To the Editor,
Searching the literature for quality demands or biological variation for arterial partial pressure of oxygen (pO2), we found no criteria that could be used for defining the necessary limits for neither bias nor Coefficient of Variation (CV%) of pO2.
Usually, according to the Stockholm criteria of 1999 and further revised in Milan 2014 [1, 2], the criteria should preferably be based on medical needs according to application, or if that cannot be done, according to intra- and interindividual variation.
When we constructed the Tempus600 system [3] and later documented the effect on biochemical parameters by transporting sample tubes through the system, we experienced an unacceptable effect on true values on especially O2-related parameters [4].
As no data is yet available through the Westgard: (Ricos et al.) or EFLMs databases on Coefficient of Variations and biological variation of pO2, we discussed with our clinicians how they use the pO2 in clinical application. Based on these clinical judgements and through the known information on the normal ranges, we defined the precondition for our considerations for pO2 quality specifications in arterial blood as follows:
The normal reference range is 11–14 kPa
An acceptable value in an emergency patient is down to 10 kPa
The definition of respiratory failure is a value below 8 kPa
A life-threatening value is 5 kPa
According to the clinician’s decision limits for acting on changes in pO2, the following could be deduced:
It is clinically relevant to detect a decrease from 10 to 9 and from 9 to 8 kPa
It is clinically necessary to detect a decrease of 0.5 kPa at the level of 5 kPa
An acceptable total error at a level of 5 kPa is required
We assumed that in such situations the biological variation is unimportant, as the actual value is what determines the patient’s situation and the clinician’s corrective medical interventions. Therefore, at these critical values it is important to establish as accurately as possible the value, but also to detect an improvement or deterioration over time.
Consequently, the considerations on acceptable variation comprise both sampling, other pre-analytical and transport sources of variation, and the analytical variation itself. The measurement should thus be capable of detecting a change in pO2 from 10 to 9 kPa with at least a 95 % certainty.
In evaluating this decrease, the bias is of no importance if measured on the same instrument and thus, the critical difference mandatory to be able to detect with 95 % certainty is 1 kPa. Using analysis of variance, the difference should be within 2.77 times the total analytical and pre-analytical variation. This means that pre- and analytical variation of 0.36 kPa (1 SD) will fulfill this. At the pO2 level of 9 kPa, this is 0.36 × 100/9=a CV% corresponding to 4 %.
The maximal total analytical and pre-analytical variation are thus defined to be less than 4 %.
This defined minimum quality criteria for the pO2 measurement will consequently also enable a measurement of a decrease from 8 to 5 kPa with a certainty of above 99 %, fulfilling the ability to detect this life-threatening value.
For the absolute measurement of the value 5 kPa with a total error of <0.5 kPa giving the clinician’s limit for acting (Bullet 6: It is clinical necessary to detect a decrease of 0.5 kPa at the level of 5 kPa), the bias plus 1.65 times the total analytical variation should not exceed 0.5 kPa.
(Bullet 7: An acceptable total error at a level of 5 kPa is required).
This can be achieved with a maximal bias of 0.2 kPa and a total coefficient of variation of 4 % (0.2 + 1.65 × (4 % of 5.0=0.2) ∼ 0.5). For the absolute measurement of a value, this means that the total coefficient of variation of sampling, transport, other pre-analytical sources, and the analytical uncertainty should be below 0.36 kPa at level 9 kPa and below 4 % through the range from 1 to 14 kPa.
This means that at 5 kPa, a significant change of 2.77 × 4 % of 5 kPa=2.77 × 0.2=0.55 kPa is detectable, largely fulfilling the clinical needs for the application.
If these criteria of maximum bias of 0.2 kPa and a total coefficient of variation below 4 % are fulfilled, the quality specifications are satisfactory for all applications for arterial blood measurements of oxygen tension.
For the measurement of venous values, at least the same specifications are required.
In conclusion, we suggest that the quality demands for arterial partial pressure of oxygen are:
Bias<0.2 kPa
CV<4 %
TE<0.5 kPa at a level below 5 kPa and TE<0.6 kPa at a level above 5 kPa
TE<0.2 + (1.65 × 4 % of the actual level).
We have implemented these quality demands in our laboratories until more detailed scientifically based data are available.
The calculations and resulting specifications have been distributed to members of the Nordic lab Network CCL (Laboratory medicine in Nordic countries) for criticism, and remarks from Anders Kallner were subsequently incorporated.
-
Research ethics: Not applicable.
-
Informed consent: Not applicable.
-
Author contributions: The authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: The authors state no conflict of interest.
-
Research funding: None declared.
-
Data availability: Not applicable.
References
1. Fraser, CG. General strategies to set quality specifications for reliability performance characteristics. Scand J Clin Lab Invest 1999;59:487–90. https://doi.org/10.1080/00365519950185210.Search in Google Scholar PubMed
2. Sandberg, S, Fraser, CG, Horvath, AR, Jansen, R, Jones, G, Oosterhuis, W, et al.. Defining analytical performance specifications: consensus statement from the 1st strategic conference of the European federation of clinical chemistry and laboratory medicine. Clin Chem Lab Med 2015;53:833–5. https://doi.org/10.1515/cclm-2015-0067.Search in Google Scholar PubMed
3. Hastrup, J CH, Skov Madsen, J, Backer Mogensen, C, Brandslund, I. Stabilitet af biokemiske komponenter i blodprøver der transporteres i et nyt dedikeret rørsystem, Tempus 600. In: Klinisk biokemi i norden. 2013:vol 25, 38–46 pp.Search in Google Scholar
4. Andersen, IB, Mogensen, N, Brandslund, I. Stability of biochemical components in blood samples transported by Tempus600/sysmex GLP robot reception system. J Appl Lab Med 2017;1:376–86. https://doi.org/10.1373/jalm.2016.021188.Search in Google Scholar PubMed
© 2023 the author(s), published by De Gruyter, Berlin/Boston
This work is licensed under the Creative Commons Attribution 4.0 International License.
Articles in the same Issue
- Frontmatter
- Editorial
- Unraveling the mystery of blood groups and COVID-19
- Reviews
- Serum hepcidin levels in chronic liver disease: a systematic review and meta-analysis
- Platelet distribution width (PDW) as a significant correlate of COVID-19 infection severity and mortality
- Mini Reviews
- ABO blood group-related mechanism of infection of SARS-CoV-2: an overview of systematic reviews
- Opinion Paper
- Personalized laboratory medicine in the digital health era: recent developments and future challenges
- Guidelines and Recommendations
- Algorithm of differential diagnosis of anemia involving laboratory medicine specialists to advance diagnostic excellence
- General Clinical Chemistry and Laboratory Medicine
- Rescaling creatinine makes GFR estimation equations generally applicable across populations – validation results for the Lund-Malmö equation in a French cohort of sub-Saharan ancestry
- Periodic verification of results’ comparability between several analyzers: experience in the application of the EP31-A-IR guideline
- Machine learning to optimize cerebrospinal fluid dilution for analysis of MRZH reaction
- Diagnostic performance of automated red cell parameters in predicting bone marrow iron stores
- Reliability of hemoglobin A2 value as measured by the Premier Resolution system for screening of β-thalassemia carriers
- Amino acid sequence homology of monoclonal serum free light chain dimers and tissue deposited light chains in AL amyloidosis: a pilot study
- Development of high-performance point-of-care aqueous VEGF detection system and proof-of-concept validation in RVO patients
- Detection rate of IGF-1 variants and their implication to protein binding: study of over 240,000 patients
- Analysis of a second-tier test panel in dried blood spot samples using liquid chromatography-tandem mass spectrometry in Catalonia’s newborn screening programme
- Targeted quantitative lipidomic uncovers lipid biomarkers for predicting the presence of compensated cirrhosis and discriminating decompensated cirrhosis from compensated cirrhosis
- Reference Values and Biological Variations
- Establishment of reference intervals for free light chains and immunoglobulins in Saudi population
- Cancer Diagnostics
- A predictive and prognostic model for surgical outcome and prognosis in ovarian cancer computed by clinico-pathological and serological parameters (CA125, HE4, mesothelin)
- M-protein diagnostics in multiple myeloma patients using ultra-sensitive targeted mass spectrometry and an off-the-shelf calibrator
- Cardiovascular Diseases
- Bioactive adrenomedullin (bio-ADM) is associated with endothelial dysfunction in infants and children with complex congenital heart disease undergoing open-heart surgery on cardiopulmonary bypass
- Infectious Diseases
- Monocyte distribution width as an early predictor of short-term outcome in adult patients with sepsis
- Analytical and clinical evaluations of SNIBE Maglumi chemiluminescent immunoassay for the detection of SARS-CoV-2 antigen in salivary samples
- Letters to the Editor
- Predicting hemoglobinopathies using ChatGPT
- Managing the Quality Control of multiple instruments
- Lipid droplets may interfere with urinary red blood cell and crystal counts by urinary flow cytometry
- Assessment of the lipemia index determined by the Atellica CH 930 analyzer for the detection of monoclonal immunoglobulins
- Concerning quality demands of arterial partial pressure of oxygen
- Navigating between perpendicular drop and tangent skimming methods for M-protein quantification: a call for clarification of guidelines
- Early detection of peripheral invasive candidiasis further to cytographic interferences in Sysmex XN-9000 hematology analyzer
- Congress Abstracts
- Annual meeting of the Royal Belgian Society of Laboratory Medicine: “Symphony of the Heart”
Articles in the same Issue
- Frontmatter
- Editorial
- Unraveling the mystery of blood groups and COVID-19
- Reviews
- Serum hepcidin levels in chronic liver disease: a systematic review and meta-analysis
- Platelet distribution width (PDW) as a significant correlate of COVID-19 infection severity and mortality
- Mini Reviews
- ABO blood group-related mechanism of infection of SARS-CoV-2: an overview of systematic reviews
- Opinion Paper
- Personalized laboratory medicine in the digital health era: recent developments and future challenges
- Guidelines and Recommendations
- Algorithm of differential diagnosis of anemia involving laboratory medicine specialists to advance diagnostic excellence
- General Clinical Chemistry and Laboratory Medicine
- Rescaling creatinine makes GFR estimation equations generally applicable across populations – validation results for the Lund-Malmö equation in a French cohort of sub-Saharan ancestry
- Periodic verification of results’ comparability between several analyzers: experience in the application of the EP31-A-IR guideline
- Machine learning to optimize cerebrospinal fluid dilution for analysis of MRZH reaction
- Diagnostic performance of automated red cell parameters in predicting bone marrow iron stores
- Reliability of hemoglobin A2 value as measured by the Premier Resolution system for screening of β-thalassemia carriers
- Amino acid sequence homology of monoclonal serum free light chain dimers and tissue deposited light chains in AL amyloidosis: a pilot study
- Development of high-performance point-of-care aqueous VEGF detection system and proof-of-concept validation in RVO patients
- Detection rate of IGF-1 variants and their implication to protein binding: study of over 240,000 patients
- Analysis of a second-tier test panel in dried blood spot samples using liquid chromatography-tandem mass spectrometry in Catalonia’s newborn screening programme
- Targeted quantitative lipidomic uncovers lipid biomarkers for predicting the presence of compensated cirrhosis and discriminating decompensated cirrhosis from compensated cirrhosis
- Reference Values and Biological Variations
- Establishment of reference intervals for free light chains and immunoglobulins in Saudi population
- Cancer Diagnostics
- A predictive and prognostic model for surgical outcome and prognosis in ovarian cancer computed by clinico-pathological and serological parameters (CA125, HE4, mesothelin)
- M-protein diagnostics in multiple myeloma patients using ultra-sensitive targeted mass spectrometry and an off-the-shelf calibrator
- Cardiovascular Diseases
- Bioactive adrenomedullin (bio-ADM) is associated with endothelial dysfunction in infants and children with complex congenital heart disease undergoing open-heart surgery on cardiopulmonary bypass
- Infectious Diseases
- Monocyte distribution width as an early predictor of short-term outcome in adult patients with sepsis
- Analytical and clinical evaluations of SNIBE Maglumi chemiluminescent immunoassay for the detection of SARS-CoV-2 antigen in salivary samples
- Letters to the Editor
- Predicting hemoglobinopathies using ChatGPT
- Managing the Quality Control of multiple instruments
- Lipid droplets may interfere with urinary red blood cell and crystal counts by urinary flow cytometry
- Assessment of the lipemia index determined by the Atellica CH 930 analyzer for the detection of monoclonal immunoglobulins
- Concerning quality demands of arterial partial pressure of oxygen
- Navigating between perpendicular drop and tangent skimming methods for M-protein quantification: a call for clarification of guidelines
- Early detection of peripheral invasive candidiasis further to cytographic interferences in Sysmex XN-9000 hematology analyzer
- Congress Abstracts
- Annual meeting of the Royal Belgian Society of Laboratory Medicine: “Symphony of the Heart”