10:30 in all FHCs, centrifuged in the FHC and collected by couriers once a day to the Health Care Laboratory. Delivery time of samples from central FHCs to laboratory after sample collection ranges from 15 min (as the nearest FHC is in the same building with laboratory) to 120 min (furthest FHC is 65 km to laboratory) with a mean of 90 min and 35 km. From district FHCs delivery time ranges ranges from 60 min (the nearest FHC is 30 km away from the laboratory) to 240 min (furthest FHC is 270 km away from the laboratory) with a mean of 180 min and 120 km. So, preanalytical variabilities are highly difficult to control in public health care laboratories when compared with a hospital laboratory.

Being aware problems of visual determination of serum indices, in June 2014 automated serum index was adapted to clinical chemistry analyzers Beckman Coulter AU2700 (Beckman Coulter, Brea, CA, USA). Analyzers use HIL reagent to measure interfering substances with spectrophotometric measurement which are carried out at following wavelengths; lipemic judgment at 660 and 800 nm, icterus judgment at 480, 570, 600 and 800 nm and hemolytic judgment at 410, 480, 600 and 800 nm. By using formulas analyzer calculates interfering substance semi-quantitatively into one of these six categories; normal, one positive, two positive, three positive, four positive and five positive. Table 1 summarizes the warning flags of analyzer and their approximate concentration given by manufacturer information sheet.

Table 1:

Warning flags given by analyzer and their approximate concentration.

FlagHEM (mg/dL hemoglobin)ICT (mg/dL bilirubin)LIP (mg/dL intralipid)
No interference<50<2.5<40
+50–992.5–4.940–99
++100–1995.0–9.9100–199
+++200–29910.0–19.9200–299
++++300–50020–40300–500
+++++>500>40>500

Warning flags derived from analyzer were integrated to Laboratory Information System (LIS). Results of serum indices could be seen in a separate column among patient’s test results both by laboratory technicians and clinical biochemistry specialists.

Twenty-eight routine test are performed on clinical chemistry analyzers; alanine aminotransferaz (ALT), albumin, alkaline phosphatase, amylase, Antistreptolysin O, aspartat aminotransferase (AST), total bilirubin, direct bilirubin, blood urea nitrogen, calcium, cholesterol, chloride, C-reactive protein, creatine kinase, creatinine, gamma glutamyltransferase, glucose (Glu), HDL cholesterol, iron, iron binding capacity, lactate dehydrogenase, LDL cholesterol, magnesium, phosphate, potassium, rheumatoid factor, total protein, triglyceride, sodium, and uric acid.

According to the information sheet given by the manufacturer, tests influenced by hemolysis at different hemolysis levels are given in Table 2. Test results affected by hemolysis are rejected in agreement with the information sheet by clinical biochemistry specialists.

Table 2:

Affected tests at different hemolysis levels.

Hemolysis influenced parametersFlags
Hem 1Hem 2Hem 3Hem 4Hem 5
AST+++++
Tbil+++++
Dbil+++++
Fe+++++
LDH+++++
K+++++
ALP++++
CK++++
Mg++++
Amy+++
Cl+++
Na+++
TP+++
ALT+++
GGT+++
Chol++
Glucose++
P++
TG++
UIBC++
LDL-C++
All Tests+
  1. Hemolysis levels were defined according to Hb concentration (mg/dL) as following (+): Hem 1, 50–99; (++): Hem 2, 100–199; (+++): Hem 3, 200–299; (++++): Hem 4, 300–500; (+++++): Hem 5, >500.

In this study we evaluated data from LIS in terms of sample rejection rates and rejected test results 01 June–31 August 2013 which is before adapting hemolysis index and after 01 June–31 August 2014. Statistical analyses were performed using OpenEpi version 3.01 (www.openepi.com). In order to compare proportional differences between two periods chi-square test was used and p<0.05 considered as statistically significant.

Results

After implementation of HIL indices detection of hemolyzed sample increased by five-fold compared with visual detection. We found rejected sample rate increased (p<0.001) but rejected test rate decreased (p<0.001) (Table 3).

Table 3:

Recorded data of our laboratory for 3 months period before and after implementing automated serum indices.

Before hemolysis indexAfter hemolysis index
Method for detectionVisualAutomated
Total sample accepted52,61358,349
Total test accepted612,797636,228
Hemolyzed sample2941480
Rejected test33893075
Rejected test rate (%)0.550.48
Rejected sample rate (%)0.562.53

Within a total of 1480 rejected sample the most frequent group was found to be Hem 1 with 75.74% and rejected test rate 72.33% (Table 4).

Table 4:

Number and percentage of hemolysis index and rejected test after adapting automated serum indices.

FrequencyRejected test
Hem 11121 (75.74%)2224 (72.33%)
Hem 2308 (20.81%)622 (20.32%)
Hem 336 (2.43%)128 (4.16%)
Hem 415 (1.02%)101 (3.28%)
Total14803075

Most of the patients were female (65.61%) and when analyzed by means of age the most rejected samples belonged to 19–65 years population (Table 5).

Table 5:

Distribution of patients according to age and sex after implementing HIL indices.

AgeHem 1Hem 2Hem 3Hem 4Total hemolyzed sampleTotal sample accepted
MaleFemaleMaleFemaleMaleFemaleMaleFemaleMaleFemaleMaleFemale
0–1 years512001007268148
1–18 years466221143004708018912968
19–65 years299608681698194437980012,93635,638
65 years ↑396112221412538927701930
Total389732103205122451050997117,66540,684
34.4%65.6%30.3%69.7%

The most rejected test in Hem 1 and Hem 2 group was AST, Hem 3 group AST and ALT, Hem 4 group Glu (Table 6).

Table 6:

Distribution of rejected tests according to hemolysis degree after introducing serum indices.

Hemolysis influenced parametersHem 1Hem 2Hem 3Hem 4
AST8562372711
Tbil1565361
Dbil1574551
Fe56648198
LDH2035571
K2866951
ALP5562
CK710
Mg5353
Amy20
Cl51
Na51
TP10
ALT2711
GGT72
Chol12
Glucose14
P1
TG11
UIBC8
LDL-C12

Discussion

In this study our aim was to evaluate the effect of automated HIL usage on rejected sample rate and rejected test rate. Our results revealed that rejected sample rate was increased but rejected test rate was decreased.

Sample quality is one of the major quality indicators in laboratory diagnostics and for precise measurements factors affecting results, patient safety and clinical decision-making have to be identified and managed. Hemolysis, one of the most rejection motives, is recommended to be determined by automated analyzers because visual determination is considered unreliable and subjective [10], [16]. Automated chemistry analyzers can detect hemolysis with increased productivity, high throughput and decreased error rates. Also, it improves detection of mildly hemolyzed samples which have free hemoglobin levels of 30–60 mg/dL or cannot be identified visually or underestimated by laboratory staff [17]. So, incorrect results especially rising from mild hemolysis can damage patient safety.

In our study after introduction of automated serum indices to analyzers, rejected samples due to hemolysis raised significantly when compared the same period 1 year ago. Vermeer et al. reported that after introduction of serum indices for hemolysis detection in spite of visual determination hemolyzed sample (>0.5 g/L) were found to be six-fold increased [18]. Also in another study, Tamimi et al. [19] reported that after implementation of serum indices sample rejection rates were raised to 0.21% from 0.13% and the most significant sample rejection reason was hemolysis. They concluded that the increase in sample rejection rates assisted by automated analytical systems improved staff’s awareness. These findings were showed consistency with our data but none of these studies categorized hemolysis degree or rejected test profile.

In a study performed by Adiga and Yogish [20] they found that majority of the hemolyzed samples determined by automated analyzer had small to intermediate degree of hemolysis and reported that hemolysis index estimation defeats limitations of visual assessment and provide more unbiased and exact results. In our study we found 75.74% of hemolyzed samples were in Hem +1 group which reflects 50–99 md/dL hemoglobin presence was underestimated or could not be determined by laboratory staff with naked eye.

In our study rejected test rate was significantly lower after introduction of serum indices to laboratory than visual inspection period. This may be a consequence of test based rejection according to manufacturers’ suggestions in automation period. But in visual determination period all tests requested in the same sample were rejected regardless whether or not influenced test existed. Among rejected tests according to manufacturers’ suggestions, the most rejected test in Hem 1 and Hem 2 group was AST, Hem 3 group AST and ALT, Hem 4 group Glu. In the literature we could not find any data about sample rejection rates or test distribution after introducing HIL indices to analyzers. Further comparison studies are needed to see the effect of automated HIL indices usage on test rejection rates and rejected parameters.

In a study published by Söderberg et al. [21] which consisted primary health care centers, nursing homes and emergency departments analyzed hemolysis prevalence and age and gender distribution. They revealed that patients over median age had higher frequency of hemolysis which may be a consequence of difficulties at venipuncture in aged population. They also found in primary heath care centers women had higher frequency of hemolyzed sample. In our study samples from patients between 19 and 65 years old women had higher prevalence of hemolysis. Age and sex distribution analysis of total accepted samples to laboratory showed consistency with our finding, as the majority of patients were women between 19 and 65 years old.

Finally, this study has some limitations. First of all, in this retrospective study we were not able to find data about scale of hemolysis in visual evaluation period and rejected parameters prevalence cause of recording failures. Secondly, studies suggest laboratories should verify hemolysis effect on parameters provided by manufacturer [22]. In our study we relied on manufacturers’ documents cause of resource limitations.

In conclusion, to provide accurate test results and patient safety, interfering substances should be detected on chemistry analyzers by HIL indices which is fast and highly reliable.

The authors declare that there is no conflict of interests regarding the publication of this article.

References

1. Beastall GH. Adding value to laboratory medicine: a professional responsibility. Clin Chem Lab Med 2013;51:221–7.10.1515/cclm-2012-0630Search in Google Scholar PubMed

2. Carraro P, Plebani M. Errors in a stat laboratory: types and frequencies 10 years later. Clin Chem 2007;53:1338–42.10.1373/clinchem.2007.088344Search in Google Scholar PubMed

3. Lippi G, Blanckaert N, Bonini P, Green S, Kitchen S, Palicka V, et al. Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories. Clin Chem Lab Med 2008;46:764–72.10.1515/CCLM.2008.170Search in Google Scholar PubMed

4. Aykal G, Yegin A, Aydın O, Yılmaz N, Ellidag HY. The impact of educational interventions on reducing the rejection rates in the preanalytical phase. Turk J Biochem 2014;39:562–6.10.5505/tjb.2014.47113Search in Google Scholar

5. Lippi G, Gianfranco C, Camilla M. Critical review and meta-analysis of spurious hemolysis in blood samples collected from intravenous catheters. Biochem Med 2013;23:193–200.10.11613/BM.2013.022Search in Google Scholar

6. Lippi G, Mario P, Savatore DS, Gianfranco C. Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories. Crit Rev Cli Lab Sci 2011;48:143–53.10.3109/10408363.2011.600228Search in Google Scholar PubMed

7. Kirchner MJ, Funes VA, Adzet CB, Clar MV, Escuer MI, Girona JM, et al. Quality indicators and specifications for key processes in clinical laboratories: a preliminary experience. Clin Chem Lab Med 2007;45:672–7.10.1515/CCLM.2007.122Search in Google Scholar

8. Lippi G, Salvagno GL, Blanckaert N, Giavarina D, Green S, Kitchen S, et al. Multicenter evaluation of the hemolysis index in automated clinical chemistry systems. Clin Chem Lab Med 2009;47:934–9.10.1515/CCLM.2009.218Search in Google Scholar PubMed

9. Lippi G, Salvagno GL, Montagnana M, Giorgio B, Gian CG. Influence of hemolysis on routine clinical chemistry testing. Clin Chem Lab Med 2006;44:311–6.10.1515/CCLM.2006.054Search in Google Scholar PubMed

10. Howanitz P, Lehman CM, Jones BA, Mejer FA, Horowitz GL. Practices for identifying and rejecting hemolyzed specimens are highly variable in clinical laboratories. Arch Pathol Lab Med 2015;139:1014–9.10.5858/arpa.2014-0161-CPSearch in Google Scholar PubMed

11. Luksic AH, Gabaj NN, Miler M, Dukic L, Bakliza A, Simundic AM. Visual assessment of hemolysis affects patient safety. Clin Chem Lab Med 2018;56:574–81.10.1515/cclm-2017-0532Search in Google Scholar PubMed

12. Simundic AM, Nikolac N, Ivankovic V, Ruzic DF, Magdic B, Kvaternik M, et al. Comparison of visual vs. automated detection of lipemic, icteric and hemolyzed specimens: can we rely on a human eye? Clin Chem Lab Med 2009;47:1361–5.10.1515/CCLM.2009.306Search in Google Scholar PubMed

13. Lippi G, Cadamuro J, Meyer A, Simundic AM. Practical recommendations for managing hemolyzed samples in clinical chemistry testing. Clin Chem Lab Med 2018;56:718–27.10.1515/cclm-2017-1104Search in Google Scholar PubMed

14. Gabaj NN, Miler M, Vrtaric A, Hemar M, Filipi P, Kocijancic M, et al. Precision, accuracy, cross reactivity and comparability of serum indices measurement on Abbott Architect c8000, Beckman Coulter AU5800 and Roche Cobas 6000 c501 clinical chemistry analyzers. Clin Chem Lab Med 2018;56:776–88.10.1515/cclm-2017-0889Search in Google Scholar PubMed

15. Clinical and Laboratory Standards Institute. Hemolysis, icterus and lipemia/turbidity indices as indicators of interference in clinical laboratory analysis; approved guideline. C56A. Wayne, PA: Clinical and Laboratory Standards Institute, 2012.Search in Google Scholar

16. McCaughey EJ, Vecellio E, Lake R, Li L, Burnet L, Chesher D, et al. Current methods of haemolysis detection and reporting as a source of risk to patient safety: a narrative review. Clin Biochem Rev 2016;37:143–51.Search in Google Scholar

17. Simundic AM, Topic E, Nikolac N, Lippi G. Hemolysis detection and management of hemolyzed specimens. Biochem Med 2010;20:154–9.10.11613/BM.2010.018Search in Google Scholar

18. Vermeer HJ, Thomassen E, de Jonge N. Automated processing of serum indices used for interference detection by the laboratory information system. Clin Chem 2005;51:244–7.10.1373/clinchem.2004.036301Search in Google Scholar PubMed

19. Tamimi W, Martin-Ballesteros J, Brearton S, Alenzi FQ, Hasanato R. Evaluation of biological specimen acceptability in a complex clinical laboratory before and after implementing automated grading serum indices. Br J Biomed Sci 2012;69:103–7.10.1080/09674845.2012.12069134Search in Google Scholar

20. Adiga U, Yogish S. Hemolytic index – a tool to measure hemolysis in vitro. J Biotechnol Biochem 2016;2:49–52.Search in Google Scholar

21. Söderberg J, Jonsson PA, Wallin O, Grankvist K, Hultdin J. Haemolysis index- an estimate of preanalytical quality in primary health care. Clin Chem Lab Med 2009;47:940–4.10.1515/CCLM.2009.227Search in Google Scholar PubMed

22. Lim YK, Cha YJ. Proposal of modified HIL-indices for determining hemolysis, icterus and lipemia interference on the Beckman Coulter AU5800 Automated Platform. Lab Med Online 2017;7:66–72.10.3343/lmo.2017.7.2.66Search in Google Scholar

Received: 2018-11-12
Accepted: 2018-12-17
Published Online: 2019-01-26
Published in Print: 2019-10-25

©2019 Walter de Gruyter GmbH, Berlin/Boston

Articles in the same Issue

  1. Frontmatter
  2. Review Article
  3. Does vitamin D prevent radiotherapy-induced toxicity?
  4. Research Articles
  5. Compliance of medical biochemistry education in medical schools with national core education program 2014
  6. The importance of parathormone in determining the deficiency of vitamin D
  7. Association between serum vitamin D level and liver MRI T2 star in patients with β-thalassemia major
  8. Role of O-GlcNAcylation and endoplasmic reticulum stress on obesity and insulin resistance
  9. Effects of cellular energy homeostasis modulation through AMPK on regulation of protein translation and response to hypoxia
  10. Perceived barriers to diabetes management at home: a qualitative study
  11. The effect of automated hemolysis index measurement on sample and test rejection rates
  12. Identification of immune-related genes in thymus of breast cancer mouse model exposed to different calorie restriction
  13. Effect of xylitol on gut microbiota in an in vitro colonic simulation
  14. Fibrinopeptide-A and fibrinopeptide-B may help to D-dimer as early diagnosis markers for acute mesenteric ischemia
  15. Plasma homocysteine and aminothiol levels in idiopathic epilepsy patients receiving antiepileptic drugs
  16. Apelin-13 serum levels in type 2 diabetic obese women: possible relations with microRNAs-107 and 375
  17. An evaluation of biomarkers indicating endothelial cell damage, inflammation and coagulation in children with Henoch-Schönlein purpura
  18. Enteroprotective effect of Tsukamurella inchonensis on streptozotocin induced type 1 diabetic rats
  19. The in vitro cytotoxicity, genotoxicity and oxidative damage potential of dapagliflozin, on cultured human blood cells
  20. Investigation and isolation of peptide based antiglycating agents from various sources
  21. Effect of skin-to-skin contact on the placental separation time, mother’s oxytocin and pain levels: randomized controlled trial
  22. The protective role of oleuropein against diethylnitrosamine and phenobarbital induced damage in rats
  23. Letter to the Editor
  24. ICD code specific reference ranges
Downloaded on 18.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/tjb-2018-0462/html
Scroll to top button