Home Area dose measurements by thermoluminescence dosemeters for environmental monitoring
Article
Licensed
Unlicensed Requires Authentication

Area dose measurements by thermoluminescence dosemeters for environmental monitoring

  • J. Pillath , A. Knaps and R. Hille
Published/Copyright: March 11, 2022
Become an author with De Gruyter Brill

Abstract

For more than 25 years, thermoluminescence dosemeters (TLDs) have been used at Research Centre Jülich for the dose determination of photon radiation in environmental monitoring. Dose measurements are performed in six-month exposure intervals at 78 measuring locations in total at the outer fence and some additional locations inside and outside the research centre. Long-term variations in natural background radiation are analysed in comparison with climate data and the local conditions near the measuring locations and discussed with a view to determining the area-specific parameters. Taking the special conditions of Research Centre Jülich into account, the problems relating to the detection limit of 0.1 mSv/a, stipulated in the guideline for the emission and immission monitoring of nuclear facilities, are considered in more detail compared to natural background radiation.

Abstract

Seit über 25 Jahren werden am Forschungszentrum Jülich Thermolumineszenz-Dosimeter (TLD) zur Dosisbestimmung der Photonenstrahlung in der Umgebungsüberwachung eingesetzt. Die Dosismessungen erfolgen an insgesamt 78 Messstellen am Außenzaun und einigen zusätzlichen Orten innerhalb und außerhalb des Geländes des Forschungszentrums in halbjährlichen Expositionsintervallen. Langzeitige Schwankungen im natürlichen Strahlungsuntergrund sowie Unterschiede zwischen den Halbjahresdosen werden im Vergleich mit den Klimadaten und örtlichen Beschaffenheiten in der Nähe der Messorte analysiert und im Hinblick auf die Bestimmung der ortsspezifischen Parameter diskutiert. Unter Berücksichtigung der speziellen Gegebenheiten des Forschungszentrums Jülich wird die Problematik der in der Richtlinie zur Emissions- und Immissionsüberwachung kerntechnischer Anlagen geforderten Nachweisgrenze von 0,1 mSv/a gegenüber dem natürlichen Strahlungsuntergrund näher betrachtet.

References

1 Yamashita, T.; Ohnishi, H.; Yasuno, Y.; Takenaga, M.; Yamamoto, O.: Construction and Characterictics of CaSO4: Tm TLD. Technical Report Vol. 18, April 1972, Matsushita Electric Industrial Co.Search in Google Scholar

2 Weyrauch, M.; Weise, K.: Untersuchungen der Unsicherheiten bei der Messung mit Thermolumineszenz-Dosimetern. Report PTB-N-16 of the Physikalisch-Technische Bundesanstalt, Braunschweig, September 1993Search in Google Scholar

3 DIN 25 483 (1998): Verfahren zur Umgebungsüberwachung mit integrierenden Festkörperdosimetern. DIN Deutsches Institut für Normung e.V. (German Institute for Standardization), Beuth Verlag, BerlinSearch in Google Scholar

4 Umweltradioaktivität und Strahlenbelastung im Jahr 2001, Federal Government information, Federal Ministry for the Environment, Nature Conservation and Nuclear SafetySearch in Google Scholar

5 Verordnung über den Schutz vor Schäden durch ionisierende Strahlen (Radiation Protection Ordinance), as amended by the promulgation of 30 June 1989, Federal Law Gazette I, p. 1321Search in Google Scholar

6 Richtlinie zur Emissions- und Immissionsüberwachung kerntechnischer Anlagen (REI) of 19 August 1993, GMBl. Nr. 29, S. 501Search in Google Scholar

7 Czarnecki, J.; Völkle, H.; Baggenstoß, M.; Schüler, J.: Die Bestimmung von Netto-Dosen in der Umgebung von Kernkraftwerken mit Hilfe von Thermolumineszenzdosimetern: Auswerteverfahren und erreichbare Nachweisgrenzen, Annual Conference of the Professional Association for Radiation Protection (19.–22. 10. 1982 in Munich).Search in Google Scholar

8 Czarnecki, J.: An Interpretation Method of TLD-Environmental Data. Health Physics 45 (1983) 173–179Search in Google Scholar

9 Empfehlungen zur Überwachung der Umweltradioaktivität, Loose-leaf edition of the Professional Association for Radiation Protection, FS-78-15 AKU, p. 3.4.1, 30. März 1988Search in Google Scholar

Received: 2003-04-07
Published Online: 2022-03-11

© 2003 Carl Hanser Verlag, München

Articles in the same Issue

  1. Frontmatter
  2. Calendar of events · veranstaltungskalender
  3. Summaries/Kurzfassungen
  4. Editorial
  5. Radiation Protection: measuring and modeling
  6. Modeling, simulation and measurement of radiation exposure using electronic personal dosimeters on realistic and ICRU phantoms
  7. Books · Bücher
  8. Emerging systemic risks in the 21th century
  9. Status and prospects of electronic personal dosemeters for neutron/photon radiation fields
  10. Mathematical efficiency calibration in gamma spectroscopy
  11. Note · Mitteilung
  12. Karl-Wirtz-Preis der Kerntechnischen Gesellschaft e.V
  13. Recent trends in radioanalytical purification procedures and application to decommissioning of nuclear facilities
  14. Migration of iodine-129 and iodine-127 in soils
  15. Uncertainties of measurements in radiation protection
  16. The response of various neutron dose meters considering the application at a high energy particle accelerator
  17. Books · Bücher
  18. Grundkurs Strahlenschutz Praxiswissen für den Umgang mit radioaktiven Stoffen
  19. Application of modern radiochemical techniques to the determination of Sr-89/Sr-90 in environmental samples
  20. Books · Bücher
  21. Indemnification of Nuclear Damage in the Event of a Nuclear Accident
  22. Influence of calcium on the uptake of ingested strontium
  23. Books · Bücher
  24. Utilisation and Reliability of High Power Proton Accelerators
  25. Area dose measurements by thermoluminescence dosemeters for environmental monitoring
  26. Technical Note
  27. Some comments on Herndon’s nuclear georeactor
  28. Advances · Patente
  29. Patent
  30. Books · Bücher
  31. Possible Implication of Draft ICRP Recommendations
Downloaded on 23.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/kern-2003-0078/html
Scroll to top button