Granularity Index of the SYSMEX XE-5000 hematology analyzer as a replacement for manual microscopy of toxic granulation neutrophils in patients with inflammatory diseases
Abstract
Background: When certain inflammatory processes occur, toxic granulation neutrophils (TGNs) appear in the blood showing prominent cytoplasmic granules. Currently, the granularity of TGNs is analyzed by manual microscopy of blood smears. The SYSMEX XE-5000 is an automated hematology analyzer, which can measure toxic granulation of TGNs by calculating the Granularity (GI) Index. In this study we investigated if the GI-Index is suitable as a parameter for the TGN granularity in inflammatory diseases.
Methods: An evaluation of the toxic granulation neutrophil (TGN) granularity by manual microscopy, the GI-Index and the C-reactive protein (CRP) concentrations of 158 patients were determined. Blood samples from 40 healthy individuals were incubated with lipopolysaccharide (LPS) for in vitro kinetic measurements of the GI-Index. Furthermore, time course measurements of the GI-Index and CRP concentrations of 100 intensive care unit patients were performed.
Results: The GI-Index correlated with the microscopic rating of TGNs (n=158; rs=0.839; p<0.0001). When incubating the blood samples with LPS, the neutrophils displayed hypogranulation 30 min after incubation and a hypergranulation after 90 min. In vivo, the GI-Index indicated changes of the bacterial infection status 1 day earlier than the CRP concentration. The correlation of CRP and GI-Index varied between the patient cohorts (n=158; rs=0.836) (n=100; r=0.177), depending on the cause and extent of inflammation.
Conclusions: The GI-Index is suited to quantify the granularity of TGNs. The GI-Index is an automated, standardized parameter available on a 24 h basis. We suggest that it replace the time-consuming, subjective and semiquantitative microscopic procedure.
©2011 by Walter de Gruyter Berlin Boston
Artikel in diesem Heft
- Editorial
- Method comparison: where do we draw the line?
- Review
- Current role of liquid chromatography coupled to mass spectrometry in clinical toxicology screening methods
- Minireview
- Molecular diagnostics for pharmacogenomic testing of fluoropyrimidine based-therapy: costs, methods and applications
- Opinion Paper
- Preanalytical quality improvement: from dream to reality
- Point
- Towards more complete specifications for acceptable analytical performance – a plea for error grid analysis
- Counterpoint
- Closing the brain-to-brain loop in laboratory testing
- General Clinical Chemistry and Laboratory Medicine
- Harmonization of growth hormone measurements with different immunoassays by data adjustment
- Comparison of the results from two different External Quality Assessment Schemes supports the utility of robust quality specifications
- High-throughput LC-MS/MS method for monitoring sirolimus and everolimus in the routine clinical laboratory
- Performance of the Roche Total Mycophenolic Acid® assay on the Cobas Integra 400®, Cobas 6000® and comparison to LC-MS/MS in liver transplant patients
- Diagnostic efficiency of truncated area under the curve from 0 to 2 h (AUC0–2) of mycophenolic acid in kidney transplant recipients receiving mycophenolate mofetil and concomitant tacrolimus
- Good performance of an immunoassay based method for nevirapine measurements in human breast milk
- Measurement of plasma amino acids by Ultraperformance® Liquid Chromatography
- Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers
- Granularity Index of the SYSMEX XE-5000 hematology analyzer as a replacement for manual microscopy of toxic granulation neutrophils in patients with inflammatory diseases
- Direct chromogenic substrate immuno-capture activity assay for testing of factor VII-activating protease
- The antinuclear antibody assay: developing criteria for reflexive anti-dsDNA antibody testing in a laboratory setting
- A simple liquid chromatography-tandem mass spectrometry method for measuring metanephrine and normetanephrine in urine
- Monoclonal gammopathy missed by capillary zone electrophoresis
- Adaptation of the Diazyme Direct Enzymatic HbA1c Assay for a microplate reader at room temperature
- Validation and Outcome Studies
- Laboratory assessment of iron status in pregnancy
- Cardiovascular Diseases
- Blood cells characteristics as determinants of acute myocardial infarction
- No evidence for an association between the rs2824292 variant at chromosome 21q21 and ventricular fibrillation during acute myocardial infarction in a German population
- Letters to the Editor
- Letter to the Editor Reply: Statistical methods for assessment of added usefulness of new biomarkers
- Reply to Vavrova et al. Clin Chem Lab Med 2011;49:89–92
Artikel in diesem Heft
- Editorial
- Method comparison: where do we draw the line?
- Review
- Current role of liquid chromatography coupled to mass spectrometry in clinical toxicology screening methods
- Minireview
- Molecular diagnostics for pharmacogenomic testing of fluoropyrimidine based-therapy: costs, methods and applications
- Opinion Paper
- Preanalytical quality improvement: from dream to reality
- Point
- Towards more complete specifications for acceptable analytical performance – a plea for error grid analysis
- Counterpoint
- Closing the brain-to-brain loop in laboratory testing
- General Clinical Chemistry and Laboratory Medicine
- Harmonization of growth hormone measurements with different immunoassays by data adjustment
- Comparison of the results from two different External Quality Assessment Schemes supports the utility of robust quality specifications
- High-throughput LC-MS/MS method for monitoring sirolimus and everolimus in the routine clinical laboratory
- Performance of the Roche Total Mycophenolic Acid® assay on the Cobas Integra 400®, Cobas 6000® and comparison to LC-MS/MS in liver transplant patients
- Diagnostic efficiency of truncated area under the curve from 0 to 2 h (AUC0–2) of mycophenolic acid in kidney transplant recipients receiving mycophenolate mofetil and concomitant tacrolimus
- Good performance of an immunoassay based method for nevirapine measurements in human breast milk
- Measurement of plasma amino acids by Ultraperformance® Liquid Chromatography
- Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers
- Granularity Index of the SYSMEX XE-5000 hematology analyzer as a replacement for manual microscopy of toxic granulation neutrophils in patients with inflammatory diseases
- Direct chromogenic substrate immuno-capture activity assay for testing of factor VII-activating protease
- The antinuclear antibody assay: developing criteria for reflexive anti-dsDNA antibody testing in a laboratory setting
- A simple liquid chromatography-tandem mass spectrometry method for measuring metanephrine and normetanephrine in urine
- Monoclonal gammopathy missed by capillary zone electrophoresis
- Adaptation of the Diazyme Direct Enzymatic HbA1c Assay for a microplate reader at room temperature
- Validation and Outcome Studies
- Laboratory assessment of iron status in pregnancy
- Cardiovascular Diseases
- Blood cells characteristics as determinants of acute myocardial infarction
- No evidence for an association between the rs2824292 variant at chromosome 21q21 and ventricular fibrillation during acute myocardial infarction in a German population
- Letters to the Editor
- Letter to the Editor Reply: Statistical methods for assessment of added usefulness of new biomarkers
- Reply to Vavrova et al. Clin Chem Lab Med 2011;49:89–92