Evaluation of Uncertainty of Measurement in Routine Clinical Chemistry - Applications to Determination of the Substance Concentration of Calcium and Glucose in Serum
-
Solveig Linko
Abstract
We studied the uncertainty of measurement for the calcium and glucose (amount of) substance concentrations in serum. The evaluation follows a four-step procedure, which complies with the ISO document Guide to the Expression of Uncertainty in Measurement (GUM). The applications were chosen to represent commonly used measuring systems in medical laboratories. The uncertainty components are quantified using observations of the measuring system, and information from calibration certificates, instrument specifications and literature. The evaluation focuses on the measurement step but empirical terms are used to illustrate how the pre-analytical phase and patient-related issues can be accounted for.
The software GUM Workbench® was used to facilitate calculations and to visualize the importance of each uncertainty component. The combined standard uncertainties (uc) for the measurands were ≤2% including the pre-analytical uncertainty sources. The patient-related source is discussed in relation to clinician's diagnosis and decision-making. The evaluation, as carried out here for calcium and glucose substance concentration measurements, can easily be applied to many other measurands in clinical chemistry.
This work emphasizes that the internal quality control can provide much of the information needed in the uncertainty evaluation, and that external quality assessment (EQA) schemes are important in the control of the uncertainty evaluated by the individual laboratories. Due to statistical and metrological limitations routine EQA schemes should themselves not be used as a means of uncertainty evaluation.
Copyright (c) 2002 by Walter de Gruyter GmbH & Co. KG
Articles in the same Issue
- Fecal Elastase-1 as a Test for Pancreatic Function: a Review
- α2-Macroglobulin Deletion Polymorphism and Plasma Levels in Late Onset Alzheimer's Disease
- High-Speed Detection of the Two Common Paraoxonase Polymorphisms Leu55 → Met and Gln192 → Arg by Real-Time Fluorescence PCR and Melting Curves
- Rapid Detection of Polymorphisms of the Nitric Oxide Cascade
- Salivary Electrolyte Concentrations Are Associated with Cystic Fibrosis Transmembrane Regulator Genotypes
- Granulocyte-Colony Stimulating Factor (G-CSF) and Macrophage-Colony Stimulating Factor (M-CSF) in Colorectal Cancer Patients
- Sialic Acid, Intercellular Adhesion Molecule-1 and Rheumatoid Arthritis: a Study on the Erythrocyte Membrane
- Determination of Organochlorine Pesticide Residues in the Blood of Healthy Individuals
- Reference Values for a Heterogeneous Ferritin Assay and Traceability to the 3rd International Recombinant Standard for Ferritin (NIBSC Code 94/572)
- The Circulating Levels of Cardiac Natriuretic Hormones in Healthy Adults: Effects of Age and Sex
- Plasma Free Amino Acids in Taiwan Chinese: the Effect of Age
- Regression-Based Reference Limits for Urinary Amino Acids in a Pediatric Population
- Evaluation of Uncertainty of Measurement in Routine Clinical Chemistry - Applications to Determination of the Substance Concentration of Calcium and Glucose in Serum
- Comparison of Quick and Owren Prothrombin Time with Regard to the Harmonisation of the International Normalised Ratio (INR) System
- The Accuracy of Calculated Base Excess in Blood
- Aspects of the Development of Clinical Chemistry and Laboratory Medicine in the former East Germany
- University Degree in Clinical Laboratory Sciences
- Process Control Reduces the Laboratory Turnaround Time
- Who Needs Two Different Citrate Concentrations in Sample Tubes Used for Coagulation Tests?
- Quantifying the Capability of Detection of Measurement Procedures in Clinical Laboratory Sciences
- Yet another Example of Lack of Laboratory Expertise in Certain Biomedical Publications
- Counterimmunoelectrophoresis: Fast, Easy and Cost-effective Method for the Detection of Autoantibodies to Intracellular Antigens
- Meetings and Awards
Articles in the same Issue
- Fecal Elastase-1 as a Test for Pancreatic Function: a Review
- α2-Macroglobulin Deletion Polymorphism and Plasma Levels in Late Onset Alzheimer's Disease
- High-Speed Detection of the Two Common Paraoxonase Polymorphisms Leu55 → Met and Gln192 → Arg by Real-Time Fluorescence PCR and Melting Curves
- Rapid Detection of Polymorphisms of the Nitric Oxide Cascade
- Salivary Electrolyte Concentrations Are Associated with Cystic Fibrosis Transmembrane Regulator Genotypes
- Granulocyte-Colony Stimulating Factor (G-CSF) and Macrophage-Colony Stimulating Factor (M-CSF) in Colorectal Cancer Patients
- Sialic Acid, Intercellular Adhesion Molecule-1 and Rheumatoid Arthritis: a Study on the Erythrocyte Membrane
- Determination of Organochlorine Pesticide Residues in the Blood of Healthy Individuals
- Reference Values for a Heterogeneous Ferritin Assay and Traceability to the 3rd International Recombinant Standard for Ferritin (NIBSC Code 94/572)
- The Circulating Levels of Cardiac Natriuretic Hormones in Healthy Adults: Effects of Age and Sex
- Plasma Free Amino Acids in Taiwan Chinese: the Effect of Age
- Regression-Based Reference Limits for Urinary Amino Acids in a Pediatric Population
- Evaluation of Uncertainty of Measurement in Routine Clinical Chemistry - Applications to Determination of the Substance Concentration of Calcium and Glucose in Serum
- Comparison of Quick and Owren Prothrombin Time with Regard to the Harmonisation of the International Normalised Ratio (INR) System
- The Accuracy of Calculated Base Excess in Blood
- Aspects of the Development of Clinical Chemistry and Laboratory Medicine in the former East Germany
- University Degree in Clinical Laboratory Sciences
- Process Control Reduces the Laboratory Turnaround Time
- Who Needs Two Different Citrate Concentrations in Sample Tubes Used for Coagulation Tests?
- Quantifying the Capability of Detection of Measurement Procedures in Clinical Laboratory Sciences
- Yet another Example of Lack of Laboratory Expertise in Certain Biomedical Publications
- Counterimmunoelectrophoresis: Fast, Easy and Cost-effective Method for the Detection of Autoantibodies to Intracellular Antigens
- Meetings and Awards