To the Editor,

It is widely believed that return to normalcy after the COVID-19 pandemic will include widespread testing for both active (presence of viral nucleic acids) and recovered (presence of anti-SARS CoV2 immunoglobulins) infections [1]. The latter can inform if individuals were exposed to the virus and if they presumably developed immunity. The required analytical performance characteristics of COVID-19 antibody (Ab) tests have not been discussed in detail. Suboptimal analytical performance can create confusion and may lead to false reassurances, especially when carried out on large populations that have yet to be exposed to the virus and in the absence of a gold standard comparative method [2]. Here, we analyze the importance of assay specificity and calculate the required threshold for the tests to provide clinically relevant information.

Most discrete clinical tests provide two types of results: positive or negative for a specific parameter of interest. Positive results can be true positives (TP; positive with assayed parameter present) or false positives (FP; positive without assayed parameter present). Negative results can be true negatives (TN; negative without assayed parameter present) or false negatives (FN; negative with assayed parameter present). The analytical requirements for the adoption of a clinical test depend on its intended use. It is widely claimed that COVID-19 serological Ab tests could be used for the safe return of presumably immune people to normalcy. In this case, FN Ab results are almost inconsequential as the tested individuals will be told that they are negative and thus have no presumed immunity, when, in fact, they do, leading at most to some unnecessary extension of their isolation practices, but not viral spread. But FP results are consequential as these individuals may be reassured of presumed immunity that they do not have, rendering them vulnerable to infection and viral spread. The rate of FP results depends on the analytical specificity of the assay.

The analytical parameter that reflects the required analytical specificity in order to adopt such a test or not (based on the FP/FN ratio) is the positive predictive value (PPV). The % PPV reflects somebody’s chances of having the disease if the test is positive and equals: PPV=TP/(TP+FP)×100. The required PPV value depends on the intended use of the test and the presumed risk/benefit value of its application. The PPV is dependent not only on the sensitivity and specificity of the test but also on the disease prevalence. Below, we present simulation data for COVID-19 infection based on two US prevalence values. The one at 0.2% (one infected per 500 individuals) and another one at 1% (one infected per 100 individuals) are based on preliminary estimations of prevalence in some countries (we remind that prevalence varies widely among countries, regions, cities, etc. and is continuously changing). We also fixed the sensitivity of the test to 100% (no FN, as a best case scenario) and varied the specificity from 90% onward, calculating the PPV in each case (Table 1). These data show that in order to achieve a minimum of 90% PPV (allowing one FP out of every nine TP), the analytical specificity of the COVID-19 Ab test must exceed 99.9% in both scenarios. With most clinical tests, such exquisite specificities are quite rare and particularly challenging to achieve.

Table 1:

Positive predictive value of a test with a 100% sensitivity and variable specificity, at two disease prevalences.

% Specificity at 100% sensitivityCases per 100,000 peopleTrue positives (TP)True negatives (TN)False positives (FP)False negatives (FN)Positive predictive value (PPV), %
Prevalence 1:500
 9020020089,820998002.0
 9520020094,810499003.8
 9820020097,804199609.1
 9920020098,802998016.6
 99.520020099,301499028.6
 99.920020099,700100066.9
 99.9520020099,75050080
Prevalence 1:100
 901000100089,100990009.1
 951000100094,0504950016.8
 981000100097,0201980033.6
 991000100098,010990050.2
 99.51000100098,505495069
 99.91000100098,90199090.1
 99.951000100098,95050095.2
  1. The % PPV was calculated as TP/(TP+FP)×100. For comments see text.

We urge COVID-19 Ab kit manufacturers to check thoroughly the specificity of their COVID-19 Ab assays by using sera from patients with various autoimmune conditions (about 5% prevalence in the US population) [3] and patients positive for heterophilic antibodies [4] (about 1% prevalence in the general population; these patients have serum antibodies with broad and unpredictable specificities against human and animal antibodies) and other infections with diverse viruses (optimally including SARS-CoV and other ‘common-cold’ CoVs), to make sure of no cross-reactivity [5]. Universal standardizations and assay harmonizations would contribute tremendously to minimal analytical variability between kits [6].

The high urgency for new SARS CoV-2 serological tests should not compromise assay quality standards. FP results could lead to vital errors both in prevalence prediction studies and in misleadingly reassuring patients of presumed immunity, thus allowing them to return to societal activities at a great risk of getting infected and further spreading the virus.


Corresponding author: Eleftherios P. Diamandis, PhD, MD, Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, Canada; Department of Pathology and Laboratory Medicine, Mount Sinai Hospital, Toronto, Canada; Department of Clinical Biochemistry, University Health Network, Toronto, Canada; and Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, 60 Murray St [Box 32], Flr 6 – Rm L6-201-1, Toronto, ON, M5T 3L9, Canada, Phone: +416-586-8443, Fax: +416-619-5521

  1. Research funding: None declared.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

References

1. Petherick A. Developing antibody tests for SARS-CoV-2. Lancet 2020;395:1101–2.10.1016/S0140-6736(20)30788-1Search in Google Scholar

2. Trevethan R. Sensitivity, specificity, and predictive values: foundations, pliabilities, and pitfalls in research and practice. Front Public Health 2017;5:307–12.10.3389/fpubh.2017.00307Search in Google Scholar

3. Cooper GS, Stroehla BC. The epidemiology of autoimmune diseases. Autoimmun Rev 2003;2:119–25.10.1016/S1568-9972(03)00006-5Search in Google Scholar

4. Preissner CM, Dodge LA, O’Kane DJ, Singh RJ, Grebe SK. Prevalence of heterophilic antibody interference in eight automated tumor marker immunoassays. Clin Chem 2005;51: 208–10.10.1373/clinchem.2004.040501Search in Google Scholar PubMed

5. Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev 2004;25:105–20.Search in Google Scholar

6. Tate J, Myers GL. Harmonization of clinical laboratory test results. eJIFCC 2016;27:5–14.Search in Google Scholar

Received: 2020-04-21
Accepted: 2020-04-22
Published Online: 2020-05-09
Published in Print: 2020-06-25

©2020 Walter de Gruyter GmbH, Berlin/Boston

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. Critical role of laboratory medicine in the global response to the COVID-19 pandemic
  4. Reviews
  5. Hematologic, biochemical and immune biomarker abnormalities associated with severe illness and mortality in coronavirus disease 2019 (COVID-19): a meta-analysis
  6. COVID-19: progression of disease and intravascular coagulation – present status and future perspectives
  7. IFCC Recommendations
  8. Molecular, serological, and biochemical diagnosis and monitoring of COVID-19: IFCC taskforce evaluation of the latest evidence
  9. Biosafety measures for preventing infection from COVID-19 in clinical laboratories: IFCC Taskforce Recommendations
  10. Opinion Papers
  11. The critical role of laboratory medicine during coronavirus disease 2019 (COVID-19) and other viral outbreaks
  12. Potential preanalytical and analytical vulnerabilities in the laboratory diagnosis of coronavirus disease 2019 (COVID-19)
  13. Laboratory diagnostics within a modular hospital at the time of Coronavirus disease 2019 (COVID-19) in Wuhan
  14. Original Articles
  15. Analytical performances of a chemiluminescence immunoassay for SARS-CoV-2 IgM/IgG and antibody kinetics
  16. Comparison of throat swabs and sputum specimens for viral nucleic acid detection in 52 cases of novel coronavirus (SARS-Cov-2)-infected pneumonia (COVID-19)
  17. Routine blood tests as a potential diagnostic tool for COVID-19
  18. Laboratory predictors of death from coronavirus disease 2019 (COVID-19) in the area of Valcamonica, Italy
  19. The hemocyte counts as a potential biomarker for predicting disease progression in COVID-19: a retrospective study
  20. Prominent changes in blood coagulation of patients with SARS-CoV-2 infection
  21. The value of urine biochemical parameters in the prediction of the severity of coronavirus disease 2019
  22. Letters to the Editor
  23. COVID-19 infections are also affected by human ACE1 D/I polymorphism
  24. No significant correlation between ACE Ins/Del genetic polymorphism and COVID-19 infection
  25. ACE Ins/Del genetic polymorphism and epidemiological findings in COVID-19
  26. Laboratory abnormalities in patients with COVID-2019 infection
  27. Laboratory abnormalities in children with novel coronavirus disease 2019
  28. Clinical laboratory and SARS-CoV-2 infection: where do we stand?
  29. Clinical chemistry tests for patients with COVID-19 – important caveats for interpretation
  30. Antibody tests for COVID-19: drawing attention to the importance of analytical specificity
  31. Erythrocyte sedimentation rate is associated with severe coronavirus disease 2019 (COVID-19): a pooled analysis
  32. One disease, different features: COVID-19 laboratory and radiological findings in three Italian patients
  33. Decreased “WBC*LYM” was observed in SARS-CoV-2-infected patients from a fever clinic in Wuhan
  34. Assessment of immune response to SARS-CoV-2 with fully automated MAGLUMI 2019-nCoV IgG and IgM chemiluminescence immunoassays
  35. Coinfection of SARS-CoV-2 and multiple respiratory pathogens in children
  36. The friendly use of chloroquine in the COVID-19 disease: a warning for the G6PD-deficient males and for the unaware carriers of pathogenic alterations of the G6PD gene
  37. Potential interference of hydroxychloroquine-glucuronide metabolite on therapeutic drug monitoring of hydroxychloroquine using a mass spectrometry detector
Downloaded on 23.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2020-0554/html?lang=en
Scroll to top button