A novel method based solely on field programmable gate array (FPGA) units enabling measurement of time and charge of analog signals in positron emission tomography (PET)
-
Marek Pałka
, Paweł Moskal
, Tomasz Bednarski , Piotr Białas , Eryk Czerwiński , Łukasz Kapłon , Andrzej Kochanowski , Grzegorz Korcyl , Jakub Kowal , Paweł Kowalski , Tomasz Kozik , Wojciech Krzemień , Marcin Molenda , Szymon Niedźwiecki , Monika Pawlik , Lech Razyński , Zbigniew Rudy , Piotr Salabura , Neha Gupta-Sharma , Michał Silarski , Artur Słomski , Jerzy Smyrski , Adam Strzelecki , Wojciech Wiślicki , Marcin Zieliński und Natalia Zoń
Abstract
This article presents an application of a novel technique for precise measurements of time and charge based solely on a field programmable gate array (FPGA) device for positron emission tomography (PET). The described approach simplifies electronic circuits, reduces the power consumption, lowers costs, merges front-end electronics with digital electronics, and also makes more compact final design. Furthermore, it allows to measure time when analog signals cross a reference voltage at different threshold levels with a very high precision of ~15 ps (rms) and thus enables sampling of signals in a voltage domain.
Acknowledgments
We acknowledge the technical and administrative support by M. Adamczyk, T. Gucwa-Ryś, A. Heczko, M. Kajetanowicz, G. Konopka-Cupiał, J. Majewski, W. Migdał, and A. Misiak and the financial support by the Polish National Center for Development and Research through grant INNOTECH-K1/IN1/64/159174/NCBR/12, the Foundation for Polish Science through MPD programme, and the EU and MSHE Grant No. POIG.02.03.00-161 00-013/09.
Conflict of interest statement
Authors’ conflict of interest disclosure: The authors stated that there are no conflicts of interest regarding the publication of this article. Research funding played no role in thestudy design; in the collection, analysis, and interpretationof data; in the writing of the report; or in the decision tosubmit the report for publication.
Research funding: None declared.
Employment or leadership: None declared.
Honorarium: None declared.
References
1. Moskal P, Salabura P, Silarski M, Smyrski J, Zdebik J, Zieliński M. Novel detector systems for the positron emission tomography. Bio-Algorithms Med-Syst 2011;7:73; arXiv: 1305.5187.Suche in Google Scholar
2. Moskal P, Bednarski T, Białas P, Ciszewska M, Czerwiński E, Heczko A, et al. Strip-PET: a novel detector concept for the TOF-PET scanner. Nucl Med Rev 2012;15:C68; arXiv: 1305.5562.Suche in Google Scholar
3. Moskal P, Bednarski T, Białas P, Ciszewska M, Czerwiński E, Heczko A, et al. TOF-PET detector concept based on organic scintillators. Nucl Med Rev 2012;15:C81; arXiv: 1305.5559.Suche in Google Scholar
4. Wu J, Hansen S, Shi Z. ADC and TDC implemented using FPGA. FERMILAB-CONF-07-601-E.Suche in Google Scholar
5. Kalisz J, Szplet R, Pasierbinski J, Poniecki A. Field programmable gate array based time-to-digital converter with 200-ps resolution. IEEE Trans Instrum Measure 1997;46:51–5.10.1109/19.552156Suche in Google Scholar
6. Wu J, Shi Z. The 10-ps wave union TDC: improving FPGA TDC resolution beyond its cell delay. IEEE Nuclear Science Symposium Conference, October 19–25, 2008.10.1109/NSSMIC.2008.4775079Suche in Google Scholar
7. Traxler M, Bayer E, Kajetanowicz M, Korcyl G, Maier L, Michel J, et al. A compact system for high precision time measurements (<14 ps RMS) and integrated data acquisition for a large number of channels. J Instrum 2011;6:C12004.10.1088/1748-0221/6/12/C12004Suche in Google Scholar
©2014 by Walter de Gruyter Berlin/Boston
Artikel in diesem Heft
- Frontmatter
- Original Articles
- 3D PET image reconstruction based on the maximum likelihood estimation method (MLEM) algorithm
- List-mode reconstruction in 2D strip PET
- Calibration of photomultipliers gain used in the J-PET detector
- A novel method for calibration and monitoring of time synchronization of TOF-PET scanners by means of cosmic rays
- Plastic scintillators for positron emission tomography obtained by the bulk polymerization method
- J-PET analysis framework for the prototype TOF-PET detector
- Trigger-less and reconfigurable data acquisition system for positron emission tomography
- A novel method based solely on field programmable gate array (FPGA) units enabling measurement of time and charge of analog signals in positron emission tomography (PET)
- Review
- Positron emission tomography (PET) in oncology: current applications and future perspectives
Artikel in diesem Heft
- Frontmatter
- Original Articles
- 3D PET image reconstruction based on the maximum likelihood estimation method (MLEM) algorithm
- List-mode reconstruction in 2D strip PET
- Calibration of photomultipliers gain used in the J-PET detector
- A novel method for calibration and monitoring of time synchronization of TOF-PET scanners by means of cosmic rays
- Plastic scintillators for positron emission tomography obtained by the bulk polymerization method
- J-PET analysis framework for the prototype TOF-PET detector
- Trigger-less and reconfigurable data acquisition system for positron emission tomography
- A novel method based solely on field programmable gate array (FPGA) units enabling measurement of time and charge of analog signals in positron emission tomography (PET)
- Review
- Positron emission tomography (PET) in oncology: current applications and future perspectives