Abstract
Objectives
The increasing adoption of point-of-care testing (POCT) in primary healthcare highlights the need for robust quality assurance (QA) procedures to ensure result reliability. The patient split sample approach, i.e. comparing POCT and central laboratory results from the same patient, can be a valuable QA tool, though practical guidance for its implementation remains scarce. This study aimed to develop clear, user-friendly recommendations for non-laboratory personnel in primary healthcare laboratories on when and how to perform such comparisons and to recommend acceptable limits for the compared results.
Methods
In 2023, an expert group was established, composed of medical specialists in laboratory medicine, researchers, and laboratory advisors. The recommendations were formulated based on relevant literature, the professional experience of the group members, and Noklus’ educational framework for primary care laboratories. Pragmatic acceptability limits were established taking several approaches into account. The draft recommendations were audited among more than 120 Noklus employees and final consensus was reached in 2024.
Results
Comparing POCT results with central laboratory results is recommended when: (A) no suitable EQA program exists for a given POCT; (B) appropriate internal quality control materials are lacking or inadequate; or (C) POCT results contradict clinical expectations. Acceptable limits for the compared results were set at 15 or 20 %, depending on the measurand. Results outside these limits should be reviewed using a structured checklist.
Conclusions
These are the first published recommendations for using patient split samples in primary healthcare that are designed to be simple and user-friendly for non-laboratory personnel, facilitating widespread adoption.
Introduction
The use of point-of-care testing (POCT) is rapidly growing in laboratory medicine, offering advantages like accessibility, faster results, immediate decision-making, and ease of use. However, as for all medical laboratory testing, accurate POCT results are critical, since inaccurate results can lead to misdiagnosis, inappropriate treatment decisions, and adverse patient outcomes [1], 2]. In primary healthcare settings, where POCT is conducted by staff with varying laboratory expertise, robust quality assurance (QA) protocols are essential to ensure reliable results [3].
Key factors of QA protocols for POCT include internal quality control (IQC) and participation in external quality assessment (EQA) programs, along with education and guidance on test operation and result interpretation [4]. Additionally, some primary healthcare laboratories perform quality control by using patient split sample comparisons, also known as “parallel testing”, with a central laboratory. This process involves comparing results from the same patient using two different measuring systems, often using two different measurement procedures, typically analyzing a capillary or venous whole blood sample on a POCT instrument and a venous sample (serum, plasma or whole blood) on a central laboratory instrument. However, there are few published recommendations on when and how these split sample comparisons should be conducted, and to our knowledge, there are no publications about what are considered acceptable deviations of the compared results.
A recent publication from International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) committee for POCT [5] recommends routine split sample comparisons where appropriate, but without specifying detailed procedures. Australia and the United Kingdom recommend split sample testing as an alternative to EQA in cases where EQA programs are unavailable, though specific guidelines on implementation are lacking [6], 7]. However, there are some specific recommendations such as in Denmark, where systematic monthly comparisons with defined performance specifications are recommended for selected analytes as a replacement for EQA [8]. However, these recommendations are currently under revision. Further, the Canadian society of clinical chemists point of care testing interest group have published guidelines on how often split sample testing should be performed and the number of patient samples required, depending on the complexity of the POCT instrument and the testing setting [9]. However, these guidelines recommend quite comprehensive split sample routines making it challenging for the primary healthcare laboratories to follow. In addition, no recommendations are given on performance specifications or what to do if results are outside the limits.
The Norwegian organization for quality improvement of laboratory examinations (Noklus) with about 3,500 participants has more than 30 years of experience in ensuring POCT quality in primary healthcare [10]. Noklus provides EQA schemes, along with education and guidance for primary healthcare laboratories. It also offers online quality assurance (QA) protocols to support comprehensive quality management of POCT, including IQC [4]. However, formal recommendations for conducting patient split sample comparisons with central laboratories are lacking.
The aim of this study was to develop practical and easy-to-follow recommendations for primary healthcare laboratories on when, and how, to compare POCT results with central laboratory results using patient split samples as part of the QA protocol, and to establish acceptability limits for the compared results.
Materials and methods
Working group in Noklus
In March 2023, a working group in Norway was established to develop guidelines for comparing POCT results with central laboratory results using a split-sample approach. This multidisciplinary working group consisted of medical specialists in laboratory medicine, experienced researchers, and laboratory advisors who have extensive expertise in guiding POCT users in primary healthcare settings. The recommendations were formulated based on the current IQC guidelines provided by Noklus, a review of relevant literature, and the collective expertise of the working group members.
The recommendations were agreed on after a thorough process involving the whole Noklus organization. More than 120 employees (including 60 laboratory advisors and 20 medical specialists in laboratory medicine) were encouraged several times to give input on the draft recommendations. In addition, the project was presented and audited at meetings on different occasions, making it possible to give input during the whole process. The working group had in total 12 meetings, and after reviewing all inputs, the working group reached a consensus in December 2024.
Establishing pragmatic acceptability limits for split sample comparisons
Even though there are publications describing how to establish analytical performance specifications (APSs) for medical devices [11], it is not straight forward to establish acceptability limits for split sample comparisons. All APSs, also those based on biological variation, should be targeted for the intended use [12]. In this situation the intended use is for split samples. i.e. acceptable limits for differences between results in POCT in primary healthcare and results obtained at a central laboratory. It is important to underline that these limits will be applied by people who work in e.g. general practitioners’ offices and have little or no education in laboratory medicine. It is therefore important that the limits are pragmatic, simple and easy to remember.
The working group discussed and reviewed different approaches to establish pragmatic limits for the most commonly used analytes in primary healthcare settings in Norway, ensuring these limits realistically reflect the expected variability when comparing POCT results with those from central laboratories.
When establishing the acceptability limits for split sample comparisons it has to be taken into account that there are several factors that can contribute to discrepancies between POCT and laboratory results. This includes bias between the compared instruments, imprecision of both the POCT and laboratory measuring system, instability of the analyte during transport, differences in sample materials (e.g. capillary vs. venous blood), and measuring system-specific interferences (differences in method selectivity). Given these potential sources of variation, a relatively large difference between the two results may be observed without indicating an out-of-control situation. Therefore, the stable analytical imprecision and potential bias of both the measuring systems, along with preanalytical variables such as sample collection, should be considered when defining the split sample limits.
Several approaches were taken into account when establishing the pragmatic split sample acceptability limits for selected measurands; 1) desirable level of APSs for analytical imprecision and bias derived from biological variation [13] and from selected references [14], [15], [16], [17], 2) recommended APSs for desirable level of standard measurement uncertainty from selected references [18], [19], [20], and 3) EQA limits for POCT used by Noklus. Calculations based on both the total allowable error concept and the maximum allowable uncertainty concept were made for all these numbers (data not shown), and a pragmatic acceptability limit was chosen for each measurand.
Results
Recommendations on when, and how, to compare POCT results with central laboratory results
Prior to conducting split sample comparisons with a central laboratory, it is important to ensure that specific prerequisites are fulfilled. The reference limits should be comparable for the POCT and laboratory measuring system, the measurand must be stable during transport, and both the POCT and laboratory samples should be collected simultaneously, preferably from the same puncture, to minimize pre-analytical variability.
Using split sample comparisons with a central laboratory should not be the preferred approach of QA for POCT in primary healthcare and random split sample testing should be avoided. The reason for this is the complexity involved in interpreting discrepancies between POCT and laboratory results because of several factors that can contribute to the deviation.
The most reliable practices for ensuring the accuracy and precision of POCT are regular performance of IQC and participation in EQA programs. However, as a substitute for traditional quality control, there are specific circumstances (scenarios) where split-sample comparisons may be useful.
When EQA programs for POCT are unavailable or unsuitable for the specific test in question
When appropriate IQC materials for POCT are not accessible, or when the available materials are either unsuitable or only present in clinically irrelevant concentrations
When POCT results are unexpected given a patient’s clinical history
Detailed recommendations for each of these scenarios are given in Table 1. The pragmatic acceptability limits for the difference between a POCT result and a central laboratory result were set to 15 or 20 % dependent on the measurand (Table 2). Should a split sample result fall outside these limits, a troubleshooting process should be initiated. The working group has developed a checklist to assist in identifying potential reasons for the deviating result which includes a step-by-step procedure to identify underlying issues (Table 3). In addition, a checklist was made to eliminate analytical and user-errors before sending a split sample to a central laboratory in situations when obtaining unexpected POCT result for one individual patient (scenario C) (Table 4).
Recommendations for primary healthcare laboratories on when (scenarios A, B and C), and how, to compare results with a central laboratory using patient split samples.
Scenarios | A (if no suitable EQA scheme) | B (if no suitable IQC material) | C (if unexpected POCT result for one patient) |
---|---|---|---|
Aim of the split sample analysis | Accuracy assessment of the POCT device | Internal analytical quality control of the POCT device | Finding the correct result for one individual patient |
How many samples? | 2 patient samples in different clinically relevant concentrations | 1 patient sample in a clinically relevant concentration (use different concentrations each time) | 1 patient sample |
How frequent? | 2–4 times a year | Weekly or monthly dependent on the type of POCT device (see reference [21]) | When the POCT result is unexpected (analytical and user errors must be eliminated before performing split sample, see Table 4) |
How to assess the results against the acceptability limits? | Assess each result separately | Plot results, assess each result separately, assess the plot again after 5 results, assess the plot continually thereafter to assess trends | Assess the result separately |
What to do if results are outside the limits? (Troubleshooting) | Analyze IQC material (if possible), find and correct the error (use the troubleshooting guide, see Table 3) | Find and correct the error (use the troubleshooting guide, see Table 3), already released patient results should be reviewed carefully | Use the result from central lab. If this patient has consecutive results outside the limits, do not use POCT results for this patient |
Pragmatic acceptability limits for split sample comparisons (difference between the POCT result and the central laboratory result) for selected measurands.
Measurand | Acceptable limits for split sample comparisons, % |
---|---|
Glucose, HbA1c, hemoglobin, erythrocytes, EVF | 15 |
Leukocytes, thrombocytes, CRP, INR, cholesterol, LDL, HDL, triglycerides, D-dimer, troponin T, U-albumin, U-ACR | 20 |
-
HbA1c, glycated hemoglobin type A; EVF, erythrocyte volume fraction; CRP, C-reactive protein; INR, International Normalized Ratio; LDL, low-density lipoprotein; HDL, high-density lipoprotein; ACR, albumin/creatinine ratio.
Step-by-step troubleshooting guide for deviating split sample results in situations where no suitable EQA scheme is available (scenario A) or no suitable IQC material is available (scenario B).
Troubleshooting guide for deviating split sample results (in scenario A or B) | |
Sample sent to the central laboratory | Was the blood sampling performed correctly? |
Was the sample handled correctly? e.g. storage, centrifugation | |
Was the transport time acceptable (within stability of the measurand)? | |
Central laboratory measurement | Was there any problems or issues? |
POCT measurement | Was the blood sampling performed correctly? |
Was the measurement performed according to the recommended instruction for use? | |
POCT test reagents/strips/cassettes | Were the test reagents handled and stored correctly? e.g. temperature |
Was the test reagent used within the expiration date? | |
POCT instrument | Was the instrument operating in the correct mode? |
Has maintenance been performed recently? | |
Was results from any electronic QC cartridge within acceptable limits? | |
EQA results | Were the recent EQA results within acceptable limits?a |
IQC results | Were the recent IQC results within acceptable limits?b |
-
aIn scenario A, no suitable EQA scheme is available and thus no EQA results exits. bScenario B, no suitable IQC materials is available and thus no IQC results exist.
Check list before sending split samples to a central laboratory for comparison in situations when receiving unexpected POCT result for one individual patient (scenario C). Analytical and user errors must be eliminated before performing split sample, therefore, “yes” on all questions is required.
Check list before sending a sample with unexpected POCT result (scenario C) | |
The patient | Was the sample collected from the correct patient? |
Sample analyzed at the POCT measuring system | Was the blood sampling performed correctly? |
Was the measurement performed according to the recommended instruction for use? | |
Was the patient sample analyzed twice? | |
POCT instrument and reagents/strips/cassettes | Was the instrument operating in the correct mode? |
Were the test reagents handled and stored correctly? e.g. temperature | |
QC results | Was results from any electronic QC cartridge within acceptable limits? |
Were the recent IQC results within acceptable limits? |
Discussion
This study presents recommendations for primary healthcare laboratories on when, and how, to compare POCT results with central laboratory results as part of the QA protocol using patient split samples. To our knowledge, this is the first study presenting pragmatic and user-friendly guidance for non-laboratory personnel and includes recommendation of which acceptability limits to use. The patient split sample approach serves as a valuable quality assurance tool in primary healthcare when IQC or EQA is not suitable or possible, ensuring the accuracy and reliability of POCT by comparing results with those obtained from a central laboratory.
While split sample testing is designed to verify POCT accuracy, it is important to acknowledge that central laboratory measuring systems may also produce inaccurate results compared to a reference measurement procedure. Therefore, before implementing split sample testing, it is advisable to review the laboratory’s performance in EQA programs to better interpret potential discrepancies. It is also important to assess the stability of the measurand in question to ensure that it remains stable during transport.
One primary application of split sample testing is in situations where no EQA programs are available for the POCT device (scenario A, Table 1). In this scenario, we recommend using at least two patient samples in clinically relevant concentration levels (e.g., one normal and one high/low level) approximately 2–4 times per year. The results should be recorded and assessed against predefined acceptability limits. If one or more results are outside the limits, root cause analysis should be conducted using a dedicated troubleshooting checklist (Table 3).
Some POCT instruments lack dedicated IQC control materials, while others provide controls that behave significantly differently from patient samples, limiting their effectiveness. In such cases, split sample testing provides a practical alternative for monitoring the POCT instrument (scenario B, Table 1). For practical reasons, we recommend using one patient sample in clinically relevant concentration level on a weekly or monthly basis. The frequency can be determined by using the scoring system recommended for IQC routines in primary healthcare [21], however, daily split sample testing is not recommended, as it would result in an unnecessary burden on workload and offers minimal additional benefit [22].
As for all QA results, the split sample results should be recorded, and deviating results should be investigated with appropriate corrective actions taken if clinically necessary. Identified errors should be corrected, and it must be assessed whether previously reported patient results were affected and should be recalled. With appropriate guidance, primary care laboratories should be capable of managing QA processes effectively, even when staff have limited laboratory experience.
Although routine IQC procedures recommend additional quality checks when new reagents are introduced or after instrument maintenance [21], 23], patient split sample testing may not be practical in these situations due to the delayed availability of central laboratory results. Instead, an alternative approach involves analysing a patient sample using both the old and new reagent or pre- and post-maintenance, allowing potential discrepancies to be detected. These results should be documented and reviewed over time, similar to standard analytical quality assurance.
In scenario C (Table 1), the primary goal of split sample testing is to ensure individual patient result accuracy rather than analytical quality assurance. In this situation, it is assumed that the primary care laboratory participates in EQA programs and use IQC materials at a regular basis or uses a split sample approach as recommended in this paper. It is important that user errors and errors related to the operation procedure are eliminated before sending the sample to the laboratory for comparison as outlined in Table 4. If errors are detected, these should be corrected and the sample re-analysed. If no errors are detected and the POCT result appears inconsistent with the clinical presentation, confirmation through laboratory testing (such as a split sample comparison) may be warranted.
If significant discrepancies between POCT and laboratory results are observed, preference should be given to the laboratory result given that the quality control routines in the central laboratory are sound. In cases of recurring deviations which may be caused by e.g. interferences present in the sample from the patient, consideration should be given to permanently analysing that patient’s samples at the central laboratory. Possible causes of persistent discrepancies include patient-related interferences, such as e.g. the presence of antiphospholipid antibodies affecting POCT INR measurements [24].
There are currently no published recommendations for which acceptable limits to use in split sample comparisons. In this study, we have used a pragmatic approach when establishing the acceptability limits to ensure ease of use in primary healthcare settings. The limits were determined taking into account several factors, including biological variation, state-of-the-art, EQA limits, and recommendations from publications, using both the total allowable error (TEa) concept and the maximum allowable uncertainty (MAU) concept [25], 26]. Because these recommendations are intended for non-laboratory personnel, the aim was to provide easy-to-remember numbers and similar limits for as many measurands as possible. Therefore, we did not establish the exact number for each measurand.
The limits must account for various factors influencing differences between POCT and laboratory results and is therefore wide. Consequently, the split sample approach serves as a less sensitive quality assurance tool than traditional EQA and IQC and can thus only detect large errors. This is why we recommend that this approach should never be the preferred choice of performing quality assurance of POCT in primary healthcare [22], it should only be done in special circumstances as described above (scenario A-C). Although sending patient samples to a clinical laboratory may be less expensive than purchasing commercial control materials, we do not recommend replacing traditional IQC and EQA with the split sample approach since it might be less sensitive, and the interpretation of the results can be more challenging. In addition, ethical guidelines generally require that patients provide oral (or written) informed consent before an additional venous sample is collected (for measurement of the same measurand) solely for quality assurance purposes. However, if a venous sample has already been obtained for POCT, usually no such consent is needed.
The strength of this study is that it reflects the experience and opinion of the whole Noklus organisation and not just the opinion of the authors. The recommendations are a product of the organisation’s principles for education and guidance of the primary healthcare for more than 30 years [10]. The limitations are that no experts outside Norway contributed, and that the recommendations are valid only for quantitative tests such as for example glucose meters, CRP and HbA1c POCT devices, it does not include qualitative rapid tests such as for example SARS-CoV-2 and Streptococcus A, or semiquantitative test such as urine strips.
In conclusion, recommendations for patient split sample procedures have been developed to ensure reliable POCT results in primary healthcare laboratories. These procedures provide a quality assurance framework that can be used in special situations making it possible to detect significant errors and ensure safe patient management. Importantly, they are designed to be simple and user-friendly for non-laboratory personnel, facilitating widespread adoption in primary healthcare settings. Future work will monitor the usability of these recommendations among the Noklus’ participants.
-
Research ethics: Not applicable.
-
Informed consent: Not applicable.
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Use of Large Language Models, AI and Machine Learning Tools: None declared.
-
Conflict of interest: The authors state no conflict of interest.
-
Research funding: None declared.
-
Data availability: Not applicable.
References
1. Price, CP, Smith, I, Van den Bruel, A. Improving the quality of point-of-care testing. Fam Pract 2018;35:358–64. https://doi.org/10.1093/fampra/cmx120.Search in Google Scholar PubMed
2. Banfi, G, Bozic, B, Cihan, M, Pasalic, D, Pennestri, F, Plebani, M. Point-of-care testing, near-patient testing and patient self-testing: warning points. Clin Chem Lab Med 2024;62:2388–92. https://doi.org/10.1515/cclm-2024-0525.Search in Google Scholar PubMed
3. Lewandrowski, K, Gregory, K, Macmillan, D. Assuring quality in point-of-care testing: evolution of technologies, informatics, and program management. Arch Pathol Lab Med 2011;135:1405–14. https://doi.org/10.5858/arpa.2011-0157-ra.Search in Google Scholar PubMed
4. Stavelin, A, Sandberg, S. Analytical performance specifications and quality assurance of point-of-care testing in primary healthcare. Crit Rev Clin Lab Sci 2024;61:164–77. https://doi.org/10.1080/10408363.2023.2262029.Search in Google Scholar PubMed
5. Khan, AI, Pratumvinit, B, Jacobs, E, Kost, GJ, Kary, H, Balla, J, et al.. Point-of-care testing performed by healthcare professionals outside the hospital setting: consensus based recommendations from the IFCC committee on point-of-care testing (IFCC C-POCT). Clin Chem Lab Med 2023;61:1572–9. https://doi.org/10.1515/cclm-2023-0502.Search in Google Scholar PubMed
6. Point of Care Testing Implementation Guide. Australasian Association of Clinical Biochemists (AACB) 2019. https://www.aacb.asn.au/AACB/AACB/Resources/Guidelines-and-Position-Statements.aspx?hkey=9aa4a6c0-6aeb-4bc2-a6fc-38074ebab935 [Accessed 19 Feb 2025].Search in Google Scholar
7. Guidance: management and use of IVD point of care test devices. GOV.UK; 2021. https://www.gov.uk/government/publications/in-vitro-diagnostic-point-of-care-test-devices/management-and-use-of-ivd-point-of-care-test-devices.[Accessed 19 Feb 2025].Search in Google Scholar
8. Hansen-Nord, G, Olsen, PP, Rem, J, Guldager, N, Holm, NU, Junge, OF, et al.. Kvalitetssikring og kvalitetskrav til laboratoriemedicinske aktiviteter i almen praksis. Regionernes Lønnings- og takstnævn (RLTN) og praktiserende lægers organisation (PLO). Danmark: Konsensusdokument; 2010.Search in Google Scholar
9. Venner, AA, Beach, LA, Shea, JL, Knauer, MJ, Huang, Y, Fung, AWS, et al.. Quality assurance practices for point of care testing programs: recommendations by the Canadian society of clinical chemists point of care testing interest group. Clin Biochem 2021;88:11–7. https://doi.org/10.1016/j.clinbiochem.2020.11.008.Search in Google Scholar PubMed
10. Stavelin, A, Sandberg, S. Harmonization activities of Noklus – a quality improvement organization for point-of-care laboratory examinations. Clin Chem Lab Med 2019;57:106–14. https://doi.org/10.1515/cclm-2018-0061.Search in Google Scholar PubMed
11. Sandberg, S, Fraser, CG, Horvath, AR, Jansen, R, Jones, G, Oosterhuis, W, et al.. Defining analytical performance specifications: consensus statement from the 1st strategic conference of the European federation of clinical chemistry and laboratory medicine. Clin Chem Lab Med 2015;53:833–5. https://doi.org/10.1515/cclm-2015-0067.Search in Google Scholar PubMed
12. Sandberg, S, Carobene, A, Bartlett, B, Coskun, A, Fernandez-Calle, P, Jonker, N, et al.. Biological variation: recent development and future challenges. Clin Chem Lab Med 2023;61:741–50. https://doi.org/10.1515/cclm-2022-1255.Search in Google Scholar PubMed
13. Aarsand, AK, Fernandez-Calle, P, Webster, C, Coskun, A, Gonzales-Lao, E, Diaz-Garzon, J, et al.. The EFLM biological variation database. https://biologicalvariation.eu/ [Accessed 19 Aug 2024].Search in Google Scholar
14. CLSI. Point-of-Care coagulation testing and anticoagulation monitoring, 2nd ed. USA: Clinical and Laboratory Standards Institute; 2020, CLSI Guideline POCT14.Search in Google Scholar
15. Stavelin, A, Ronneseth, E, Gidske, G, Solsvik, AE, Sandberg, S. Using three external quality assurance schemes to achieve equivalent international normalized ratio results in primary and secondary healthcare. Clin Chem Lab Med 2023;61:419–26. https://doi.org/10.1515/cclm-2022-1080.Search in Google Scholar PubMed
16. Cole, J, Sampson, M, van Deventer, HE, Remaley, AT. Reducing lipid panel error allowances to improve the accuracy of cardiovascular risk stratification. Clin Chem 2023;69:1145–54. https://doi.org/10.1093/clinchem/hvad109.Search in Google Scholar PubMed
17. Noklus. The Norwegian organization for quality improvement of laboratory examinations. https://www.noklus.no/diagnostikk-av-diabetes/ [Accessed 27 Mar 2025].Search in Google Scholar
18. Panteghini, M. Analytical performance specifications for combined uncertainty budget in the implementation of metrological traceability. Clin Chem Lab Med 2024;62:1497–504. https://doi.org/10.1515/cclm-2023-1325.Search in Google Scholar PubMed
19. Braga, F, Panteghini, M. Performance specifications for measurement uncertainty of common biochemical measurands according to milan models. Clin Chem Lab Med 2021;59:1362–8. https://doi.org/10.1515/cclm-2021-0170.Search in Google Scholar PubMed
20. van den Besselaar, A, Stavelin, A, Kitchen, S, Bryant, M, Tripodi, A, Scalambrino, E, et al.. Defining a metrologically traceable and sustainable calibration hierarchy of international normalized ratio for monitoring of vitamin K antagonist treatment in accordance with international organization for standardization (ISO) 17511:2020 standard: communication from the international Federation of clinical chemistry and laboratory Medicine-SSC/ISTH working group on prothrombin time/international normalized ratio standardization. J Thromb Haemostasis 2024;22:1236–48. https://doi.org/10.1016/j.jtha.2023.12.013.Search in Google Scholar PubMed
21. Gidske, G, Sandberg, S, Fossum, AL, Binder, S, Langsjoen, EC, Solsvik, AE, et al.. Point-of-care testing in primary healthcare: a scoring system to determine the frequency of performing internal quality control. Clin Chem Lab Med 2022;60:740–7. https://doi.org/10.1515/cclm-2021-1258.Search in Google Scholar PubMed
22. Stavelin, A, Petersen, PH, Solvik, U, Sandberg, S. Internal quality control of prothrombin time in primary care: comparing the use of patient split samples with lyophilised control materials. Thromb Haemost 2009;102:593–600. https://doi.org/10.1160/th09-02-0082.Search in Google Scholar
23. Holt, H, Freedman, DB. Internal quality control in point-of-care testing: where’s the evidence? Ann Clin Biochem 2016;53:233–9. https://doi.org/10.1177/0004563215615148.Search in Google Scholar PubMed
24. Jepsen, SY, Larsen, JB, Christensen, TD, Grove, EL, Maegaard, M, Hvas, AM. Warfarin monitoring and interference by lupus anticoagulant in patients with antiphospholipid syndrome. Thromb Res 2022;211:127–32. https://doi.org/10.1016/j.thromres.2022.01.021.Search in Google Scholar PubMed
25. Oosterhuis, WP, Bayat, H, Armbruster, D, Coskun, A, Freeman, KP, Kallner, A, et al.. The use of error and uncertainty methods in the medical laboratory. Clin Chem Lab Med 2018;56:209–19. https://doi.org/10.1515/cclm-2017-0341.Search in Google Scholar PubMed
26. Panteghini, M, Sandberg, S. Total error vs. measurement uncertainty: the match continues. Clin Chem Lab Med 2016;54:195–6. https://doi.org/10.1515/cclm-2015-1036.Search in Google Scholar PubMed
© 2025 the author(s), published by De Gruyter, Berlin/Boston
This work is licensed under the Creative Commons Attribution 4.0 International License.