Abstract:
Motor functions can be hindered in consequence to a stroke or a spinal cord injury. This often results in partial paralyses of the upper limb. The effectiveness of rehabilitation therapy can be improved by the use of rehabilitation robotics and Functional Electrical Stimulation (FES). We consider a hybrid arm weight support combining both. In order to compensate the effect of FES-induced muscle fatigue, we introduce a method to substitute the decreasing level of FES support by cable-driven robotics. We evaluated the approach in a trial with one healthy subject performing repetitive arm lifting. The controller automatically adapted the support and thus no increase in user generated volitional effort was observed when FES induced muscle fatigue occured.
1 Introduction
A stroke or a spinal cord injury (SCI) can result in injuries of the central nervous system (CNS), which may lead to a partial or complete loss of motor function in parts of the body. Rehabilitation therapy has been shown to improve recovery of motor functions after a stroke or SCI [1], [2]. Two already widely used methods during rehabilitation therapy are Functional Electrical Stimulation (FES) and rehabilitation robotics. Robotic systems are mostly used to provide external assistance to the paralyzed limb in order to facilitate movement. FES also supports the patient during motion, with the advantage that it activates the muscles, thereby strengthening them while retarding muscle atrophy and reducing spasticity [3]. It also improves proprioception and kinesthetic feedback. However, the trajectory of an FES assisted movement is harder to control [4]. Another main disadvantage of using FES during therapy is the quick occurrence of muscle fatigue since the CNS activates muscle fibers in a much more efficient way than FES does [5].
Studies have proven the effectiveness of both methods in improving the recovery of motor functions [6], [7]. In addition, recent research shows that combining robotic assisted rehabilitation with FES further increases the benefits of rehabilitation therapy [8], [5].
Our approach uses the support provided by the robotic system to compensate the effects of FES-induced muscle fatigue. During longer therapy sessions, the stimulation intensity required to achieve a certain level of muscle activity increases as the electrically stimulated muscles get fatigued. In this case, we increase the proportion of support provided by the robotics.
2 Methods
2.1 Experimental setup
Experiments were performed using the setup shown in Figure 1. A stimulator (RehaStim, Hasomed GmbH, Germany), an active cable-driven robotic system (Diego, Tyromotion GmbH, Austria), an EMG Amplifier (Porti 32, TMSi, Netherlands) and a computer running Linux were used. The control algorithms were implemented in Simulink® (Mathworks, USA) and a modified Linux real-time target was used to generate an executable binary that runs in soft real-time at 25 Hz. Internal variables and parameters were visualized and set using a Python-GUI.
![Figure 1 Experimental Setup: The required elbow rope tension uTe and the recruitment level rλ^\(r_{\hat{\lambda}}\) to hold the arm at a shoulder flexion angle θu are scaled with the support factors sr and sq respectively. The total scale factor stotal = sr + sq has to be within [0, 1]. The EMG is recorded and processed to obtain the FES induced muscle recruitment level λ^\(\hat{\lambda}\) and the volitional muscle activity γ^\(\hat{\gamma}\). The recruitment controller calculates the required stimulation intensity q which is then generated by the stimulator using the current I and the pulse width pw. The cable-driven robotics generates the tension vectors srTE and srTW.](/document/doi/10.1515/cdbme-2017-0007/asset/graphic/j_cdbme-2017-0007_fig_001.jpg)
Experimental Setup: The required elbow rope tension uTe and the recruitment level
The stimulation pulses were transmitted to the muscle via two self-adhesive hydro-gel transcutaneous electrodes overlapping the anterior and medial part of the deltoid muscle. The stimulator generated bi-phasic stimulation pulses of current I ∈ [0, 100] mA and pulse width pw ∈ [100, 500] μs. Thereby I and pw were computed to realize the normalized pulse charge q according to [9].
Two ropes of the robotics system were attached to the forearm’s elbow and wrist joints. The actuators, located above the user, apply a tension to the ropes and attached sensors provided data from which the position of the arm and thus the shoulder flexion angle θu were calculated.
2.2 Control idea
Two control algorithms provide adaptive weight support to the user’s arm. The first control algorithm calculates the rope tensions required to hold the arm at the current angle θu. A second algorithm computes the needed stimulation intensity to maintain θu. Both methods are scaled by factors to determine their amount of weight support provided to the user’s arm. A scale factor controller adapts these two factors if FES-induced muscle fatigue is detected.
2.3 FES weight support controller
The virtual weight compensation using FES is implemented as presented in [10]. The FES induced muscle recruitment level
As muscle fatigue occurs, the muscle recruitment level
2.4 Robotics weight support controller
The weight support applied by the robotics to compensate the gravitational forces of the arm is implemented as described in (11). The required rope tension at the elbow
In contrast to [11], no additional identification routine is necessary due to an alternative determination of the necessary arm parameters. The weights of the upper arm and forearm are considered to be a percentage of the total body weight [12]. The lengths of both arm segments are measured by means of the robotics at the beginning of each experiment. The patient dependent factor sr ∈ [0, 1] scales the provided elbow rope tension to
2.5 Scale factor controller
Our approach assumes that the user alternates between holding his arm at a lower angle and an upper angle by following a given reference for θu. The reference is shown to the subject, who is asked to follow it by using his residual voluntary activity. The scale factor controller updates the scale factors sq and sr at the end of every iteration k after the subject’s arm comes down from the upper angle. The mean of q during the phase when the upper angle is reached, is averaged over the last three iterations to yield
To achieve this, the FES scale factor sq is adapted by an integrating controller as
where KI is the adjustable gain of the integrator and the error e[k] is defined as
Thereby sq, init is the initial value of sq at the beginning of the experiment. The total support level stotal = sr + sq is maintained by sr[k] = stotal − sq[k].
2.6 Experimental procedure
The complete setup was implemented as described in Section 2.1. At first, the maximum tolerated stimulation intensity qmax of the subject was detected by increasing q until the subject reported discomfort. Then q was increased until qmax was reached and afterwards decreased to zero to identify the required
For the healthy subject, a lower threshold σq = 0.75 qmax and at the same time a high stotal = 0.9 were chosen. Thus, it was possible to provoke relatively quickly occurring muscle fatigue. The subject had to complete 20 iterations of a movement, successively raising and lowering the arm to achieve a shoulder flexion angle θu that is within two reference bands at 50° ± 5° (lower angle) and 90° ± 5° (upper angle). For this reason, the actual desired reference band was visualized to the subject.
3 Results
The results of the trial with one healthy subject (age 21 years) are shown in Figure 2. FES-induced muscle fatigue occurred starting at the iteration around 90-100 s, whereby the recruitment controller increased q to achieve the scaled desired muscle recruitment level sq

Experimental results of a trial with a healthy subject. The subject had to hold θu between two visualized reference bands.
The estimated volitional muscle activity
4 Conclusion
The use of a robotic system to compensate the effects of muscle fatigue appears to be a promising approach to eliminate one of the biggest drawbacks of FES during arm lifting. Further trials with stroke or SCI patients have to be performed in order to confirm the effectiveness of the procedure. Due to the fast setup and only one containing fast identification routine, the presented method is relatively simple to apply. The main limitation of the current approach is that changes in q are only analyzed when the flexion angle is near the chosen upper angle. More work needs to be done on tracking muscle fatigue during arbitrary movements. A further development using the approach of [13] will enable longer and therefore more efficient therapy sessions.
Acknowledgement
We would like to thank Axelgaard Manufacturing Co., USA for donating the used stimulation electrodes.
Author’s Statement
Research funding: The work was conducted within the research project BeMobil, which is supported by the German Federal Ministry of Education and Research (BMBF) (FKZ16SV7069K). Conflict of interest: Authors state no conflict of interest. Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use complies with all the relevant national regulations, institutional policies and was performed in accordance with the tenets of the Helsinki Declaration. The performed trials have been approved by the ethics committee of the Berlin Chamber of Physicians.
References
Feys HM, De Weerdt WJ, Selz BE, Cox Steck GA, Spichiger R, Vereeck LE, et al. Effect of a therapeutic intervention for the hemiplegic upper limb in the acute phase after stroke. Stroke 1998;29:785–92.10.1161/01.STR.29.4.785Search in Google Scholar
Sumida M, Fujimoto M, Tokuhiro A, Tominaga T, Magara A, Uchida R. Early rehabilitation effect for traumatic spinal cord injury. Arch Phys Med Rehabil. 2001;82:391–5.10.1053/apmr.2001.19780Search in Google Scholar
Doucet BM, Lam A, Griffin L. Neuromuscular electrical stimulation for skeletal muscle function. Yale J Biol Med. 2012;85:201–15.Search in Google Scholar
Vette AH, Masani K, Kim JY, Popovic MR. Closed-loop control of functional electrical stimulation-assisted arm-free standing in individuals with spinal cord injury: a feasibility study. Neuromodulation 2009;12:22–32.10.1111/j.1525-1403.2009.00184.xSearch in Google Scholar
Hu XL, Tong RKY, Ho NSK, Xue JJ, Rong W, Li LSW. Wrist rehabilitation assisted by an electromyography-driven neuromuscular electrical stimulation robot after stroke. Neurorehabilitation and Neural Repair 2015;29:767–76.10.1177/1545968314565510Search in Google Scholar
Kawashima N, Popovic MR, Zivanovic V. Effect of intensive functional electrical stimulation therapy on upper-limb motor recovery after stroke: Case study of a patient with chronic stroke. Physiother Can 2013;65:20–8.10.3138/ptc.2011-36Search in Google Scholar
Volpe BT, Lynch D, Rykman-Berland A, Ferraro M, Galgano M, Hogan N, et al. Intensive sensorimotor arm training mediated by therapist or robot improves hemiparesis in patients with chronic stroke. Neurorehabil Neural Repair. 2008;22:305–10.10.1177/1545968307311102Search in Google Scholar
Miyasaka H, Orand A, Ohnishi H, Tanino G, Takeda K, Sonoda S. Ability of electrical stimulation therapy to improve the effectiveness of robotic training for paretic upper limbs in patients with stroke. Med Eng Phys. 2016;38:1172–5.10.1016/j.medengphy.2016.07.010Search in Google Scholar
Klauer C, Raisch J, Schauer T. Linearisation of electrically stimulated muscles by feedback control of the muscular recruitment measured by evoked emg. Proc. of the 17th International Conference on Methods and Models in Automation and Robotics, IEEE, Międzyzdroje, Poland; 2012. p. 108–13.10.1109/MMAR.2012.6347902Search in Google Scholar
Klauer C, Passon A, Raisch J, Schauer T. Virtual weight-compensating exoskeleton using λ-controlled fes. Proc. of Automed 2013, Dresden, Germany; 2013.Search in Google Scholar
Knuth S, Passon A, Dähne F, Niedeggen A, Schmehl I, Schauer T. Adaptive arm weight support using a cable-driven robotic system. Springer International Publishing; 2017. p. 1317–21.10.1007/978-3-319-46669-9_215Search in Google Scholar
de Leva P. Adjustments to zatsiorsky-seluyanov’s segment inertia parameters. J Biomech. 1996;29:1223–30.10.1016/0021-9290(95)00178-6Search in Google Scholar
Klauer C, Ferrante S, Ambrosini E, Shiri U, Dähne F, Schmehl I, et al. A patient-controlled functional electrical stimulation system for arm weight relief. Med Eng Phys. 2016;38:1232–43.10.1016/j.medengphy.2016.06.006Search in Google Scholar PubMed
©2017 Arne Passon et al., licensee De Gruyter.
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.
Articles in the same Issue
- Assessment of upper extremity movement performance of patients with obstetric brachial plexus palsy
- Analysis of ergonomic and unergonomic human lifting behaviors by using Inertial Measurement Units
- Comparison of dynamic balancing responses following outward lateral perturbations during walking of healthy and post-stroke subjects
- A physically-motivated model describing the dynamic interactions between residual limb and socket in lower limb prostheses
- Modular finger and hand motion capturing system based on inertial and magnetic sensors
- Functional electrical stimulation for home-based upper-limb stroke rehabilitation
- Compensating the effects of FES-induced muscle fatigue by rehabilitation robotics during arm weight support
- Method for instrumentation of orthotic joints for measurement of internal joint loads
- User-centered design of a patient’s work station for haptic robot-based telerehabilitation after stroke
- Modelling of therapeutic action during mirror therapy
- Adjustment of instantaneous reference walking speed during synchronized robot-based gait rehabilitation
- Application of head-mounted devices with eye-tracking in virtual reality therapy
- Design of a pressure sensitive matrix for analyzing direct haptic patient-therapist interaction in motor rehabilitation after stroke
- PDMS electrodes for recording and stimulation
- User-centred design as an important component of technological development
Articles in the same Issue
- Assessment of upper extremity movement performance of patients with obstetric brachial plexus palsy
- Analysis of ergonomic and unergonomic human lifting behaviors by using Inertial Measurement Units
- Comparison of dynamic balancing responses following outward lateral perturbations during walking of healthy and post-stroke subjects
- A physically-motivated model describing the dynamic interactions between residual limb and socket in lower limb prostheses
- Modular finger and hand motion capturing system based on inertial and magnetic sensors
- Functional electrical stimulation for home-based upper-limb stroke rehabilitation
- Compensating the effects of FES-induced muscle fatigue by rehabilitation robotics during arm weight support
- Method for instrumentation of orthotic joints for measurement of internal joint loads
- User-centered design of a patient’s work station for haptic robot-based telerehabilitation after stroke
- Modelling of therapeutic action during mirror therapy
- Adjustment of instantaneous reference walking speed during synchronized robot-based gait rehabilitation
- Application of head-mounted devices with eye-tracking in virtual reality therapy
- Design of a pressure sensitive matrix for analyzing direct haptic patient-therapist interaction in motor rehabilitation after stroke
- PDMS electrodes for recording and stimulation
- User-centred design as an important component of technological development