Home Evaluation of the Konelab 20XT clinical chemistry analyzer
Article
Licensed
Unlicensed Requires Authentication

Evaluation of the Konelab 20XT clinical chemistry analyzer

  • Nataša Stojanović , Dunja Rogić and Ana Stavljenić-Rukavina
Published/Copyright: June 11, 2014

Abstract

The Konelab 20XT (Thermo Electron Oy, Finland) is a clinical chemistry analyzer for colorimetric, immunoturbidimetric and ion-selective electrode measurements. The aim of our work was to evaluate the analytical performances of the Konelab 20XT according to the European Clinical Chemistry Laboratory Standards Guidelines. A total of 30 analytes including substrates, enzymes, electrolytes and specific proteins were tested. Investigation results showed low imprecision (within-run coefficient of variation was below 3.5% and between-day coefficient of variation was less than 2.5% for most analytes at all three levels studied) and acceptable accuracy of the analyzer. No significant sample- or reagent-related carry-over was found. It was demonstrated that the analytical system operates within the claimed linearity ranges. The results compared well with those obtained by instruments routinely used in our laboratory (Olympus AU2700, Behring Nephelometer II). In general, the data on interference by hemoglobin, hyperbilirubinemia and turbidity are in accordance with known facts. However, slight hemolysis was found to interfere with the alkaline phosphatase (ALP) assay and mild lipemia affected the glucose assay. The Konelab 20XT is an easy-to-use analyzer that is suitable for routine and emergency analyses in small laboratories.


Corresponding author: Dunja Rogić, PhD, Clinical Institute of Laboratory Diagnostics, Clinical Hospital Center Zagreb, 10000 Zagreb, Croatia Phone: +385-1-23-88-380, Fax: +385-1-23-12-079,

References

1 ECCLS. Guidelines for the evaluation of analyzers in clinical chemistry. ECCLS document, vol. 3. Berlin, Germany: Beuth Verlag GmbH, 1986:1–32. Search in Google Scholar

2 Ferrero C, Modenese A, Bonini PA, Ceriotti F, Carobene A, Franzin M, et al. Multicentre evaluation of KONE Optima Analysis System. Clin Chem Lab Med 1998; 36: 475–84. 10.1515/CCLM.1998.080Search in Google Scholar PubMed

3 NCCLS. Evaluation of precision performance of clinical chemistry devices: tentative guidelines. NCCLS document EP5-T2, 2nd ed., 1992. Search in Google Scholar

4 Fraser CG, Hyltoft Petersen P, Ricos C, Haeckel R. Proposed quality specifications for the imprecision and inaccuracy of analytical systems for clinical chemistry. Eur J Clin Chem Biochem 1992; 30: 311–7. Search in Google Scholar

5 Ricoś C, Alvarez V, Cava F, García-Lario, JV, Hernández A, Himénez CV, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest 1999; 59: 491–500. 10.1080/00365519950185229Search in Google Scholar PubMed

6 Westgard. Biological variation database and desirable quality specification. The 2001 update. www.westgard.com. Search in Google Scholar

7 Broughton PM, Glowenlock AH, McCormack JJ, Neill DW. A revised scheme for the evaluation of automatic instruments. Ann Clin Biochem 1974; 11: 207–18. 10.1177/000456327401100164Search in Google Scholar PubMed

8 NCCLS. Evaluation of the linearity of quantitative analytical methods. NCCLS document EP6-P2. Proposed guideline, 2nd ed., 2002. Search in Google Scholar

9 Passing H, Bablok W. A new biometrical procedure for testing the quality of measurement from two different analytical methods. Part I. J Clin Chem Clin Biochem 1983; 21: 709–20. Search in Google Scholar

10 Glick MR, Ryder KW, Jackson SA. Graphical comparison of interferences in clinical chemistry instrumentation. Clin Chem 1986; 32: 470–5. 10.1093/clinchem/32.3.470Search in Google Scholar

11 Guder WG, Narayanan S, Wisser H, Zuwta B. List of analytes. Preanalytical variables. In: Guder WG, Narayanan S, Wisser H, Zuwta B, editors. From the patient to the laboratory, 2nd ed. Darmstadt, Germany: GIT Verlag GmbH, 2001:8,16,20. Search in Google Scholar

12 Young DS. Effects of preanalytical variables on clinical laboratory tests. 2nd ed. Washington, USA: AACC Press, 1997:3-306–8. Search in Google Scholar

13 Painter PC. Method evaluation and technology assessment. In: Lewandrowski K, editor. Clinical chemistry laboratory management and clinical correlations. New York, USA: Lippincot Williams & Wilkins, 2002:130–2. Search in Google Scholar

14 Bonini P, Ceriotti F, Keller F, Brauer P, Stolz H, Pascual C, et al. Multicentre evaluation of the Boehringer Mannheim/Hitachi 747 analysis system. Eur J Clin Chem Biochem 1992; 30: 881–99. Search in Google Scholar

Received: 2005-2-2
Accepted: 2005-4-4
Published Online: 2014-6-11
Published in Print: 2005-6-1

© Walter de Gruyter Berlin New York

Articles in the same Issue

  1. Spectrophotometric determination of urinary iodine by the Sandell-Kolthoff reaction subsequent to dry alkaline ashing. Results from the Czech Republic in the period 1994–2002
  2. Comparison of iodine contents in gastric cancer and surrounding normal tissues
  3. Effects of urine dilution on quantity, size and aggregation of calcium oxalate crystals induced in vitro by an oxalate load
  4. Is idiopathic recurrent calcium urolithiasis in males a cellular disease? Laboratory findings in plasma, urine and erythrocytes, emphasizing the absence and presence of stones, oxidative and mineral metabolism: an observational study
  5. Circulating levels of nitrated apolipoprotein A-I are increased in type 2 diabetic patients
  6. Apolipoprotein A5 gene polymorphism –1131T→C: association with plasma lipids and type 2 diabetes mellitus with coronary heart disease in Chinese
  7. A simple method for estimating equilibrium constants for serum testosterone binding resulting in an optimal free testosterone index for use in elderly men
  8. Standardized quantification of circulating peripheral tumor cells from lung and breast cancer
  9. Plasma homocysteine and markers for oxidative stress and inflammation in patients with coronary artery disease – a prospective randomized study of vitamin supplementation
  10. Comparison of methods for calculating serum osmolality: multivariate linear regression analysis
  11. Radioimmunoassay for plasma C-type natriuretic peptide determination: a methodological evaluation
  12. Evaluation of the Konelab 20XT clinical chemistry analyzer
  13. A guide to defining the competence required of a consultant in clinical chemistry and laboratory medicine
  14. Time course of matrix metalloproteinase-9 and tissue inhibitor of metalloproteinase-1 in patients presenting with acute coronary syndrome
  15. Screening for gestational diabetes mellitus
  16. Systematic terminology for specialities and disciplines related to clinical laboratory
Downloaded on 25.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2005.111/html?lang=en
Scroll to top button